← Product Code [MNR](/productcode/MNR) · K024322

# EMBLA N7000 (K024322)

_Medcare Flaga · MNR · Oct 27, 2003 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K024322

## Device Facts

- **Applicant:** Medcare Flaga
- **Product Code:** [MNR](/productcode/MNR.md)
- **Decision Date:** Oct 27, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.2375
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology

## Indications for Use

The Embla N7000 is intended for use by a physician or trained technician for the acquisition of EEG and polysomnography (PSG) signals and transmission of these signals to a PC during neurophysiologic or sleep examinations. The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments, The use of the Embla N7000 system does not involve any patient monitoring or diagnosis.

## Device Story

Embla N7000 is a full polysomnography and EEG acquisition system. It collects EEG, bipolar, and respiratory signals, plus auxiliary inputs (e.g., CPAP, CO2 machines). Device amplifies and digitizes signals, transmitting them via Ethernet to a PC for external analysis. Used in hospitals, sleep labs, and clinics by physicians or technicians. System does not perform patient monitoring or diagnosis; it serves as a data acquisition front-end. Benefits include streamlined cabling and flexible, rugged design for clinical/research environments. Healthcare providers use the transmitted data on a PC for subsequent review and clinical decision-making.

## Clinical Evidence

Bench testing only. Validation included design review, internal verification, and practical testing in a simulated sleep lab setting. Hazard analysis performed and verified. External testing by SEMKO confirmed compliance with EMC requirements and electrical safety standards (UL2601).

## Technological Characteristics

System features 32 referential EEG channels, 8 bipolar channels, and respiratory signal inputs. Uses internal microprocessor and Flash memory. Connectivity via standard Ethernet with signal-transformer isolation. Powered by internal mains isolation transformer. Complies with EN60601-1, UL-2601, and EN60601-1-2. Software level of concern is minor.

## Regulatory Identification

A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.

## Predicate Devices

- Artisan™ Acquisition System ([K992283](/device/K992283.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the word "Medcare" in bold black font. Above the word, there is a handwritten string of characters, "K024322". The handwritten string is oriented diagonally in the upper left corner of the image.

OCT 2 7 2003

### 510(k) Summary

December 19th, 2002

#### Submitter

Medcare Flaga Vesturhlid 7 105 Reykjavik Iceland Europe

Tel: 011 354 510 2000 Fax: 011 354 510 2010

Contact Person: Berglind Hallgrímsdóttir

e-mail: berglind@medcare.is, guality@medcare.is

#### Device

Trade Name: Common Name: Classification Product Code Classification Panel Embla N7000 EEG Amplifier Electroencephalograph, Class II GWQ Neurology

#### Substantial Equivalence

Artisan™ Acquisition System For EEG and polysmnography.

Airsep Corporation

290 Creekside Drive Buffalo, NY 14228

#### 510(k) Number K992283

#### Description of the Device

The Embla N7000 system is a full polysomnography system with expanded EEG capability. It is used to perform online sleep studies in the sleep lab, hospital or clinical environment under the supervision of a clinician or sleep technician. It does not do the recording, monitoring or analysis of this data.

The Embla N7000 system integrates digital technology and precision engineering into a flexible, rugged, full polysomnography system. The system itself, featuring an Ethernet network connection, is simple to assemble and the cables have been streamlined to provide a comfortable and reliable system. Meeting the most demanding clinical and research needs by offering 32 referential EEG channels, 8 bipolar channels plus an extensive set of respiratory signals, the system also includes auxiliary inputs for additional devices such as CPAP and CO2 machines.

{1}------------------------------------------------

# Medcare

#### Intended Use of the device

The Embla N7000 is intended for use by a physician or trained technician for the acquisition of EEG and polysomnography (PSG) signals and transmission of these signals to a PC during neurophysiologic or sleep examinations. The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments,

The use of the Embla N7000 system does not involve any patient monitoring or diagnosis.

#### Technological Charaterics

Both Artisan and Embla N7000 are used for amplifying EEG and sleep signals in clinical and research laboratories. Both are designed for collection and transmission of sleep data to a personal computer. Both systems are only acquisition systems from amplifying and digitizing EEG and other polygraph channels and their transmission to a PC.

The amplification and data acquisition functions are similar for both systems. The main difference is the extended hardware possibilities of re-referencing in the Embla N7000. Also the Embla N7000 is controlled by software that is stored in internal Flash memory and runs on internal microprocessors. Artisans programs are also stored in internal Flash memory and run on a PLD (Programmable Logic Device). The difference between a microprocessor and a PLD is mainly how the commands are processed in each device. The software for PLDs sets up the configuration of the PLDs electronics to make it perform certain logical operations. The software for microprocessors performs the same logical operations by a sequence of instructions executed in the processors CPU. The software development was done according to the firmware development process at Medcare Flaga and it was verified that the software requirements are met. The software level of concern for both products was minor.

The patient mains-isolation is equivalent for the two devices. The N7000 has internal mains isolation transformer to power the device. The Artisan is powered from external mains transformer that provides the same isolation. In both cases the mains isolation fulfills the reinforced requirements of over 4000V manufacturing test voltage, set by the EN60601-1 and UL-2601.

The isolation from external devices is done in similar way in both devices. In Artisan, a special external box containing AD converters and opto-couplers are used to isolate the signals, but in N7000 the similar converters and opto-couplers are built in the device.

