← Product Code [GCJ](/productcode/GCJ) · K024251

# SIGHTFIX (K024251)

_World of Medicine Lemke GmbH · GCJ · Mar 18, 2003 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K024251

## Device Facts

- **Applicant:** World of Medicine Lemke GmbH
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Mar 18, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The SightFix is intended to allow access and observation of body cavities and the surgical field during diagnostic and therapeutic procedures in laparoscopy and open surgery.

## Device Story

SightFix comprises a rigid endoscope and a manually operated flexible holding arm. Endoscope contains optics and light fiber; connects to external camera (SightCam) and light source (halogen, xenon, or quartz). Holding arm provides stable positioning of endoscope during diagnostic and therapeutic procedures in laparoscopy and open surgery. Operated by surgical staff in OR setting. Provides visual access to surgical field; assists in maintaining stable view without manual holding. Benefits include improved visualization and reduced fatigue for surgical team.

## Clinical Evidence

Bench testing only. Compliance with IEC 60601-2-18:1996 (safety of endoscopic equipment) and European sterilization standards (EN 554:1994, EN 556:2001-01, EN 1174-1:1996).

## Technological Characteristics

Rigid endoscope with optic and light fiber; manually operated flexible holding arm. Materials: stainless steel and aluminum. Sterilization: autoclavable. Connectivity: compatible with standard 180W endoscopic light sources via adapters. Standards: IEC 60601-2-18:1996, EN 554, EN 556, EN 1174-1.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- MGB LAPALUX Telescope ([K982013](/device/K982013.md))
- Robotrac ™ Retractor Arm (Unitrac) ([K893121](/device/K893121.md))
- KSEA Endoscope Holder ([K990334](/device/K990334.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K024251

### 510(K) SUMMARY SightFix

MAR 1 8 2003

#### I. Submitter:

WORLD OF MEDICINE Lemke GmbH Danziger Strasse 21 82194 Gröbenzell Germany

#### II. Device Names:

- 1 . Classification Name: Endoscope and Accessory
- 2. Common or Usual Name: Laparoscope and Endoscope Holder
- 3. Proprietary Name:

SightFix

#### III. Classification:

Class II. The device is described in 21 C.F.R. § 876.1500. The product code for the device is GCJ.

#### IV. Predicate Device:

- MGB LAPALUX Telescope (K982013) manufactured by MGB Endoskopische . Geräte GmbH
- Robotrac ™ Retractor Arm (Unitrac) (K893121) manufactured by Aesculap Instruments Corp.
- KSEA Endoscope Holder (K990334) manufactured by Karl Storz Endoscopy-● America, Inc.

#### V. Intended Use:

The SightFix is intended to allow access and observation of body cavities and the surgical field during diagnostic and therapeutic procedures in laparoscopy and open surgery.

#### VI. Device Description:

The SightFix consists of a rigid endoscope and a flexible holding arm. The rigid endoscope of the SightFix contains an optic and a light fiber. The SightFix endoscope is designed to be used with the camera SightCam and the light source SightLight manufactured by WORLD OF MEDICINE Lemke GmbH. In addition, the use of adapters allows an easy attachment of standard 180W endoscopic light sources (halogen, xenon, quartz) to the endoscope. The holding arm of the SightFix is a manually operated endoscope holder to ensure an optimal positioning of the endoscope. Both the rigid endoscope and some components of the holding arm are autoclavable.

{1}------------------------------------------------

#### VII. Substantial Equivalence:

The SightFix described in this notification is similar in intended use, design, material and technological characteristics to the MGB LAPALUX Telescope (K982013) manufactured by MGB Endoskopische Geräte GmbH, the Robotrac TM Retractor Arm (Unitrac) (K893121) manufactured by Aesculap Instruments Corp. and the KSEA Endoscope Holder (K990334) manufactured by Karl Storz Endoscopy-America, Inc.

Both the endoscope of the SightFix and the predicate device MGB LAPALUX Telescope (K982013) are intended to allow access and observation of body cavities during diagnostic and therapeutic procedures in laparoscopy. In addition, both devices are rigid and autoclavable.

The holding arm of the SightFix and the predicate devices Robotrac ™ Retractor Arm (Unitrac) (K893121) and KSEA Endoscope Holder (K990334) are devices all intended to hold an endoscope during diagnostic and therapeutic endoscopic procedures. Moreover, the SightFix holding arm and the Robotrac TM Retractor Arm (Unitrac) (K893121) are both intended to be used during open surgery. All three devices contain stainless steel and aluminum and can be sterilized.

The differences between the SightFix and the predicate devices are minor and raise no new questions of safety and effectiveness. Accordingly, WORLD OF MEDICINE Lemke GmbH believes that the SightFix is substantially equivalent to the predicate devices currently on the market.

## VIII. Performance Data:

The device complies with the International standards IEC 60601-2-18:1996 (Particular requirements for the safety of endoscopic equipment). Some Components of the SightFix comply with the European Standards EN 554:1994, EN 556:2001-01 and EN 1174-1:1996 (Sterilization of Medical Devices). In addition, the device bears the CE mark in accordance with the European Medical Device Directive 93/42/EEC.

{2}------------------------------------------------

Image /page/2/Picture/12 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure with three overlapping profiles, suggesting a sense of community or interconnectedness. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the figure.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# MAR 1 8 2003

World of Medicine Lemke GmbH c/o Ms. Susanne Raab 91 Trowbridge Street Cambridge, Massachusetts 02138

Re: K024251

Trade/Device Name: SightFix Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: December 13, 2002 Received: December 23, 2002

Dear Ms. Raab:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Susanne Raab

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

for Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## STATEMENT OF INDICATIONS FOR USE

.

| APPLICANT:                | WORLD OF MEDICINE Lemke GmbH |
|---------------------------|------------------------------|
| 510(K) NUMBER (if known): | K024251                      |
| DEVICE NAME:              | SightFix                     |

INDICATIONS FOR USE:

The SightFix is intended to allow access and observation of body cavities and the surgical field during diagnostic and therapeutic procedures in laparoscopy and open surgery.

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Per 21 C.F.R. § 801.109)

(Optional Format 1-2-96)

Miriam C. Provost

(Division Sign-Off) Division of General, Restorative and Neurologicas Devices

510(k) Number K024251

---

**Source:** [https://fda-staging.innolitics.com/device/K024251](https://fda-staging.innolitics.com/device/K024251)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
