K030237 · Kapp Surgical Instrument, Inc. · KWI · Aug 29, 2003 · Orthopedic
Device Facts
Record ID
K030237
Device Name
KAPP CUSTOM RADIAL HEAD ELBOW IMPLANT
Applicant
Kapp Surgical Instrument, Inc.
Product Code
KWI · Orthopedic
Decision Date
Aug 29, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3170
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The KAPP Radial Elbow Implant is intended to be used for: replacement of the proximal end of the radius: Replacement of the radial head for degenerative, or post-traumatic disabilities presenting pain, crepitation and decreased motion at the radio-humeral and or proximal radio-ulnar joint with either joint destruction or subluxation visible on xray and or resistance to conservative treatment; Primary replacement after fracture of the radial head; Symptomatic sequelae after radial head resection; Revision following failed radial head arthroplasty. The device is intended for custom single use with bone cement.
Device Story
Custom radial head elbow implant; replaces proximal end of radius. Device consists of implant and sizers; fixed via intermedullary canal stem using bone cement. Used by orthopedic surgeons in clinical/surgical settings to treat degenerative or post-traumatic joint conditions, fractures, or failed prior arthroplasty. Provides mechanical replacement of radial head to restore joint function and reduce pain. Single-use device.
Clinical Evidence
No clinical data provided; substantial equivalence based on design and material characteristics.
Indicated for patients requiring proximal radial head replacement due to degenerative or post-traumatic disabilities (pain, crepitation, decreased motion, joint destruction, or subluxation), primary fracture, symptomatic sequelae post-resection, or revision of failed arthroplasty.
Regulatory Classification
Identification
An elbow joint radial (hemi-elbow) polymer prosthesis is a device intended to be implanted made of medical grade silicone elastomer used to replace the proximal end of the radius.
Predicate Devices
Avanta Radial Head Implant
Wright Medical Metallic Radial Head Implants
Submission Summary (Full Text)
{0}------------------------------------------------
Kapp Surgical Instrument, Inc.
# AUG 2 9 2003
K030237 PG 142
## Attachment 1 - 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92, upon which the substantial equivalence determination is based.
#### Summary Information:
| Applicant: | Kapp Surgical Instrument Co., Inc.<br>4919 Warrensville Center Road<br>Warrensville Heights, Ohio 44128<br>Tel: (216) 587-4400<br>Fax: (216) 587-0411 |
|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact : | Albert Santilli, President |
| Prepared: | January 22, 2003 |
## Device Identification:
| Proprietary Name: | Kapp Custom Radial Head Elbow Implant |
|----------------------------------------|--------------------------------------------------------------------------------|
| Common Name: | Radial Head Implant |
| Classification Name<br>and Regulation: | Prosthesis, Elbow, Hemi-radial, polymer,<br>21 CFR 888.3170, 87KWI |
| Predicate Device(s): | Avanta Radial Head Implant and<br>Wright Medical Metallic Radial Head Implants |
#### Device Description:
The Kapp Custom Radial Head Elbow Implant includes various sizes of implants and accessories including sizers. The implant allows for replacement of the proximal radial head.
#### Indications for use:
The KAPP Custom Radial Elbow Implant is intended to be used for: replacement of the proximal end of the radius: Replacement of the radial head for degenerative, or posttraumatic disabilities presenting pain, crepitation and decreased motion at the radiohumeral and or proximal radio-ulnar joint with either joint destruction or subluxation visible on x-ray and or resistance to conservative treatment; primary replacement after fracture of the radial head; symptomatic sequelae after radial head resection; revision following failed radial head arthroplasty. It is intended for custom single use with bone cement.
