MODIFICATION TO POSITIVE TOUCH POWDER-FREE LATEX EXAMINATION GLOVES WITH PROTEIN AND CHEMOTHERAPY LABELING CLAIM

K024292 · Allegiance Healthcare Corp. · LYY · Jan 9, 2003 · General Hospital

Device Facts

Record IDK024292
Device NameMODIFICATION TO POSITIVE TOUCH POWDER-FREE LATEX EXAMINATION GLOVES WITH PROTEIN AND CHEMOTHERAPY LABELING CLAIM
ApplicantAllegiance Healthcare Corp.
Product CodeLYY · General Hospital
Decision DateJan 9, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

Positive Touch Powder-Free Latex Examination Gloves With Protein and Chemotherapy Labeling Claim are disposable devices intended for medical purposes that are worn on the examiner's hands or fingers to prevent contamination between patient and examiner. These gloves have been tested for use with chemotherapy drugs using ASTM F 739-99a.

Device Story

Disposable patient examination glove; made of natural rubber latex with polymer coating. Used by healthcare professionals in clinical settings to prevent cross-contamination between patient and examiner. Tested for resistance to chemotherapy drugs per ASTM F 739-99a. Provides barrier protection; reduces potential for latex protein sensitization due to low protein content (≤50 µg/g).

Clinical Evidence

Bench testing only. Primary skin irritation and guinea pig maximization tests showed no irritation or sensitization potential. Physical properties (tensile strength, elongation) met ASTM D3578-01. Barrier defects met 21 CFR §800.20 and ASTM D3578-01 (AQL=2.5). Protein content verified at ≤50 µg/g via ASTM 5712-95. Chemotherapy drug resistance tested per ASTM F 739-99a.

Technological Characteristics

Material: Natural rubber latex with polymer coating. Form factor: Non-sterile, various sizes (XS-XL). Standards: ASTM D3578-01 (physical properties, powder levels), ASTM 5712-95 (protein content), ASTM F 739-99a (chemotherapy drug permeation).

Indications for Use

Indicated for use as a disposable patient examination glove worn on the examiner's hands or fingers to prevent cross-contamination between patient and examiner. Contains 50 micrograms or less of total water extractable protein per gram. Tested for use with chemotherapy drugs.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Ko24292 Image /page/0/Picture/2 description: The image shows the word "Allegiance" in a bold, stylized font. To the left of the word is a symbol resembling a plus sign, but it is made up of many small squares. The word "Allegiance" is written in all lowercase letters, and the letters are connected to each other. #### STATEMENT OF SAFETY AND EFFECTIVENESS XIII. #### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS POSITIVE TOUCH POWDER-FREE LATEX EXAMINATION GLOVES WITH PROTEIN & CHEMOTHERAPY LABELING CLAIM | Manufacturer: | Allegiance Healthcare Corporation<br>1500 Waukegan Road<br>McGaw Park, Illinois USA | |-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Regulatory Affairs Contact: | Erica Sethi<br>1500 Waukegan Road, Bldg. WM<br>McGaw Park, IL 60085 | | Telephone: | (847) 785-3337 | | Date Summary Prepared: | 11/20/02 | | Common Name: | Patient Examination Glove | | Classification: | Glove, Examination (Latex) | | Predicate Devices: | Positive Touch Powder-Free Latex Examination Gloves With Protein<br>and Chemotherapy Labeling Claim | | Description: | Positive Touch Powder Free Examination Gloves With Protein and<br>Chemotherapy Labeling Claim are formulated using latex and are coated<br>with a Polymer coating. These gloves are offered non-sterile in sizes X-<br>small, Small, Medium, Large and X-large. | | Intended Use: | Positive Touch Powder-Free Latex Examination Gloves With Protein and<br>Chemotherapy Labeling Claim are disposable devices intended for<br>medical purposes that are worn on the examiner's hands or fingers to<br>prevent contamination between patient and examiner. These gloves have<br>been tested for use with chemotherapy drugs using ASTM F 739-99a. | {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the word "Allegiance" in a bold, stylized font. To the left of the word is a symbol resembling a plus sign, but it is made up of many small dots. The text is in black and the background is white. #### STATEMENT OF SAFETY AND EFFECTIVENESS XIII. ### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (con't) POSITIVE TOUCH POWDER-FREE LATEX EXAMINATION GLOVES WITH PROTEIN AND CHEMOTHERAPY LABELING CLAIM | Substantial Equivalence | Positive Touch Powder Free Examination Gloves With Protein and<br>Chemotherapy Labeling Claim are substantially equivalent to Positive<br>Touch Powder-Free Latex Examination Gloves in that they provide the<br>following characteristics: | |-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | - intended use | | | - size, design, product features | | | - made of natural rubber latex | - physical characteristics Summary of Test Results: | Test | Result | |-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Primary Skin Irritation | Glove does not display irritation potential. | | Guinea Pig Maximization | Glove does not display sensitization potential. | | Ultimate Elongation<br>& Tensile Strength | Glove meets or exceeds requirements for rubber<br>examination gloves per ASTM D3578-01. | | Barrier Defects | Glove meets or exceeds requirements per 21 CFR<br>§800.20 and ASTM D3578-01, AQL=2.5. | | Powder Level | Glove meets powder level requirements for "Powder Free"<br>designation per ASTM D 3578-01. | | Protein Labeling Claim | Glove meets requirements for protein claim of 50 microgram or less<br>of total water extractable protein per gram of glove using the ASTM<br>Lowry test method (ASTM 5712-95). | {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's symbol, which is a stylized representation of a human figure. The symbol is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter of the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## JAN 0 9 2003 Ms. Erica Sethi Manager, Regulatory Affairs Allegiance Healthcare Corporation 1500 Waukegan Road, Building WM McGaw Park, Illinois 60085 Re: K024292 Trade/Device Name: Positive Touch Powder-Free Latex Examination Gloves Contains 50 Micrograms or Less of Total Extractable Protein Per Gram and Chemotherapy Claim Regulation Number: 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LYY Dated: November 20, 2002 Received: December 24, 2002 Dear Ms. Sethi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Ms. Sethi Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely vour Timothy Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ K024 292 # Allegiance Allegiance Healthcare Corporation 1500 Waukegan Road McGaw Park, Illinois 60085-6787 847.473.1500 FAX: 847.785.2460 Page 1 of 1 | Applicant: | Allegiance Healthcare Corporation | |----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number: | K024292 | | Device Name: | Positive Touch Powder-Free Latex Examination Gloves With Protein,<br>Chemotherapy Labeling Claim | | Indications For Use: | A patient examination glove is a disposable device intended for medical<br>purposes that is worn on the examiner's hand or finger to prevent contamination<br>between patient and examiner. This latex glove contains 50 micrograms or less<br>of total water extractable protein per gram. In addition, this glove has been<br>tested for use with chemotherapy drugs. | (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use Over-The Counter Use or (Per 21 CFR 801.109) (Division Sign-Off) Hospital. Division of Anesthesiolog Infection Control, D
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...