← Product Code [GEX](/productcode/GEX) · K024189

# NLITE SYSTEM, MODEL#: A00-1025/A00-1033 (K024189)

_Icn Photonics , Ltd. · GEX · Jun 20, 2003 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K024189

## Device Facts

- **Applicant:** Icn Photonics , Ltd.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Jun 20, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The NLite System is indicated for use in Dermatological and Plastic Surgery applications and this device is intended for use in the treatment of Inflammatory Acne Vulgaris.

## Device Story

The NLite System is a flashlamp-pumped, pulsed dye laser used in dermatology and plastic surgery. The system comprises a main console containing the laser resonator, external optics, high-voltage delivery system, internal cooler, fluid circulation system, control unit, and user interface. It utilizes a flexible fiber optic delivery device with an optical handpiece and a footswitch for pulsing control. The device is operated by a physician or trained clinician in a clinical setting. The laser energy is delivered to the target tissue to treat inflammatory acne vulgaris. The system is designed to be safe and effective for its intended use, with design and manufacture adhering to international standards to minimize risks to the operator and patient.

## Clinical Evidence

Clinical data provided to demonstrate safety and efficacy for the treatment of Inflammatory Acne Vulgaris. No specific study metrics (e.g., p-values, CIs) are detailed in the provided documentation.

## Technological Characteristics

Flashlamp-pumped, pulsed dye laser. System includes laser resonator, external optics, high-voltage delivery system, internal cooler, fluid circulation system, control unit, and user interface. Delivery via flexible fiber optic and optical handpiece. Footswitch-controlled. Class II device (21 CFR 878.4810).

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- ICN Photonics Ltd NLite System ([K000811](/device/K000811.md))
- ICN Photonics Ltd NLite System ([K013461](/device/K013461.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K024189

JUN 2 0 2003

## 510(k) Summary Statement ICN Photonics NLite System

1/2

### 1. General Information

- Submitter: ICN Photonics Ltd Units 1&2 Heol Rhosyn Dafen Parc Llanelli, Carmarthenshire Wales, UK, SA14 8QG
Dr Mike Kiernan - Director of Clinical Research Contact Person:

Summary Preparation Date: December 17th 2002

### 2. Names

Device Name: NLite System

Primary Classification Name: Laser Powered Surgical Instrument

### 3. Predicate Devices

- ICN Photonics Ltd NLite System, cleared August 2000, 510(k) approval . number K000811, expanded Indications for Use 510(k) approval number K013461, January 2002.
## 4. Product Description

The NLite System is a flashlamp pumped, pulsed dye laser consisting of the following:

- Main laser console incorporating the laser resonator and external optics, high . voltage delivery system, internal cooler, fluid circulation system, control unit and user interface;
- Flexible fibre optic delivery device and optical handpiece; ●
- Footswitch for pulsing control. ●

## 5. Indications for Use

The NLite system is indicated for use in the specialties of Dermatology and Plastic Surgery, and in particular for the treatment of Inflammatory Acne Vulgaris`.

{1}------------------------------------------------

024i89 2/2

## 6. Rationale for Substantial Equivalence

The laser system described in this submission is cleared for marketing under 510(k) approval K000811, August 2000. No modifications to the laser system have been undertaken.

## 7. Safety and Efficacy Information

Clinical data has been provided to demonstrate that the NLite system is safe and effective for the described indications for use.

#### 8. Conclusion

The NLite system has been found to be substantially equivalent to the predicate devices, specifically in technological design and operation and similar in nature to the desired physiological interactions. The design and manufacture of the device is in accordance with the relative international standards and the potential risk to operator and patient have been minimized.

The clinical data provided has demonstrated that the specific indications for use is met and the safety and efficacy of the system has been proved.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus or a series of interconnected human profiles.

Public Health Service

JUN 2 0 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Dr. Mike Kiernan Director of Clinical Research ICN Photonics Ltd Units 1 & 2 Heol Rhosyn Dafen Parc Llanelli, Carmarthenshire Wales, UK, SA 14 8QG

Re: K024189

Trade/Device Name: NLite System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: March 20, 2003 Received: March 24, 2003

Dear Mr. Kiernan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Dr. Mike Kiernan

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

for Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use Statement as Requested by FDA

KO24189 Not Known 510(k) Number (if known):

Device Name: NLite System

Indications for Use:

The NLite System is indicated for use in Dermatological and Plastic Surgery applications and this device is intended for use in the treatment of Inflammatory Acne Vulgaris.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Miriam C. Provost

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number K024189

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use

---

**Source:** [https://fda-staging.innolitics.com/device/K024189](https://fda-staging.innolitics.com/device/K024189)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
