← Product Code [GKZ](/productcode/GKZ) · K030263

# IMMUNICON CELLTRACKS ANALYZER (K030263)

_Immunicon Corp. · GKZ · Nov 24, 2003 · Hematology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K030263

## Device Facts

- **Applicant:** Immunicon Corp.
- **Product Code:** [GKZ](/productcode/GKZ.md)
- **Decision Date:** Nov 24, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5220
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

The intended use of the CellTracks device is to enumerate fluorescently labeled leukocyte subsets that are immuno-magnetically selected and aligned.

## Device Story

The CellTracks Analyzer performs static cytometry to enumerate leukocyte subsets. Input: blood samples prepared with ferrofluid-conjugated monoclonal antibodies (targeting CD45) and fluorescently labeled antibodies (anti-CD3-APC, anti-CD4-PE). Operation: sample mixture loaded into CellTracks Cartridge; MagNest fixture uses magnetic field to align labeled cells along nickel lines on cartridge surface. Analyzer uses CD-based optical technology to scan and track aligned cells; software analyzes fluorescent signatures to classify and count cell subsets. Used in clinical laboratory settings by trained personnel. Output: quantitative cell counts. Benefits: provides automated, standardized enumeration of specific leukocyte populations for clinical assessment.

## Clinical Evidence

Clinical study of 149 HIV+ patients and 150 normal donors. Evaluated CD3 and CD4 lymphocyte counts across three sites. Compared CellTracks Analyzer to BD FACSCalibur. Results: CD4 slope 0.95, R2 0.9802; CD3 slope 0.77, R2 0.9375. Nonclinical testing included 20-day precision (CVs 3.68%-6.60%) and linearity (0-2,000 cells/ul). Interference testing showed no adverse effects from lipids, hematocrit, platelets, WBCs, or free hemoglobin.

## Technological Characteristics

Static cytometry system using immuno-magnetic selection. Components: MagNest (steel-yoked magnets), CellTracks Cartridge with nickel lines, CD-technology optical scanner. Reagents: ferrofluids (75-100nm radius magnetic core with polymeric layer) and fluorescent monoclonal antibodies. Connectivity: Bar code sample identification. Software: Risk analysis per ISO 14971; documentation per FDA guidance.

## Regulatory Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

## Predicate Devices

- Becton Dickinson FACSCalibur System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Page 1 of 3

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY INSTRUMENT ONLY TEMPLATE

A. 510(k) Number:
K030263

B. Instrument Name:
Immunicon CellTracks™ Analyzer

C. System Descriptions:
1. Modes of Operation:
Scanning Mode and Imaging Mode
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☐ or No ☐

In lieu of a Hazard Analysis Immunicon submitted a Risk Analysis which conforms to ISO 14971, ISO: Medical Devices—Application of Risk Management to Medical Devices. Immunicon did submit software documentation which conforms to FDA’s software documentation guidance document.

3. Sample Identification:
The CellTracks Cartridge uses Bar Code labels

4. Specimen Sampling and Handling:
Specimen sampling and handling is manual using Immunicon Cell Save sample preparation tubes.

5. Assay Types:
Static cytometry

6. Reaction Types:
Fluorescently labeled leukocyte subsets are immuno-magnetically selected and aligned in a scanning cytometer.

7. Calibration:
Calibration is accomplished by running the Calibration test protocol each day of patient testing using the re-usable CellTracks Calibrator Cartridge which contains photo-stable fluorescent beads that are fixed in position in a solid matrix.

8. Quality Control:
BD Multi-Check control liquid quality control reagents manufactured by Becton Dickinson are recommended to check the instrument, reagents, and operator technique

D. Other Supportive Performance Characteristics Data Not Covered In The “L. Performance Characteristics” Section Of The SE Decision Summary.
1. Clinical Testing:
Accuracy was tested using NCCLS EP9-A, Method Comparison and Bias Estimation Using Patient Samples. Three geographically diverse clinical

{1}

Page 2 of 3

study sites were used with blood samples from 149 HIV+ patients and 150 samples from normal donors. The Cell Tracks device was compared with the Becton Dickinson FACSCalibur System. Results of the testing indicated a slope of .95 and an R² of 0.9802 for CD4 and a slope of 0.77 and an R² of 0.9375 for CD3.

2. Nonclinical Testing:

a. Linearity: 0-2,000 cells/μl.
b. Precision: A 20 day precision study was performed according to NCCLS EP-5-A, Evaluation of Precision Performance of Clinical Chemistry Devices using three CellTracks Analyzers and Becton Dickinson BD Multi-Check Controls. The normal CD3 control within-run CV’s ranged from 3.68% to 4.34%. The low CD3 control within-run CV’s ranged from 3.89% to 6.60%.
c. Interference Testing: Lipid levels from 0 to @ 5,000 mg/dL, hematocrits from 43.3 to 76.2%, platelet counts from 210 x 10³ to 2,040 x 10³, and free hemoglobin in plasma levels of 10.3 to 12.4 g/dL had adverse affect on the CD3 or CD4 counts.

E. Other Supportive Information:

1. Intended Use: The intended use of the CellTracks device is to enumerate fluorescently labeled leukocyte subsets that are immuno-magnetically selected and aligned.

2. Principles of Operation: The CellTracks analyzer is used in conjunction with reagents containing ferrofluids and fluorescent labeled monoclonal antibodies. Ferrofluid consists of a magnetic core surrounded by a polymeric layer coated with antibodies for capturing target cells. In this example, all leukocytes are immuno-magnetically labeled by targeting the CD45 antigen. Ferrofluid particles are colloidal (75-100 nanometers in radius) and when mixed with a sample containing the target white blood cells, the anti-CD45 monoclonal antibodies conjugated to the ferrofluid (anti-CD45 ferrofluid) bind with the CD45 antigen associated with the white blood cells. Fluorescent reagents (Becton Dickinson reagents, and anti-CD3-APC and anti-CD4-PE, are added for identification and enumeration of cell subsets. The reagent/sample mixture is dispensed into a CellTracks Cartridge that is inserted into a MagNest, a fixture of two magnets yoked together by steel. The strong magnetic field causes the labeled target cells to move to the surface of the cartridge, where the CellTracks analyzer uses compact disk (CD) technology to focus and track along nickel lines on the surface of the Cell Tracks Cartridge and to analyze the fluorescent signatures of aligned cells. The software enumerates and classifies cell subsets based on fluorescence.

3. Regulation Number: 21 CFR 864.5220, Automated differential cell counter

4. Device Classification: Class II

{2}

Page 3 of 3

5. Classification Panel: Hematology and Pathology Devices Panel (81)

6. Product Code: GKZ

F. Conclusion:
The Immunicon CellTracks Analyzer has been shown to be substantially equivalent to the Becton Dickinson FACSCalibur System (K840195).

---

**Source:** [https://fda-staging.innolitics.com/device/K030263](https://fda-staging.innolitics.com/device/K030263)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