{1}------------------------------------------------
Comparison to the Predicate Devices:
page 2 of 2
The legally marketed predicate devices to which this device is substantially equivalent is the Avanta Radial Head Implant and Wright Medical Technologies Metallic Radial Head Implants.
| Attribute | Candidate Device | Predicate Devices | |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name | Kapp Custom Radial<br>Head Elbow Implant | Advanta Radial Head<br>Implant | WMT Metallic Radial<br>Head Implants |
| Use | Single Use | Single Use | Single Use |
| Fixation | Intermedullary Canal<br>Stem | Intermedullary Canal<br>Stem | Intermedullary Canal<br>Stem |
| Constraint | Non-constrained | Non-constrained | Non-constrained |
| Material | 316L Stainless Steel | Co-Cr/CpTi | Ti, Ti6Al4V, Co-Cr |
| Sizes | 6 custom sizes – 2, 2.5,<br>3, 4, 5, and 6 | 3 sizes: 1, 2, 3 | 7 sizes: 0, 1, 1.5, 2, 2.5,<br>3, 4 |
| Indications for use: | The KAPP Custom<br>Radial Elbow Implant is<br>intended to be used for:<br>replacement of the<br>proximal end of the<br>radius: Replacement of<br>the radial head for<br>degenerative, or post-<br>traumatic disabilities<br>presenting pain,<br>crepitation and<br>decreased motion at the<br>radio-humeral and or<br>proximal radio-ulnar<br>joint with either joint<br>destruction or<br>subluxation visible on<br>x-ray and or resistance<br>to conservative<br>treatment; primary<br>replacement after<br>fracture of the radial<br>head; symptomatic<br>sequelae after radial<br>head resection; revision<br>following failed radial<br>head arthroplasty. It is<br>intended for custom<br>single use with bone<br>cement. | The Advanta Radial<br>Head Implant is<br>intended for:<br>replacement of the<br>proximal end of the<br>radius: Replacement of<br>the radial head for<br>degenerative, or post-<br>traumatic disabilities<br>presenting pain,<br>crepitation and<br>decreased motion at the<br>radio-humeral and or<br>proximal radio-ulnar<br>joint with either joint<br>destruction or<br>subluxation visible on<br>x-ray and or resistance<br>to conservative<br>treatment; primary<br>replacement after<br>fracture of the radial<br>head; symptomatic<br>sequelae after radial<br>head resection; revision<br>following failed radial<br>head arthroplasty. | The WMT Metallic<br>Radial Head Implants<br>are intended for:<br>replacement of the<br>radial head for<br>degenerative, or post-<br>traumatic disabilities<br>presenting pain,<br>crepitation and<br>decreased motion at the<br>radio-humeral and or<br>proximal radio-ulnar<br>joint with: joint<br>destruction and or<br>subluxation visible on<br>x-ray and or resistance<br>to conservative<br>treatment; primary<br>replacement after<br>fracture of the radial<br>head; symptomatic<br>sequelae after radial<br>head resection; revision<br>following failed radial<br>head arthroplasty. |
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, with three stylized wing shapes.
Public Health Service
AUG 2 9 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Albert N. Santilli President KAPP Surgical Instrument Company, Inc. 4919 Warrensville Center Rd. Warrensville Heights, Ohio 44128
K030237 Re:
> Trade/Device Name: KAPP Custom Radial Head Elbow Implant Regulation Number: 21 CFR 888. 3170 Regulation Name: Elbow joint radial (hemi-elbow) polymer prosthesis Regulatory Class: II Product Code: KWI Dated: June 6, 2003 Received: June 11, 2003
Dear Mr. Santilli:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Albert N. Santilli
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mach n Millerson
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Attachment 3 - Statement of Indications for Use
510(k) Number:
Device Name: KAPP Surgical Instrument, Inc. Radial Head Elbow Implant
Intended Use / Indications for Use:
The KAPP Radial Elbow Implant is intended to be used for: replacement of the proximal end of the radius:
Replacement of the radial head for degenerative, or post-traumatic disabilities presenting pain, crepitation and decreased motion at the radio-humeral and or proximal radio-ulnar joint with either joint destruction or subluxation visible on xray and or resistance to conservative treatment;
Primary replace nent after fracture of the radial head;
Symptomatic sequelae after radial head resection;
Revision following failed radial head arthroplasty
The device is inlended for custom single use with bone cement
(PLEASE DO NOT WRITE BELOW THIS LINE/CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
()
for Mark N. Millham
Division of General, Festorative and Neurological De
510(k) Number ***_***_ K03023t
Prescription Use
(per 21 CFR 801.109)
OR
Over the Counter Use
Optional Format 1-2-96
A3-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.