ATAC TOTAL BILIRUBIN REAGENT

K030014 · Elan Diagnostics · CIG · Sep 26, 2003 · Clinical Chemistry

Device Facts

Record IDK030014
Device NameATAC TOTAL BILIRUBIN REAGENT
ApplicantElan Diagnostics
Product CodeCIG · Clinical Chemistry
Decision DateSep 26, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1110
Device ClassClass 2

Indications for Use

The ATAC Total Bilirubin Reagent Kit is intended for use with the ATAC Calibrator and the ATAC 8000 Random Access Chemistry System as a system for the quantitative determination of total bilirubin in serum and plasma. Total bilirubin results are used for the diagnosis and treatment of liver, hematological, and metabolic disorders, including hepatitis and gall bladder block. This reagent is intended to be used by trained personnel in a professional setting and is not intended for home use.

Device Story

Reagent kit for quantitative total bilirubin measurement in serum/plasma; used with ATAC 8000 Random Access Chemistry System. Principle: reaction with diazotized sulfanilic acid in presence of dimethylsulfoxide forms red-purple complex; absorbance measured at 546 nm. Absorbance increase proportional to bilirubin concentration. Operated by trained personnel in professional clinical laboratory settings. Output provides bilirubin concentration values to clinicians for diagnosing/treating liver, hematological, and metabolic disorders.

Clinical Evidence

Bench testing only. Precision evaluated over 7 days (n=36 per sample); total CV ranged from 2.7% to 16.3%. Linearity established up to 25 mg/dL. Interference testing performed for heparin, hemoglobin, and triglycerides. Method comparison study (n=120) against predicate device yielded Deming regression: y = 1.013x - 0.05 mg/dL (Sy.x = 0.31 mg/dL).

Technological Characteristics

Two-reagent kit: Reagent 1 (sulfanilic acid, hydrochloric acid, 50% DMSO) and Reagent 2 (sodium nitrite). Quantitative diazo colorimetric assay. Measurement at 546 nm. Requires ATAC® 8000 Random Access Chemistry System. Calibrator traceable to NIST reference materials.

Indications for Use

Indicated for quantitative determination of total bilirubin in serum and plasma for the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block. For prescription use.

Regulatory Classification

Identification

A bilirubin (total or direct) test system is a device intended to measure the levels of bilirubin (total or direct) in plasma or serum. Measurements of the levels of bilirubin, an organic compound formed during the normal and abnormal distruction of red blood cells, if used in the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block.

