IDEAL: SYSTEM WITH INTEGRATED VENOUS AIR REMOVAL, CENTRIFUGAL BLOOD PUMP, PUMP BRACKET, ADULT MEMBRANE OXYGENATOR
Applicant
Dideco S.P.A.
Product Code
KFM · Cardiovascular
Decision Date
Mar 31, 2003
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.4360
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The IDEAL, System with Integrated Venous Air Removal, Centrifugal Blood Pump, Pump Bracket, Adult Membrane Oxygenator, Heat Exchanger and Arterial Fitter hereinafter called the IDEAL is a sterile, nonpyrogenic device intended to be used in surgical procedures requiring extracorporeal gas exchange support. The Ideal is intended for use in operations on adult patients. The Ideal must not be used longer than 6 hours. Contact with blood for longer periods is inadvisable. The ldeal is intended for use with Stockert Centrifugal Pump Console
Device Story
The IDEAL is an extracorporeal hemodynamic and gas exchange support system for cardiopulmonary bypass. It integrates a hollow fiber membrane oxygenator, heat exchanger, arterial filter, active venous air removal device (defoamer), and a centrifugal blood pump. The system is designed for use in adult surgical procedures requiring extracorporeal support, operated by clinical staff in conjunction with a Stöckert Centrifugal Pump Console. The device functions by circulating blood through the circuit via centrifugal force, providing gas exchange, temperature control, and filtration of gaseous emboli and aggregates. Integration of the venous air removal and pump modules into the oxygenator assembly reduces operating blood volumes compared to traditional setups. The device is for single-use only and limited to a maximum duration of 6 hours.
Clinical Evidence
Bench testing only. In vitro studies evaluated operating blood volumes, hemolysis/cell depletion, mechanical integrity, and air handling characterization. Testing followed ISO 7199 and FDA guidance for oxygenators, defoamers, and arterial filters. Results confirmed performance comparable to predicate devices.
Technological Characteristics
Integrated extracorporeal system; hollow fiber membrane oxygenator; heat exchanger; arterial filter; centrifugal blood pump; active venous air removal (defoamer). Materials identical to Synthesis and Revolution predicates; phosphorylcholine coating. Ethylene oxide sterilized; single-use. Complies with ISO 10993-1:1995 for biocompatibility and ISO 7199 for oxygenators.
Indications for Use
Indicated for adult patients undergoing surgical procedures requiring extracorporeal gas exchange support. Contraindicated for use longer than 6 hours.
Regulatory Classification
Identification
A nonroller-type cardiopulmonary and circulatory bypass blood pump is a prescription device that uses a method other than revolving rollers to pump the blood through an extracorporeal circuit for periods lasting less than 6 hours for the purpose of providing either:(i) Full or partial cardiopulmonary bypass (i.e., circuit includes an oxygenator) during open surgical procedures on the heart or great vessels; or(ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava.
Special Controls
*Classification* —Class II (special controls). The special controls for this device are:(i) Non-clinical performance testing must perform as intended over the intended duration of use and demonstrate the following: Operating parameters, dynamic blood damage, heat generation, air entrapment, mechanical integrity, and durability/reliability;
(ii) The patient-contacting components of the device must be demonstrated to be biocompatible;
(iii) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components; and
(iv) Labeling must include information regarding the duration of use, and a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(b)
*Nonroller-type temporary ventricular support blood pump* —(1)*Identification.* A nonroller-type temporary ventricular support blood pump is a prescription device that uses any method resulting in blood propulsion to provide the temporary ventricular assistance required for support of the systemic and/or pulmonary circulations during periods when there is ongoing or anticipated hemodynamic instability due to immediately reversible alterations in ventricular myocardial function resulting from mechanical or physiologic causes. Duration of use would be less than 6 hours.(2)
*Classification.* Class III (premarket approval).(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with FDA on or before September 8, 2015, for any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976, or that has, on or before September 8, 2015, been found to be substantially equivalent to any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976. Any other nonroller-type temporary ventricular support blood pump shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
