Pathway STM-10 is a non-implanted electrical continence device used to treat urinary incontinence and provide neuromuscular reeducation. The device connects to single-user vaginal or anal EMG/stim perineometer sensors. The patient inserts the sensor, and the device delivers electrical stimulation to the pelvic floor muscles to assist contraction, aiming to improve muscle strength and control. The device is operated by a physician or clinician who programs parameters such as frequency, duty cycle, and session length; patients do not program the device. Output is delivered via an asymmetrical balanced pulsed current. The device assists in muscle contraction, potentially benefiting patients by improving pelvic floor function.
Clinical Evidence
Bench testing only. Testing confirmed the device accurately applies programmed stimulation outputs and produces waveforms consistent with intended signals. Biocompatibility testing was performed on the vaginal and anal sensors, confirming safety.
Indicated for patients with stress, urge, or mixed urinary incontinence and for neuromuscular reeducation of the pelvic floor muscles.
Regulatory Classification
Identification
A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).
{0}------------------------------------------------
| | K030039 |
|--|-------------|
| | Page 1 of 4 |
**The Prometheus Group**
One Washington Street, #303 Dover, NH 03801
Phone: 603-749-0733 Fax: 603-749-0511
MAY 2 3 2003
# 510(k) SUMMARY Safety and Effectiveness Summary
# Pathway STM-10 Pelvic Floor Stimulator
### Submitted by:
Peter A Sullivan One Washington Street Suite 303 Dover, NH 03820 Phone: 603-749-0733 Fax: 603-749-0511
### Contact Person:
Peter Sullivan 781-582-2328
Date Submitted: April 7, 2002
{1}------------------------------------------------
K030039
Page 2 of 4
### NAME OF DEVICE
Pathway STM-10 Pelvic Floor Stimulator Trade name: Common name: Non-implanted Electrical Continence Device Classification name: 78 KPI, ClassII, (876.5320)
#### IDENTIFICATION OF PREDICATE DEVICE
The device to which we claim substantial equivalence is the Pathway CTS 2000 K023906 manufactured by The Prometheus Group.
### DESCRIPTION OF DEVICE
The Pathway STM-10 is used to stimulate the pelvic floor muscles for the treatment of incontinence. The Pathway Vaginal or Anal EMG/Stim perineometer sensor is connected to the Pathway STM-10 device to provide stimulation to the patient. This assists the patient with muscle contractions.
The Pathway STM-10 uses Pathway EMG/Stim Perineometer Sensors which are single-user sensors. The patient inserts the sensor into the vagina or rectum and uses the Pathway STM-10 to electrically stimulate the pelvic floor muscles to assist the contraction. The aim is to improve the strength and control of the pelvic floor muscles.
### INTENDED USE
Indications For Use:
- * Urinary Incontinence : Stress, Urge and Mixed Incontinence
- * Neuromuscular Reeducation
{2}------------------------------------------------
K'030039
Page 394
### SUMMARY OF TECHNICAL CHARACTERISTIC COMPARISON TO PREDICATE DEVICE
| Parameter | Pathway STM-10 | Pathway CTS 2000 |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Treatment of Urinary<br>Incontinence; Neuromuscular<br>Reeducation | Treatment of Urinary<br>Incontinence; Neuromuscular<br>Reeducation |
| Stimulator Output | 0-60 mA | 0-100 mA |
| Waveform | Asymmetrical Balanced Pulsed<br>Current | Asymmetrical Balanced Pulsed<br>Current |
| Charge/pulse at 500 ohms | 17uC | 28uC |
| Frequency | 12.5, 50 100, 200Hz | 12.5, 50, 100, 200 Hz |
| Peak pulse intensity | 60 mA | 100 mA |
| Pulse width | .3 ms fixed | .3 ms fixed |
| Ramps | 2 sec on ramp, one sec off ramp | 2 sec on ramp, one sec off ramp |
| Duty Cycle | Work/Rest of 5 seconds on, 5<br>seconds off (5/5), 5/10, 10/10,<br>10/20. | Work Time: 1 - 60 Seconds<br>Rest Time: 0 - 60 Seconds |
