HAND INNOVATIONS, INC. DISTAL RADIUS FRACTURE REPAIR SYSTEM
K030198 · Hand Innovations, Inc. · HRS · Apr 1, 2003 · Orthopedic
Device Facts
Record ID
K030198
Device Name
HAND INNOVATIONS, INC. DISTAL RADIUS FRACTURE REPAIR SYSTEM
Applicant
Hand Innovations, Inc.
Product Code
HRS · Orthopedic
Decision Date
Apr 1, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Distal Radius Fracture Repair System is intended for the fixation of fractures and osteotomies involving the distal radius.
Device Story
Distal Radius Fracture Repair System is a metallic bone fixation kit for distal radius fractures and osteotomies. System includes volar stabilization plates, dorsal intrafocal nail-plates, bone screws, and fixation pegs. Components are manufactured from Titanium TI-6AL-4V ELI and Stainless Steel. Device is used by orthopedic surgeons in a clinical/surgical setting to stabilize bone fragments. Components are provided non-sterile and require steam sterilization by healthcare professionals prior to implantation. The system is designed for permanent implantation to facilitate bone healing.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing and comparison of design, materials, and intended use to the predicate device.
Technological Characteristics
Metallic bone fixation appliance (21 CFR 888.3030). Materials: Titanium TI-6AL-4V ELI (plates, pegs, screws) and Stainless Steel 440C/17-4 (jigs, drivers). Components include volar plates, dorsal nail-plates, screws, and pegs. Provided non-sterile for steam sterilization. Permanent implant.
Indications for Use
Indicated for fixation of fractures and osteotomies of the distal radius in patients requiring surgical stabilization.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
K030198
page 1 of 4
# 1. Manufacturer and Contact Information:
| Manufacturer<br>& U. S. Distributor | Hand Innovations, Inc.<br>8905 S. W. 87th Avenue<br>Miami, FL 33175-2227 |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Information: | Richard D. Bliss, Jr.<br>Quality Systems Engineering<br>510 Stonemont Drive<br>Weston, FL 33326<br>Telephone: (954) 385-1690<br>Fax: (954) 385-1256 |
### 2. Device Classification Name:
The Orthopedic Devices Panel has classified the single/multiple component metallic bone fixation appliances and accessories as Class II. Reference 21 CFR 888.3030.
## 3. Intended Use:
The Distal Radius Fracture Repair Kit is intended for the fixation of fractures and osteotomies involving the distal radius.
### 4. Device Description and Characteristics
This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) 1990 and the FDA Modernization Act of 1997 (FDAMA).
The Distal Radius Fracture Repair System was previously cleared under 510(k) No. K023007. The system consists of a volar stabilization plate, bone screws, fixation pegs, dorsal intrafocal nail-plate, three fixed angle bone pegs and two locking screws. This 510(k) is being submitted as a modification to the original 510(k) No. K023007 to update and standardize the sterilization cycle. A copy of the clearance letter is enclosed with this submission.
The Distal Volar Radius Plate (DVR) consists of a stabilization plate, bone screws, and fixation pegs. The 3.5 mm screws are used to affix the proximal segment of the plate to the diaphysis. Pegs or screws are used for the distal bone fragment(s).
The Distal Dorsal Nail-Plate (DNP) is a bone stabilization device consisting of an intrafocal nail-plate to which three-fixed angle bone pegs and two locking screws are attached. It has a narrow distal plate-like section that lies on the surface of the distal fragment and a proximal nail-like section that is introduced into the diaphysis of the radius through the fracture site. Fixed angle pegs are used to fix the distal fragment(s) to the plate section and locking screws are used to lock the proximal fragment inside the radial shaft.
{1}------------------------------------------------
K030198
page 2 of 4
Other components used in the implantation process are identified as the DNP Jig Set of Stainless Steel SST 17-4. These items consist of the DNP Jig (left or right), drill guide, screw jig and screw guide.
A standard awl, which is a manual tool, used to increase the size of a hole or tunnel by scraping in rotation is a standard catalog item manufactured by from K-Medic under catalog no. KM-48-336.