K030154
MAR 3 1 2003
### 510(k) SUMMARY
SUBMITTER:
CONTACT PERSON:
DATE PREPARED:
DEVICE TRADE NAME:
COMMON NAMES:
CLASSIFICATION NAMES:
PREDICATE DEVICES:
Dideco S.p.A. 86. Via Statale 12 Nord 41037 Mirandola (MO) Italy
Luigi Vecchi Phone: 011 39 0535 29811 Fax: 011 39 0535 25229
January 15, 2003
IDEAL: System with Integrated Venous Air Removal, Centrifugal Blood Pump, Pump Holder, Adult Membrane Oxygenator, Heat Exchanger and Arterial Filter
Hollow Fiber Membrane Oxygenator with Integrated Arterial Filter and Heat Exchanger Venous Defoamer Centrifugal Blood Pump
Cardiopulmonary Bypass Oxygenator Cardiopulmonary Bypass Heat Exchanger Cardiopulmonary Bypass Blood Reservoir Cardiopulmonary Bypass Defoamer Cardiopulmonary Bypass Arterial Line Blood Filter Non-Roller Type Cardiopulmonary Bypass Blood Pump
Membrane Oxygenator With SYNTHESIS: Adult Integrated Arterial Filter and Hardshell Venous/Cardiotomy Reservoir Mimesys treated (Phosphorylcholine coating hereinafter called PC coating) (k022450)
Cobe Revolution Centrifugal Blood Pump, (K011835)
### DEVICE DESCRIPTION:
The Ideal, System with Integrated Venous Air Removal, Centrifugal Blood Pump, Pump Holder, Adult Membrane Oxygenator, Heat Exchanger and Arterial Filter is an extracorporeal hemodynamic and gas exchange support system for extracorporeal perfusion. The Ideal consists of a high efficiency, microporous, hollow fiber membrane oxygenator integrated with a heat exchanger and an arterial filter (Synthesis Adult Membrane Oxygenator, K022450), connected to an active venous air removal device (defoamer), a centrifuqal pump (Cobe Revolution Centrifyial Blood Pump. K011835) and a pump bracket.
## INDICATION FOR USE:
The IDEAL, System with Integrated Venous Air Removal, Centrifugal Blood Pump, Pump Bracket, Adult Membrane Oxygenator, Heat Exchanger and Arterial Filter hereinafter called the IDEAL is a sterile, nonpyrogenic device intended to be used in surgical procedures requiring extracorporeal gas
{1}------------------------------------------------
The IDEAL, System with Integrated Venous Air Removal, Centrifugal Blood Pump, Pump Bracket, Adult Membrane Oxygenator, Heat Exchanger and Arterial Fitter hereinafter called the IDEAL is a sterile, nonpyrogenic device intended to be used in surgical procedures requiring extracorporeal gas exchange support. The Ideal is intended for use in operations on adult patients. The Ideal must not be used longer than 6 hours. Contact with blood for longer periods is inadvisable. The ldeal is intended for use with Stockert Centrifugal Pump Console
#### TECHNOLOGICAL CHARACTERISTICS:
The Ideal is comprised of an adult membrane oxygenator with an integrated heat exchanger and arterial filter identical to the Synthesis predicate device. The coating is identical to the phosphoryline coating used on the Synthesis predicate device. The Ideal design has been completely reviewed with reference to the defoaming system as compared to the Synthesis' predicate device defoaming system.
The air removal system (defoamer) of the Ideal replaces the air removal function of ordinary venous reservoirs by centralizing venous line air and effectively removing it in a way substantially equivalent to the Synthesis hardshell venous/cardiotomy reservoir. The basic function of all integrated defoamer/oxygenators is the same. That is, a combination blood-gas exchange device with a separate active venous air removal section (defoamer).
The Ideal with reference to the integrated centrifugal blood pump is also identical to the Revolution Pump predicate device. The basic function of all centrifugal pumps is the same, that is, moving blood through the cardiopulmonary bypass circuit by centrifugal force.
Therefore the Ideal duplicates the functionality of the separate predicate devices.
The Ideal is ethylene oxide sterilized and has a nonpyrogenic fluid path. It is for single use only.
#### BIOCOMPATIBILITY TEST RESULTS:
A complete battery of tests were carned out in accordance with the requirements of ISO 10993-1:1995 and the FDA May 1, 1995 Memorandum on the use of the ISO 10993 standard for biocompatibility testing on the raw materials. The Synthesis device was tested for Hemolysis, Cytotoxicity, Irritation, Acute Systemic Toxicity and Mutagenicity. The Revolution device was tested for Hemolysis. Cytotoxicity, Irritation, Acute Systemic Toxicity and Mutagenicity. Biocompatibility testing performed on Synthesis and Revolution predicate devices have been taken as reference for the Ideal system as the raw materials used in the manufacturing process are identical to those used in both predicate devices. Sterility, Pyrogenicity, ETO residuals and package integrity testing were also conducted. The results of this testing met established specifications.