Predicate Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY DEVICE ONLY TEMPLATE A. 510(k) Number: K030014 B. Analyte: Total Bilirubin C. Type of Test: Quantitative Diazo Colorimetric test D. Applicant: Clinical Data E. Proprietary and Established Names: ATAC® Total Bilirubin Reagent Kit F. Regulatory Information: 1. Regulation section: 21 CFR § 862.1110 2. Classification: Class II 3. Product Code: CIG 4. Panel: Chemistry (75) G. Intended Use: 1. Intended use(s): ATAC® Total Bilirubin Reagent is for the quantitative determination of total bilirubin in serum and plasma using the ATAC® 8000 Random Access Chemistry System. 2. Indication(s) for use: ATAC® Total Bilirubin Reagent Kit is intended for use with the ATAC® Calibrator and the ATAC® 8000 Random Access Chemistry System as a system for the quantitative determination of total bilirubin in serum and plasma. Total bilirubin results are used for the diagnosis and treatment of liver, hemolytic hematological, and metabolic disorders, including hepatitis and gall bladder block 3. Special condition for use statement(s): For prescription use 4. Special instrument Requirements: ATAC® 8000 Random Access Chemistry System {1} Page 2 of 4 H. Device Description: The ATAC® Total Bilirubin Reagent Kit is a two reagent assay to be used in the ATAC® 8000 Random Access Chemistry System. It is composed of a Total Bilirubin reagent which contains sulfanilic acid and hydrochloric acid in 50% dimethylsulfoxide and a Nitrite activator containing sodium nitrite. The kit includes the ATAC® Calibrator (K030621). I. Substantial Equivalence Information: 1. Predicate device name(s): Beckman Synchron Total Bilirubin Reagent 2. Predicate K number(s): K902801 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended Use | Detection of Total Bilirubin | Detection of Total Bilirubin | | Matrix | Human serum and plasma | Human serum and plasma | | Differences | | | | Item | Device | Predicate | | Methodology | Acid Diazo Method | Time end-point Diazo Method | | Chemistry | ATAC® 8000 Random Access | Beckman Synchron | | Analyzers | Chemistry System | | J. Standard/Guidance Document Referenced (if applicable): Not Applicable K. Test Principle: When the reagent is reconstituted, the sulfanilic acid reacts with sodium nitrite to form a diazotized sulfanilic acid (diazo). In the presence of dimethylsulfoxide (DMSO) conjugated and free bilirubin in the sample react with this diazo form to form a red-purple complex. Measurement of absorbance at 546 nm is proportional to the total bilirubin concentration in the sample. L. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Total and within run imprecision were calculated from three serum controls each assayed in triplicate 12 times over a 7 day period. Eight of the 12 runs were calibrated concurrently. {2} Page 3 of 4 | Sample | n | Mean | Within-run | | Total | | | --- | --- | --- | --- | --- | --- | --- | | | | | SD | %CV | SD | %CV | | Serum 1 | 36 | 0.6 | 0.04 | 6.5 | 0.10 | 16.3 | | Serum 2 | 36 | 3.4 | 0.06 | 1.6 | 0.11 | 3.4 | | Serum 3 | 36 | 6.3 | 0.12 | 1.9 | 0.17 | 2.7 | b. Linearity/assay reportable range: Linearity was assessed between the level of detection and 25 mg/dL in three instruments using seven bilirubin linearity standards with concentrations distributed evenly over the entire range. c. Traceability (controls, calibrators, or method): The calibrator is traceable to NIST reference materials. d. Detection limit: Total imprecision was assessed at 0.6 mg/dL to have a %CV of 16.3. e. Analytical specificity: Effects of interference from additives such as sodium heparin, ammonium heparin, and lithium heparin and from hemoglobin and triglycerides were assessed. Additives were added to a serum pool spiked to 4 mg/dL of total bilirubin at concentration of 40 U/mL. Comparison of results from the bilirubin spiked pool with and without additive showed statistically insignificant biases. Hemoglobin at levels of 40, 80, 120, 160, and 200 mg/dL and triglycerides at 400, 800, 1200, 1600 and 2000 mg/dL were assessed for bias as above. The largest biased observed was for hemoglobin at 160 mg/dL which produced a positive bias of approximately 0.2 mg/dL. f. Assay cut-off: The literature derived reference range for normal values for serum and plasma is 0.3 to 1.2 mg/dL. 2. Comparison studies: a. Method comparison with predicate device: Sixty serum and 60 heparinized plasma specimens from individual adult patients were selected to contain both normal and abnormal levels of total bilirubin were analyzed with the device and it's predicate. Deming regression analysis gave the following fit ATAC 8000 = 1.013 * (Predicate) – 0.05 mg/dL with Sy.x = 0.31 mg/dL {3} Page 4 of 4 b. Matrix comparison: See above 3. Clinical studies: a. Clinical sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a and b are not applicable): See comparison study above. 4. Clinical cut-off: No clinical cut-off was determined. 5. Expected values/Reference range: Literature was reference for the normal expected values, which are below the device’s detection limit. M. Conclusion: Based upon the information provided, I recommend that the ATAC® Total Bilirubin Reagent Kit be found substantially equivalent to predicate devices according to 21 CFR § 862.1110 and 21 CFR § 862.3645.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...