K030317 · Midwest Rf, LLC · MOS · Mar 25, 2003 · Radiology
Device Facts
Record ID
K030317
Device Name
MODEL 1300GE-64: PEDIATRIC POSITIONER
Applicant
Midwest Rf, LLC
Product Code
MOS · Radiology
Decision Date
Mar 25, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Magnetic Resonance Imaging (MRI), Magnetic Resonance Angiography (MRA) and Proton Spectroscopy of the pediatric head, body and extremities. This device is used in conjunction with a General Electric 1.5 Tesla MRI System. Proton Spectroscopy is used in conjunction with the PROBE (Proton Brain Exam) option.
Device Story
The Model 1300GE-64 Pediatric Positioner is an accessory for GE Signa 1.5T MRI systems. It provides physical support and positioning for pediatric patients during MRI, MRA, and proton spectroscopy procedures of the head, body, and extremities. The device is used in clinical settings by radiology staff. It does not alter the MRI system's safety parameters, including RF power deposition, acoustic noise, or imaging performance (SNR, uniformity, spatial resolution). By stabilizing the patient, it facilitates high-quality diagnostic imaging, aiding clinicians in accurate diagnosis.
Clinical Evidence
Bench testing only. The device was evaluated for safety and effectiveness by comparing imaging performance parameters (SNR, image uniformity, geometric distortion, slice thickness/gap, and high-contrast spatial resolution) and safety parameters (RF power deposition, acoustic noise, biocompatibility) against the GE Signa 1.5T MRI system baseline. No changes in performance were observed.
Technological Characteristics
MRI accessory for pediatric positioning. Complies with UL-94 (flammability of plastic materials), IEC-60601-1 (general safety for medical electrical equipment), and IEC-60601-2-33 (safety for magnetic resonance equipment). Passive mechanical device; no electronic components or software.
Indications for Use
Indicated for pediatric patients requiring MRI, MRA, or proton spectroscopy of the head, body, or extremities using a GE 1.5T MRI system.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Submission Summary (Full Text)
{0}------------------------------------------------
| 4.0 Attachment 2: | 510(K) Summary of Safety and Effectiveness |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Model 1300GE-64 - Pediatric Positioner |
| Applicability: | Compatible with GE Signa 1.5T MRI Systems |
| Reason for 510(k): | New device |
| Classification Name: | Magnetic Resonance Diagnostic Device |
| Device Classification Panel: | Radiology |
| Device Classification Number: | 892.1000 |
| Product Code: | 90MOS |
| Common Name: | MRI Accessory |
| Proprietary Name: | Model 1300GE-64 - Pediatric Positioner |
| Establishment Registration<br>Number: | 2134565 |
| Address of MFG Facility: | Midwest RF, LLC<br>535 Norton Drive<br>Hartland, WI 53029 |
| Classification: | Class II |
| Intended Uses: | |
| Diagnostic Uses: | 2D, 3D imaging, proton density, T1 and T2<br>weighted imaging, 2D, 3D time of flight, phase<br>contrast imaging. Proton spectroscopy. |
| Anatomic Regions: | Pediatric head, body and extremities. |
{1}------------------------------------------------
Standards:
Performance Standards: None established under Section 514 Voluntary Safety Standards: UL-94 Tests for Flammability of Plastic Materials for Parts in Devices and Appliances International Electrotechnical Commission. IEC-60601-1
- IEC-60601-2-33 International Electrotechnical Commission, Part 2: particular Requirements for the Safety of Magnetic Resonance Equipment for Medical Diagnosis
Requirements for Safety
Medical Electrical Equipment, Part 1: General
## Overview
The Radiology Devices Panel considered potential concerns regarding the safe and effective operation of Magnetic Resonance Diagnostic Devices when thev recommended reclassification to Class II on July 27, 1987. After reclassification. the FDA's Center for Devices and Radiological Health (CDRH) released a draft guidance document for the content and review of Magnetic Resonance Diagnostic Device premarket notification submissions that offered clarification of these concerns. Due to considerable technological advances in MRDDs. CDRH issued an updated guidance document on November 14, 1998. The following is a summary of the information contained within this premarket notification that addresses these concerns:
The GE Signa 1.5T MRI System, operated with the Midwest RF Pediatric Positioner, is substantially equivalent to the GE Signa 1.5T MRI System operated with the legally market predicated devices listed in Attachment 8 of this PMN, within the Class II definition of Magnetic Resonance Diagnostic Device with respect to the safety parameter action levels:
### Safety Parameters
Maximum Static Magnetic Field: No change
Rate of Magnetic Field Strength Change: No change
Pediatric Positioner
{2}------------------------------------------------
| RF Power Deposition: | No change when used in<br>accordance with operator<br>instructions |
|-----------------------------------|--------------------------------------------------------------------|
| Acoustic Noise Levels: | No change |
| Biocompatibility: | No change |
| Imaging Performance Parameters | |
| Specification Volume: | No change |
| Signal-to-Noise Ratio: | No change (within measurement<br>error) |
| Image Uniformity: | No change (within measurement<br>error) |
| Geometric Distortion: | No change |
| Slice Thickness and Gap: | No change |
| High Contrast Spatial Resolution: | No change |
# General Safety and Effectiveness Concerns
The device contains detailed instructions for use. It includes indications for use, precautions, cautions, contraindications, warnings and quality assurance testing.
## Substantial Equivalence Summary
The GE Signa 1.5T MRI System, operated with the Midwest RF Pediatric Positioner addressed in this PMN, has the same intended use and technological characteristics as the GE Signa 1.5T MRI System operated with the legally market predicated devices. The use of this device does not alter the GE Signa 1.5T MRI System safety parameter specifications.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Pediatric Positioner
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 5 2003
Mr. Helmut Keidl CEO Midwest R.F., LLC 535 Norton Drive HARTLAND WI 53209 Re: K030317 Trade/Device Name: Model 1300GE-64
Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 MOS Dated: January 28, 2003 Received: January 30, 2003
Dear Mr. Keidl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
5.0 Statement of Indication for Use Attachment 3:
510(K) Number (if known): ____________________________________________________________________________________________________________________________________________________
Pediatric Positioner, Model # 1300GE-64 Device Name:
Indications For Use:
Magnetic Resonance Imaging (MRI), Magnetic Resonance Angiography (MRA) and Proton Spectroscopy of the pediatric head, body and extremities. This device is used in conjunction with a General Electric 1.5 Tesla MRI System. Proton Spectroscopy is used in conjunction with the PROBE (Proton Brain Exam) option.
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-off)<br>Division of Reproductive, Abdominal, ENT,<br>and Radiological Devices | |
|----------------------------------------------------------------------------------------------|--|
| 510(K) Number | |
Prescription Use (Per 21 CFR 801.109) OR Over-the Counter Use (Optional Format 1-2-96)
| Name | Nancy Brogdon |
|-----------------------------------------------------------------------------------------|---------------|
| (Division Sign-Off)<br>Division of Reproductive, Abdominal,<br>and Radiological Devices | |
| 510(k) Number | K030317 |
Pediatric Positioner
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.