The construction of the communication is also equivalent. Artisan uses fiber-optic cable for the isolation, using a proprietary communication standard for the point to point communications with the PC. The N7000 uses a signal-transformer to isolate the communications and communicates over standard Ethernet with the PC.

#### Test

The Embla N7000 has been tested and verified in various phases, the testing included design review, internal verification and validation as well as external testing/validation. The design was verified throughout the design process. Internal validation included bench testing and practical testing in a simulated sleep lab setting. Hazard analysis was done and appropriate measures were implemented and their effectiveness was verified. External test house, SEMKO, was used to confirm compliance to EMC requirements and standards for electrical safety including UL2601.

Testing performed and the results verify that the Embla N7000 is safe and effective for its intended use.

{2}------------------------------------------------

# Medcare

#### Compliance with guidelines and standards

Embla N/S7000 complies with the FDA "Electroencephalograph Devices Guidance for 510(k) Content, Draft Document, Version 1.0"

Embla N/S7000 is certified to carry the CE mark (CE 0413). The CE mark is a declaration that Embla N7000 is in compliance with the directive set forth by the European Union for medical devices.

Embla N/S7000 is manufactured by Medcare Flaga and the system conforms to the following standards and requlations:

Class IIa, Medical Device Directive (MDD), Annex V, "EC Directive" 93/42/EEC

EN60601-1:1990 with A1 and A12:1993, A2:1995 and A13:1996. (IEC 60601-1, 2nd ed., 1988 with A1,1991 and A2, 1995) : MEDICAL ELECTRICAL EQUIPMENT -PART 1: GENERAL REQUIREMENTS FOR SAFETY

EN60601-1-2:2001 MEDICAL ELECTRICAL EQUIPMENT -PART 1: GENERAL REQUIREMENTS FOR SAFETY - 2. COLLATERAL STANDARD: ELECTROMAGNETIC COMPATIBILITY -REQUIREMENTS AND TESTS

EN 61000-3-2:1995 and A14:2000. ELECTROMAGNETIC COMPATIBILITY (EMC) - PART 3-2: LIMITS - LIMITS FOR HARMONIC CURRENTS EMISSIONS (EQUIPMENT INPUT CURRENT UP TO AND INCLUDING16 A PER PHASE)

EN 61000-3-3:1995, ELECTROMAGNETIC COMPATIBILITY (EMC) - PART 3-3: LIMITS -LIMITATION OF VOLTAGE FLUCTUATIONS AND FLICKER IN LOW-VOLTAGE SUPPLY SYSTEMS FOR EQUIPMENT WITH RATED VOLTAGE UP TO 16 A

EN865:1997 PULSE OXIMETERS - PARTICULAR REQUIREMENTS

IEC 60601-2-25:1993 PARTICULAR REQUIREMENTS FOR THE SAFETY OF ELECTROCARDIOGRAPHS

60601-2-26:1994 PARTICULAR FOR IEC REQUIREMENTS THE SAFETY OF ELECTROENCEPHALOGRAPHS

IEC-60601-2-40:1998 PARTICULAR REQUIREMENTS FOR THE SAFETY OF ELECTROMYOGRAPHS AND EVOKED RESPONSE EQUIPMENT

EN60601-1-4 MEDICAL ELECTRICAL EQUIPMENT - PART 1: GENERAL REQUIREMENTS FOR SAFETY - 4. COLLATERAL STANDARD: PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS

UL 2601-1: MEDICAL ELECTRICAL EQUIPMENT, PART 1:GENERAL REQUIREMENTS FOR SAFETY, 2nd ed. 1997

CAN/CSA C22.2 No.601.1-M90 MEDICAL ELECTRICAL EQUIPMENT PART 1: GENERAL REQUIREMENTS FOR SAFETY

AS/NZS 3200.1.0: 1998 MEDICAL ELECTRICAL EQUIPMENT - GENERAL REQUIREMENTS FOR SAFETY - PARENT STANDARD

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, represented by three curved lines that suggest the wings and body of the bird.

Public Health Service

OCT 27 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Kumar Kulkarni Medcare Flaga c/o RTS, Incorporated 46 Loch Haven Lane Battle Creek, MI 49015

Re: K024322

Trade/Device Name: Embla N7000 Regulation Number: 21 CFR 868.2375 Regulation Name: Ventilatory Effort Recorder Regulatory Class: II Product Code: MNR Dated: July 27, 2003 Received: July 31, 2003

Dear Mr. Kulkarni:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

### Page 2 - Mr. Kumar Kulkarni

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Joflis
Chiu Lin, Ph.D.

fer

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

(Optional Format 3-10-98)

Applicant: Medare Flaga

510(k) Number (if known): K024322

Device Name: Embla N7000

Indications For Use:

The Embla N7000 is intended for use by a physician or trained technician for the acquisition of EEG and polysomnography (PSG) signals and transmission of these signals to a PC during neurophysiologic or sleep examinations. The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments.

The use of the Embla N7000 system does not involve any patient monitoring or diagnosis.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Q. Patricku 10/24/03

(Division Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices

510(k) Numb

5

---

**Source:** [https://fda-staging.innolitics.com/device/K024322](https://fda-staging.innolitics.com/device/K024322)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
