# AIGIS-SUPER (K024142)

_Sb Lucius, Inc. · EJT · Mar 10, 2003 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K024142

## Device Facts

- **Applicant:** Sb Lucius, Inc.
- **Product Code:** [EJT](/productcode/EJT.md)
- **Decision Date:** Mar 10, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3060
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

AIGIS-SUPER is intended for manufacturing - Inlay / Onlays - - Crowns - - Short span bridges - - Long span bridges - - Removable partials -

## Device Story

AIGIS-SUPER is a 53% gold-based dental casting alloy used by dental professionals for the fabrication of dental restorations. The device is processed in a dental laboratory setting to create inlays, onlays, crowns, and both short and long-span bridges. It serves as a restorative material to replace missing or damaged tooth structure, providing a high-gold appearance. The alloy is melted and cast into the desired dental prosthetic shape. Its clinical benefit is the restoration of dental function and aesthetics for patients requiring fixed or removable dental prostheses.

## Clinical Evidence

Bench testing only. Physical and mechanical properties (melting point, Vickers hardness, yield strength, elongation, and density) were compared against the predicate device. Compliance with ANSI/ADA 5 and ISO 9693 standards was reported.

## Technological Characteristics

Gold-based dental casting alloy (53% gold). Physical properties: Melting range 1,768-1,948°F, Vickers hardness 150, yield strength 300 MPa, 21% elongation, density 13.8 g/cm3. Tested per ANSI/ADA 5 and ISO 9693.

## Regulatory Identification

A noble metal alloy is a device composed primarily of noble metals, such as gold, palladium, platinum, or silver, that is intended for use in the fabrication of cast or porcelain-fused-to-metal crown and bridge restorations.

## Special Controls

*Classification.* Class II (special controls). The special control for these devices is FDA's “Class II Special Controls Guidance Document: Dental Noble Metal Alloys.” The devices are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9. See § 872.1(e) for availability of guidance information.

## Predicate Devices

- JENSEN'S JYE

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Ko24142

MAR 1 0 2003

#### 510(K) SUMMARY

SB LUCIUS, INC. Submitter of 510(k): 9778 Katella Ave. Ste. 205, Anaheim, CA 92804 Phone: (714) 530-2814, Fax : (714) 530-3448,

| Contact person: | Dae Kyu Chang       |
|-----------------|---------------------|
| Phone:          | (714) 530-2814      |
| Fax:            | (714) 530-3448      |
| E-mail:         | Gatorkr@yahoo.co.kr |

Dec 15, 2002 Date of Summary:

| Trade name:          | AIGIS-SUPER                                                     |
|----------------------|-----------------------------------------------------------------|
| Common:              | Dental casting alloy                                            |
| Classification name: | Gold based alloys and precious metal alloys for<br>clinical use |
| Product code:        | EJT                                                             |
| Classification:      | Class II                                                        |

Legally marketed device: JENSEN'S JYE 510(k) number:

### SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS

Test methods applied: as in ANSI/ADA 5 and ISO 9693

#### Comparison of composition:

|                                                                                                                                                                                                                                                                                          |                                                                                                                |                                                                                     |                                                                                                      | <br>And Am S & S & S & S & S . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . C . | C<br>MAN BAFF PARTER STATES STATE W<br>and and the commission and the program | 11 11 1 1 1 1 1 1 1 2 2 2 2 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1                                                             |                                                                            |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| with A shoulded by the drivers and sciences and science many a many a many a many a many a many a many a many a many a<br>1                                                                                                                                                              | Sublem and and and one of the first program to the first for any and on any a proprises and                    | a manufactures and any a more a more a mand on the manage of second of the programs | 11 11 1000 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100                   | TARRE REWTER                                                                                                                                                                       | STATUTE A CONSULER OF THE AT                                                  | 1<br>100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000 | I I I Juli Bala Market Bill B I<br>of any a via Branch a wound of the U.S. |
| In most and an an an a not a marting company and program more in to the comments and in<br>1                                                                                                                                                                                             | S AC ART P WHERE A CONSULARIA & A 465 469 44 - 100 400 100 100 100 100 100 100 100 100 100 100 100 100 100 100 |                                                                                     | Line more it But do status and and a more a more a more<br>200                                       | A - S - S - F - F - F - F - F - - - - - - - -<br>1                                                                                                                                 |                                                                               | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------      | of maked by a banks and the program a fire many to                         |
| for the produces accession in areas announce annual a population and the company of the proper<br>1<br><br>на композительные продавания продавания продавания пристемента пристемента по страниции по полнения по полнения по полнения по страниции по полнение по страниции по полнение | 1.14 max 8 Back armanism bent production ++ F S newspan any in a BROULAVER BEFF                                |                                                                                     | 11/2009 11/04/2019 11/09/2019 11/04/2017 11/08/2017 11/08/2017 11/08/2017 11/08/2017 11/08/2017 11/2 | Corporation Formany to interestional and is and is a market and to a comments                                                                                                      |                                                                               |                                                                                                                                                                                     |                                                                            |

## COMPOSITION WEICHT W

### Comparison of physical and mechanical properties:

| Alloy       | Melting Point Range (°F) | Hardness (Vickers) | Yield Strength (MPa) | Elongation (%) | Density (g/cm3) |
|-------------|--------------------------|--------------------|----------------------|----------------|-----------------|
| JYE         | 1,545 - 1,635            | 170                | 345                  | 23             | 12.8            |
| AIGIS-SUPER | 1,768-1,948              | 150                | 300                  | 21             | 13.8            |

{1}------------------------------------------------

#### Discussion:

Since the composition of the legally marketed alloy and the new device is very similar, it may be assumed that the biological compatibility of the alloys is also very similar.

#### Conclusion:

The main elements and their concentration are almost identical, AIGIS-SUIPER is an inlay, onlay, crown and bridge alloy. This device is dependable 53% gold alloy with a high gold appearance. AIGIS-SUPER is excellent for inlays, three-quarter crowns, long and short-span bridges. AIGIS-SUPER is substantially equivalent to JENSEN'S JYE and the minor differences between them do not affect safety or effectiveness.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## MAR 1 0 2003

Mr. Dae-Kyu Chang SB LUCIUS, Incorporated 9778 Katella Avenue, Suite 205 Anaheim, California 92804

Re: K024142

Trade/Device Name: AIGIS-SUPER Regulation Number: 872.3060 Regulation Name: Gold-based Alloys and Precious Metal Alloys for Clinical Use Regulatory Class: II Product Code: EJT Dated: December 15, 2002 Received: December 16, 200

Dear Mr. Chang

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Chang

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Susan Tanner

Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

SE LUCIUS, INC 9778 Katella Ave. Ste. 205, Anaheim, CA 92804

Phone: (714) 530-2814, Fax : (714) 530-3448,

# INDICATIONS FOR USE

510(K) Number : K₀24/4₂

Device Name(s) : AIGIS-SUPER

AIGIS-SUPER is intended for manufacturing

- Inlay / Onlays -
- Crowns -
- Short span bridges -
- Long span bridges -
- Removable partials -

(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)

concurrence of chrd,office of DEVICE EVALUATION(OED)

MEadh ke MSR

siology, General Hospital. 510(k) Number

10

---

**Source:** [https://fda-staging.innolitics.com/device/K024142](https://fda-staging.innolitics.com/device/K024142)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).