The AmeriWater Portable RO+ Model MROS09 is a stand-alone water treatment system for use in hemodialysis applications. It is designed to pretreat and purify potable water for use in making dialysate for hemodialysis. The AmeriWater Portable RO+ with Onboard Pretreatment System is intended to be used in hospitals, clinics, home-care, and dialysis centers. It provides quiet operation for bedside use and may be used for acute care cases, small dialysis wings in hospitals, or to handle extra patients at dialysis centers. Note: Federal Law restricts this device to sale by or on the order of a physician for use as a water treatment device for hemodialysis.
Device Story
Portable, stand-alone water purification system for hemodialysis; processes potable tap water into dialysate-grade water. Input: potable water; output: purified water meeting ANSI/AAMI RD62 standards. Operation: temperature-blending valve (77°F), two backwashing granular activated carbon filters (iodine number >900), regenerating water softener (sulfonated polystyrene cation resin), 1 or 5-micron polyspun prefilter, and thin-film semi-permeable reverse osmosis membrane. User-operated via external manual control dials for backwashing and regeneration after each treatment. Used in hospitals, clinics, home-care, and dialysis centers for acute care or overflow capacity. Benefits: provides bedside, quiet water treatment for hemodialysis applications.
Clinical Evidence
Bench testing only. Device performance validated against ANSI/AAMI RD62 standards for water purification. Contaminant rejection rate ≥94%.
Indicated for patients requiring hemodialysis in hospitals, clinics, home-care, or dialysis centers; used to pretreat and purify potable water for dialysate production.
Regulatory Classification
Identification
A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.
Special Controls
*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
AmeriWater Purification System for Hemodialysis (K991519)
Submission Summary (Full Text)
{0}------------------------------------------------
APR 0 7 2003
January 2, 2003
Image /page/0/Picture/3 description: The image shows the logo for AmeriWater. The logo consists of the word "AmeriWater" in a bold, sans-serif font. A registered trademark symbol is to the right of the word. Below the word is a curved line made up of a series of black squares that decrease in size from left to right.
Dayton, Ohio 45404 1257 Stanley Avenue Phone: 800/535-5585 Phone: 9371461-8833 Fax: 937/461-1988 www.ameriwater.com
Subject: 510 (K) SUMMARY
510 (K) Number:
AmeriWater Contact: Brian R. Bowman, Quality Manager
| Proprietary Name: | Portable RO+ Model MROS09 |
|----------------------|--------------------------------------------|
| Common Name: | Water Purification System |
| Classification Name: | Water Purification System for Hemodialysis |
| Classification: | Class II Medical Device under §876.5665 |
## Device Description / Intended Use:
The AmeriWater Portable RO+ Model MROS09 is a portable, stand-alone water treatment system for use in hemodialysis applications. It is designed to pre-treat and purify potable water for use in making dialysate for hemodialysis. The AmeriWater Portable RO+ Model MROS09 is intended to be used in hospitals, clinics, home-care, and dialysis centers. It provides quiet operation for bedside use and may be used for acute care cases, small dialysis wings in hospitals, or to handle extra patients at dialysis centers. The RO+ is designed to produce water that meets or exceeds AN\$I/AAMI RD62 requirements.
The water treatment components for the AmeriWater Portable RO+ Model MROS09 consist of a temperature-blending valve, two backwashing carbon filters, regenerating water softener with brine tank, a micron prefilter, and a reverse osmosis membrane. The components are conveniently packaged in ABS cabinets designed to fit in a standard washing machine pan. All components are constructed of materials that meet FDA and/or NSF standards.
The temperature-blending valve provides a constant feed water temperature of 77° F, the ideal operating temperature for reverse osmosis. The blend valve assembly is sized to accommodate all anticipated ranges of flow rates. It includes a check valve to prevent backflow of water into the hot and cold water lines and a thermometer for use in monitoring the operation and effectiveness of the temperature-blending valve.
The backwashing carbon filters are designed to remove chlorine, chloramines, and organics from the potable tap water as a means of pre-treatment before the reverse osmosis process. Page 1 of 3
{1}------------------------------------------------
The filters are in a worker/polisher series with a test port between the two filters. The carbon media is granular activated carbon with an iodine number greater than 900. The backwashing carbon filtration has been designed to meet all standards for carbon filtration as defined by the Association for the Advancement of Medical Instrumentation (AAMI). The backwash control dials are accessible externally and appropriately labeled to provide simple operation for the user. Backwashing should be completed after each treatment.
The regenerating water softener is designed to remove hardness from the water feeding the reverse osmosis process. The softener incorporates a control valve with a manual timer for regeneration. The control dials are accessible externally and appropriately labeled to provide simple operation for the user. The softener resin is an ultra high capacity, gel type, sulfonated polystyrene cation resin that provides superior service runs and longer life. Regenerations should be completed after each treatment.
The micron pre-filter is encased in an opaque polypropylene housing. The filter cartridge is a plolyspun, dual open-end cartridge designed to filter sediment and carbon fines from the water feeding the reverse osmosis machine with greater than 80% efficiency. The user has the option of selecting a 1 micron or 5 micron filter cartridge.