The components of this system will be packaged together and will also be available individually. The materials used for the various components in Distal Radius Fracture Repair System include the following:
| Name | Part No. | Material Composition |
|-----------------------------------------------------------------------------------------|------------------------------|-----------------------------|
| Volar Plate, Right | | Titanium TI-6AL-4V ELI |
| Volar Plate, Left | | Titanium TI-6AL-4V ELI |
| Dorsal Nail Plate Right | DRW-067 | Titanium TI-6AL-4V ELI |
| Dorsal Nail Plate Left | DRW-054 | Titanium TI-6AL-4VELI |
| Peg Volar Plate<br>(14,16,18,20,22,24,26,28mm) 010, DRW-055/DRW-056 (Table DRW-20) | DRW-005/056 | Titanium TI-6AL-4V ELI |
| Threaded peg 2.5 mm<br>(14,16,18,20,22,24,26,28mm) 033, DRW-057/DRW-058 (Table DRW-026) | DRW-028/033 | Titanium TI-6AL-4V ELI |
| Screw, 2.5mm (14,16,18,20,22,24,26,28mm) | DRW-034/039, DRW-059/DRW-060 | Titanium TI-6AL-4V ELI |
| Peg Driver Assembly | DRW-016 | Stainless Steel 440C |
| Drill Guide | DRW-017 | Stainless Steel 440C |
| DNP Jig, Left | DRW-062 | Stainless Steel SST-17-4 |
| DNP Jig, Right | DRW-063 | Stainless Steel SST-17-4 |
| DNP Screw Jig | DRW-064 | Stainless Steel SST-17-4 |
| DNP Screw Guide | DRW-065 | Stainless Steel SST-17-4 |
| DNP Drill Guide | DRW-066 | Stainless Steel SST-17-4 |
| Depth Gauge | Supplied by K-Medic | Off the shelf catalog item. |
| Drill, 2.0mm | Supplied by Microaire | Off the shelf catalog item. |
| Awl | Supplied by K-Medic | Off the shelf catalog item. |
| Name | Part No. | Material Composition |
| Screw Holding Sleeve | Supplied by K-Medic | Off the shelf catalog item. |
| Drill, 2.5 mm | Supplied by Microaire | Off the shelf catalog item. |
| Drill, 1.8 mm | Supplied by Microaire | Off the shelf catalog item. |
| Hex Driver | Supplied by K-Medic | Off the shelf catalog item. |
| Plate Bender | Supplied by K-Medic | Off the shelf catalog item. |
| Tissue Guide | Supplied by K-Medic | Off the shelf catalog item. |
| Screw Rack | DRW-018 | Not a device. |
| Sterilization Tray | DRW-019 | Not a device. |
{2}------------------------------------------------
Kozol98
The components within this system will be provided as non-sterile for steam sterilization by health care professionals prior to use. Instructions for sterilizations are contained in the package insert. The kit components will also be available separately provided as non-sterile for steam sterilization by the healthcare professional.
The system is packaged in a high tempered plastic sterilization tray. The tray is provided with inserts to retain the components. The tray is placed in a polymeric bag and placed into a shipping carton. All components sold separately are packaged in polymeric pouches.
See Modification of Device Section for the Kit Manufacturer's Certification Statement containing a complete listing of all components.
#### 5. Substantial Equivalence
Hand Innovations Inc. considers the modifications to this Distal Radius Fracture Repair System to be substantially equivalent to the Distal Radius Fracture Repair System. 510(k) No. K023007 that cleared FDA on December 5, 2002. with regard to the intended use, materials, biocompatibility and overall performance characteristics. The current labeling for the Distal Radius Fracture Repair System has been revised to reflect the change in the sterilization cycle.
{3}------------------------------------------------
030198
page 4 of 4
# Summary of Substantial Equivalence
| Feature | Distal Radius Fracture<br>Repair Kit | Distal Radius Fracture<br>Repair Kit<br>(Predicate Device) |
|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Hand Innovations, Inc. | Hand Innovations, Inc. |
| Packaging | Tempered Plastic suitable<br>for steam sterilization | Tempered Plastic suitable<br>For stem sterilization. |
| Sterilization method | Provided non-sterile<br>Recommend steam<br>sterilization | Provided non-sterile<br>Recommend steam<br>sterilization |
| Intended use | Distal Radius Fracture<br>Repair System is intended<br>for the fixation of fractures<br>and osteotomies involving<br>the distal radius. | Distal Radius Fracture<br>Repair System is intended<br>for the fixation of fractures<br>and osteotomies involving<br>the distal radius. |
| Implant Period | Permanent | Permanent |
| Material of Construction | Plates: Titanium | Plates: Titanium |
| Available configurations | Pegs and Screws: Titanium<br>and Stainless Steel<br>Right and Left<br>Volar and Dorsal | Pegs and Screws: Titanium<br>And Stainless Steel<br>Right and Left<br>Volar and Dorsal |
| No. of buttress pegs | 5 each size | 5 each size |
| No. of Cortex Screws | 8 each size | 8 each size |
| Buttress Peg Length mm | 16, 18, 20, 22, 24, 26, 28 | 16, 18, 20, 22, 24, 26, 28 |
| Cortex Screw Length mm | 10, 12, 14, 16, 18 | 10, 12, 14, 16, 18 |
| Specialized instruments<br>included in tray | Bending tool<br>Drill Guide<br>Depth Gauge<br>Peg Driver<br>Screw Driver<br>DNP Jig Right & Left | Bending Tool<br>Drill Guide<br>Depth Gage<br>Peg Driver<br>Screw Driver<br>DNP Jig Right & Left |
| Tool and component<br>separators and holders in<br>tray | Yes | Yes |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with three stylized wing-like shapes.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 0 1 2003
Hand Innovations, Inc. c/o Mr. Richard D. Bliss, Jr. Ouality System Engineering 510 Stonemont Drive Weston, Fl 33326
Re: K030198
Trade Name: Distal Radius Fracture Repair System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HRS Dated: January 20, 2003 Received: January 21, 2003
Dear Mr. Bliss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Mr. Richard D. Bliss, Jr.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely vours.
Mark A. Mckennon
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
030198
Device Name:
Distal Radius Fracture Repair System
Indications for Use:
The Distal Radius Fracture Repair System is intended for the fixation of fractures and osteotomies involving the distal radius.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use or Over-The-Counter Use (Pre 21 CFR 801.109) (Optional Format 1-2-96)
Mark A. Milliman
ivision Sign-Off) Division of General, Restorative al Devices
90(k) Number ***_***_ K030198
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.