| Session Duration (min) | 5, 10, 15, 20, 25, or 30. | 0-60 |
| Programmable features | None by Patient; Frequency, Duty<br>cycle, Session length by physician | None by Patient; Frequency, Duty<br>cycle, Session length by physician |
| Vaginal EMG/Stim Probe Used | Pathway Vaginal<br>EMG/Stimulation Sensor<br>K993976 | Pathway Vaginal<br>EMG/Stimulation Sensor<br>K993976 |
| Anal EMG/Stim Probe Used | Pathway Anal EMG/Stimulation<br>Sensor K993976 | Pathway Anal EMG/Stimulation<br>Sensor K993976 |
| Vaginal EMG/Stim probe<br>electrode surface area: | 2.31 cm2 | 2.31 cm2 |
| Anal EMG/Stim probe electrode<br>surface area: | 2.12 cm2 | 2.12 cm2 |
| Current Density<br>(full output @ 500 ohms) | Pathway Vaginal EMG/Stim<br>Sensor: 26.0 mA/cm2<br>Pathway Anal EMG/Stim Sensor:<br>28.3 mA/cm2<br>(Max. Instantaneous) | Pathway Vaginal EMG/Stim<br>Sensor: 43.3 mA/cm2<br>Pathway Anal EMG/Stim Sensor:<br>47.2 mA/cm2<br>(Max. Instantaneous) |
| Power Density<br>(full output @ 500 ohms) | Pathway Vaginal EMG/Stim<br>Sensor: 5.6 mW/cm2<br>Pathway Anal EMG/Stim Sensor:<br>6.1 mW/cm2<br>(maximum intensity, .3ms pulse<br>width 200Hz) | Pathway Vaginal EMG/Stim<br>Sensor: 15.6 mW/cm2<br>Pathway Anal EMG/Stim Sensor:<br>17.0 mW/cm2<br>(maximum intensity, .3ms pulse<br>width 200Hz) |
{3}------------------------------------------------
KO30039
Page 4 of 4
#### BENCH TEST DATA
A series of bench tests were performed using the Pathway STM-10 to show the device accurately applies stimulation and is substantially equivalent to the predicate device. The Pathway STM-10 was used to apply programmed stimulation outputs and the resulting waveforms were measured and compared to the intended signals.
The bench tests show the Pathway STM-10 accurately applies muscle stimulation.
### BIOCOMPATIBILITY TESTING
The Pathway Vaginal EMG/Stimulation Perineometer Sensor and the Pathway Anal EMG/Stimulation Perineometer Sensor have been laboratory tested for the safety of the materials. The Pathway Perineometer Sensors were found to be safe under the standards required for each test.
### CONCLUSION
The Pathway STM-10 is safe and effective for its intended use. The Pathway STM-10 is substantially equivalent to the predicate device.
# END OF 510(k) SUMMARY
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol of three human profiles facing right, with flowing lines representing movement or connection.
{
"markdown": ""
}
Food and Drug Administration 9200 Corporate Boulevald Rockville MD 20850
Mr. Peter A. Sullivan Engineer The Prometheus Group 2 Mallards Cove
DUXBURY MA 02332
Re: K030039
MAY 2 3 2003
Trade/Device Name: Pathway STM-10 Pelvic Floor Stimulator Regulation Number: 21 CFR 876.5320 Regulation Name: Nonimplanted electrical continence device Regulatory Class: II Product Code: 78 KPI Dated: April 7, 2003 Received: April 9, 2003 .
Dear Mr. Sullivan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean ' that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
.
{6}------------------------------------------------
KO30039
### SECTION 2 - STATEMENT OF INDICATIONS FOR USE
### 2.1 STATEMENT OF INDICATIONS FOR USE
510(k) Number:_K030039 Device Name:__Pathway STM-10 pelvic floor stimulator
### Indications for Use:
Indications For Use
- * Urinary Incontinence : Stress, Urge and Mixed Incontinence
- * Neuromuscular Reeducation
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use:<br>(Per 21 CFR 801.109) | X | OR | Over-the-Counter Use: | |
|-------------------------------------------|---|----|-----------------------|--|
|-------------------------------------------|---|----|-----------------------|--|
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
| 510(k) Number | X030039 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.