The reverse osmosis membrane is a thin film, semi-permeable membrane that removes contaminants from the water by a separation process called reverse osmosis (RO). In this process the water is permitted to pass through the semi-permeable membrane, leaving behind the dissolved solids. This is accomplished by applying a pressure greater than the osmotic pressure difference to the concentrate water. which causes the direction of water passage through the membrane to reverse. In reverse osmosis the water travels from the more concentrate side of the membrane to the less concentrate side. Contaminants are removed from the water at a rejection rate greater than or equal to 94% to achieve product water that meets or exceeds ANSI/AAMI RD62 standards.
## Substantial Equivalence:
The AmeriWater Portable RO+ Model MROS09 is the same in function and design as the AmeriWater Portable RO+ currently approved for market under the 510(k) number K924695. The basic design, intended use, and performance specifications remain the same for the new model of the RO+ product line. The changes in design that make this model different from the existing models are:
- the addition of backwashing carbon filters in place of the carbon exchange tank; and ●
- the addition of a regenerating water softener in place of the existing antiscalant system.
{2}------------------------------------------------
The backwashing carbon filters and regenerating water softener are manufactured to the same design and specifications as the backwashing carbon filters and regenerating water softeners approved for market with the AmeriWater Purification System for Hemodialysis under 510(k) number K991519.
Both the AmeriWater Portable RO+ (K924695) and the AmeriWater Portable RO+ Model MROS09 incorporate carbon filtration, water softening, micron pre-filtration, and reverse osmosis in the design. The technological characteristics and performance specifications for the device do not change in the new model. Carbon filtration is accomplished by a single carbon exchange tank in the Portable RO+ (K924695) currently marketed. The carbon tank performance is monitored daily and the tank is replaced when exhausted or at least every six months. In the new model, MROS09, the carbon filtration consists of two backwashing carbon filters with manual timers. The carbon filters are manually set to backwash after each treatment. The design and performance specifications for the backwashing carbon filters are the same as those used for the carbon filters currently marketed with the AmeriWater Purification System for Hemodialysis under 510(k) number K991519.
The AmeriWater Portable RO+ (K924695) accomplishes water softening with an onboard antiscalant system designed to bind the hardness ions so that they can be removed without scaling the reverse osmosis membrane. Model MROS09 utilizes a regenerating water softener that removes hardness from the water by a process called ion exchange. The regenerating water softener is designed with a manual timer for regeneration and is manually set by the user to regenerate after each treatment. The design and performance specifications for the regenerating water softener are the same as those used for the water softeners currently marketed with the AmeriWater Purification System for Hemodialysis under 510(k) number K991519.
In summary, the new model in the AmeriWater Portable RO+ line (Model MROS09) combines the technology of the portable reverse osmosis machine that is currently marketed under 510(k) number K924695 with the technology of the regenerating water softener and backwashing carbons currently marketed with the AmeriWater Purification System for Hemodialysis under 510(k) number K991519. There are no new technological characteristics and no new issues of safety or effectiveness raised by this design.
Prepared By:
Brian R. Pearson 1/2/03
Brian R. Bowman, 1/2/03 Quality Manager
Page 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure, composed of three curved lines.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 0 7 2003
Mr. Brian R. Bowman Quality Manager Ameri Water 1257 Stanley Avenue DAYTON OH 45404
Re: K030059
Trade/Device Name: Portable RO+ Model MROS09 Regulation Number: 21 CFR §876.5665 Regulation Name: Water purification system for hemodialysis Regulatory Class: II Product Code: 78 FIP Dated: January 2, 2003 Received: January 7, 2003
Dear Mr. Bowman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the AmeriWater logo. The logo consists of the word "AmeriWater" in a serif font. Below the word is a curved line made up of black squares that decrease in size from left to right.
Dayton, Ohio 45404 1257 Stanley Avenue Phone: 800/535-5585 Phone: 9371461-8833 Fax: 9371461-1988 www.ameriwater.com
**510 (K) Number:** K030059
Device Name: Portable RO+ Model MROS09
## STATEMENT OF INTENDED USE
The AmeriWater Portable RO+ Model MROS09 is a stand-alone water treatment system for use in hemodialysis applications. It is designed to pretreat and purify potable water for use in making dialysate for hemodialysis. The AmeriWater Portable RO+ with Onboard Pretreatment System is intended to be used in hospitals, clinics, home-care, and dialysis centers. It provides quiet operation for bedside use and may be used for acute care cases, small dialysis wings in hospitals, or to handle extra patients at dialysis centers.
*Note: Federal Law restricts this device to sale by or on the order of a physician for use as a water treatment device for hemodialysis.
*Prescription Use*
David K. Updegrove
(Division Sign-Off) Division of Reproductive, Abdom and Radiological Devic 510(k) Number _
Page 1 of 1
510K, Statement of Intended Use, 11-06-02.doc brb
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.