#### IN VITRO TEST RESULTS:
In vitro testing was carried out in accordance with the requirements of "Guidance for Cardiopulmonary Bypass Oxygenators 510(k) submissions - Final Guidance for Industry and FDA Staff" issued on November 13, 2000 - "Guidance for Blood Extracorporeal Blood Circuit Defoamer 510(k) Submission" Final Guidance for industry and FDA issued on November 29, 2000 -"Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submission" Final Guidance for industry and FDA issued on November 29, 2000 and when applicable, following the ISO 7199 (1996) standard for "Cardiovascular Implants and Articial Organs - Extra Corporeal Blood-Gas Exchangers (Oxygenator)" for providing the data necessary to demonstrate both the substantial equivalence with the predicate devices and also complying with safety and effectiveness requirements. The device aged up to 3 years was tested for operating blood volumes, hemolysis/cell deplation, mechanical integrity and air handling characterization of the integrated venous air removal device. The results of this tests met established specifications. For comparative purposes, the same testing, when applicable, has been conducted also on the Synthesis and Revolution predicate.
The result of the study showed that the device is comparable to the predicate devices concerning with all characteristics.
{2}------------------------------------------------
#### CONCLUSIONS:
The results of in vitro studies demonstrate that the Ideal performs in a manner substantially equivalent to the Synthesis with respect to the main function of ordinary venous reservoirs (defoamers) that is separation and evacuation of gaseous emboli form the venous blood. Previous in-vitro data showed that the Ideal performs in a manner substantially equivalent to the Synthesis with respect to the expected main performance parameters associated with ordinary oxygenators with integrated heat exchanger and arterial filter that is transfer of oxygen and removal of carbon dioxide, blood temperature control, removal of arterial blood components aggregates larger than 40 u. Furthermore, the Ideal performs in a manner substantially identical to the centrifying pump Revolution predicate device with respect to the basic function of an ordinary centrifygal pump that is propulsion of blood through the cardiopulmonary bypass circuit by centrifugal force.
Data collected show that integration of the active venous air removal device (defoamer) and of the centrifugal pump to the oxygenating module is advantageous in terms of lower operating blood volumes during procedures. Moreover air handling test results indicate that air bubbles are equally well handled and purged from venous blood by the Ideal integrated venous air removal device with respect to traditional extracorporeal cardiopulmonary bypass reservoirs.
Therefore it can be concluded that Ideal integrated system duplicates the functionality of the separated predicate devices. Biocompatibility and functional performances of phosphorylcholine coated devices, proved for the Synthesis predicate device, have been taken as reference for the ldeal. Additional testing has also demonstrated the effectiveness of production techniques to assure that the oxygenator is sterile and non-pyrogenic.
{3}------------------------------------------------
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.
MAR 31 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Dideco S.p.A. c/o Mr. Barry Sall Parexel International Corp. 195 West Street Waltham, MA 02451-1163
# Re: K030154
IDEAL, System with Integrated Venous Air Removal, Centrifugal Blood Pump, Pump Holder, Adult Membrane Oxygenator, Heat Exchanger and Arterial Filter Regulation Number: 21 CFR 870.4360 Regulation Name: Nonroller Type CPB Blood Pump Regulatory Class: Class III (three) Product Code: KFM Dated: January 15, 2003 Received: January 16, 2003
# Dear Mr. Sall:
We'have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 -- Mr. Barry Sall
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Ana Hlusherfor
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "dideco" in a stylized, bold, sans-serif font. The letters are connected, with the "d" and "i" forming a single shape, and the "e" and "c" similarly joined. The overall design is compact and visually unified, with a slightly rounded appearance.
DIDECO S.p.A.
510(k) Number (if known): ___________________________________________________________________________________________________________________________________________________
Device Name: IDEAL System with Integrated Venous Air Removal, Centrifygal Blood Pump, Pump Holder, Adult Membrane Oxygenator, Heat Exchanger and Arterial Filter
Indications For Use:
Ideal is intended to be used in surgical procedures requiring extracorporeal gas exchange support. Ideal is intended for use in operations on adult patients. Ideal must not be used longer than 6 hours. Contact with blood for longer periods is inadvisable. Ideal is intended for use with the Stöckert Centrifugal Pump Console.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR Over-The-Counter Use
Dva Hlulu
510
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.