← Product Code [KWZ](/productcode/KWZ) · K030047

# FREEDOM CONSTRAINED LINERS (K030047)

_Biomet, Inc. · KWZ · Jan 23, 2003 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K030047

## Device Facts

- **Applicant:** Biomet, Inc.
- **Product Code:** [KWZ](/productcode/KWZ.md)
- **Decision Date:** Jan 23, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3310
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Freedom™ Constrained Liners are intended to replace a hip joint. The devices are intended for general use in skeletally mature individuals undergoing primary and/or secondary revision surgery at high risk of hip dislocation due to a history of prior dislocation, joint or bone loss, soft laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered.

## Device Story

Freedom™ Constrained Acetabular Liners are polyethylene liners pre-assembled with a retaining ring; designed for use as a component of a total hip prosthesis. Device functions to provide stability in patients at high risk of dislocation. Used by orthopedic surgeons in clinical settings during primary or revision hip surgery. Output is a physical implant component that restricts joint motion to prevent dislocation. Benefits include reduced risk of recurrent dislocation in high-risk patients.

## Clinical Evidence

No clinical data provided; substantial equivalence based on mechanical testing and engineering analysis.

## Technological Characteristics

Polyethylene liner with pre-assembled retaining ring; metal/polymer constrained acetabular insert. Design features consistent locking mechanism and profile across five styles.

## Regulatory Identification

A hip joint metal/polymer constrained cemented or uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device prevents dislocation in more than one anatomic plane and has components that are linked together. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular component made of ultra-high-molecular-weight polyethylene with or without a metal shell, made of alloys, such as cobalt-chromium-molybdenum and titanium alloys. This generic type of device is intended for use with or without bone cement (§ 888.3027).

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Hip Joint Metal/Polymer Constrained Cemented or Uncemented Prosthesis.”

## Predicate Devices

- Ringloc® II Constrained Liners ([K021728](/device/K021728.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K030047

JAN 2 3 2003

## SUMMARY OF SAFETY AND EFFECTIVENESS

| Applicant or Sponsor: | Biomet Orthopedics, Inc.<br>P.O. Box 587<br>Warsaw, IN 46581-0587 |
|-----------------------|-------------------------------------------------------------------|
| Contact Person:       | Max T. Hebel<br>Telephone: (219) 267-6639                         |
| Proprietary Name:     | FreedomTM Constrained Liners                                      |
| Common Name:          | Constrained Acetabular Insert                                     |

Classification: Prosthesis, hip, constrained, metal/polymer (CFR 888.3310).

Device Classification: Class II

Legally Marketed Device to which Substantially Equivalence is Claimed: Ringloc® II Constrained Liners (K021728)

Device Description: The Freedom™ Constrained Acetabular Liners are polyethylene liners that will come pre-assembled with a retaining ring already in place. The liners will be available in five different styles. All the liners allow for the same range of motion. The locking mechanism and profile of that mechanism remain the same throughout the series.

Indications for Use: The Freedom™ Constrained Liners are indicated for use as a component of a total hip prosthesis in primary and revision patients at high risk of dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intra-operative instability, and for whom all other options to constrained acetabular components have been considered.

Summary of Technologies: The Freedom™ Constrained Liners-the materials, design, sizing, and indications are similar or identical to the predicate devices.

Non-Clinical Testing: Mechanical testing, published medical literature, and engineering justifications determined that the Freedom™ Constrained Liner presented no new unacceptable risks and is, therefore, substantially equivalent to the predicate device.

Clinical Testing: None provided as a basis for substantial equivalence.

## 1.007 1.5

MAILING ADDRESS SHIPPING ADDRESS 120. Box 587 56 E. Bell Drive Warsaw, IN 46581-0587 Warsaw, IN 46582 ースアスター(アステスティ) FAX - FAX
 『 FAX OFFICE 1:-MAIL (한 574.267.6639 biomet@biomet.com

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its wings. The eagle is enclosed in a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 2 3 2003

Mr. Max T. Hebel Regulatory Affairs Specialist Biomet, Inc. P.O. Box 587 Warsaw. Indiana 46581-0587

Re: K030047

Trade Name: Freedom™ Constrained Liner Regulation Number: 21 CFR 888.3310 Regulation Name: Hip joint metal/polymer constrained cemented or uncemented prosthesis Regulatory Class: II Product Code: KWZ Dated: January 2, 2003 Received: January 6, 2003

Dear Mr. Hebel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 – Mr. Max T. Hebel

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807,97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely vours,

Sincerely yours,

Mark A. Melkersen

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Page 01.

510(k) Number (if known): Device Name: Freedom™ Constrained Liners Indications for Use:

The Freedom™ Constrained Liners are intended to replace a hip joint. The devices are intended for general use in skeletally mature individuals undergoing primary and/or secondary revision surgery at high risk of hip dislocation due to a history of prior dislocation, joint or bone loss, soft laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered.

le Mark N Mellen

(Division Sign-Off)
Division of General. . . rative
and Neurological D. . . es

510(k) Number K030047

(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use (Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K030047](https://fda-staging.innolitics.com/device/K030047)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
