PRODUCT LINE-COMBO STIMULATOR/SEQUENTIAL STIMULATOR, MODEL IF-5000
K024036 · Johari Digital Healthcare , Ltd. · IPF · Dec 20, 2002 · Physical Medicine
Device Facts
Record ID
K024036
Device Name
PRODUCT LINE-COMBO STIMULATOR/SEQUENTIAL STIMULATOR, MODEL IF-5000
Applicant
Johari Digital Healthcare , Ltd.
Product Code
IPF · Physical Medicine
Decision Date
Dec 20, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Interferential Current Stimulation is indicated for : A) 1. Symptomatic relief and management of chronic and intractable pain. 2. Adjunctive treatment in the management of post-surgical and post-traumatic, acute pain conditions. B) Electrical Muscle Stimulation (EMS or Russian) is indicated for: 1 . Relaxation of Muscle spasm. 2. Prevention or Retardation of disuse atrophy. 3. Increasing local blood circulation. 4. Muscle re-education. 5. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis. 6. Maintaining or increasing range of motion.
Device Story
Combo Stimulator IF-5000 is a portable, micro-computer controlled neuromuscular electrical stimulation system; used by clinicians in clinics or by patients at home for prescribed treatments. Device delivers Interferential Current (IFT) and Electrical Muscle Stimulation (EMS/Russian) via four channels (eight electrodes). Clinicians program treatment parameters via membrane panel and LCD; amplitude and time are adjustable by patient. Unique 'Combo' feature allows sequential delivery of EMS and IFT in one session. Device operates on rechargeable batteries, enabling off-site use. Output waveforms include biphasic and sinewave. Safety features include automatic intensity reduction upon component failure and transformer-based insulation. Benefits include pain relief, muscle spasm reduction, and improved circulation/range of motion.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included verification of output current/voltage, waveform characteristics, channel isolation per ANSI 3.2.3.2 (1985), and performance under open/short circuit conditions.
Technological Characteristics
ABS molded enclosure; HMOS micro-controller; rechargeable 1.2V Ni-Mh battery power source; two output channels (four channels total); LCD interface; membrane touch control. Waveforms: Biphasic and Sinewave. Connectivity: Standalone. Sterilization: Not applicable (non-sterile).
Indications for Use
Indicated for patients requiring symptomatic relief of chronic/intractable pain, adjunctive management of post-surgical/post-traumatic acute pain, relaxation of muscle spasms, prevention/retardation of disuse atrophy, increased local blood circulation, muscle re-education, prevention of venous thrombosis via calf muscle stimulation, and maintenance/increase of range of motion.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------

Image /page/0/Picture/1 description: The image contains two quality assurance logos. The logo on the left has a large "R" in the center, surrounded by the text "REGISTER QUALITY ASSURANCE ISO 9001". The logo on the right has a crown above a check mark, with the text "UKAS QUALITY Management" below. The text "ISO 9001 Certified" is at the bottom of the image.
Johari Digital Healthcare Ltd.
REGD. OFFICE : EC-1,2,3, Electronics Complex, Light Industrial Area, Jodhpur - 342 003 (Raj) INDIA Phone : 91-291-741183, 304591 (F), Fax : 742289, E-mail : joharicare@sify.com, joharis@ndf.vsnl.net.in Website : www.jo-hari.com
ANNEXURE 'VIII'
36
## SUMMARY
DEC 2 0 2002
| JOHARI DIGITAL HEALTHCARE LTD.<br>EC-1-2-3, Electronics Complex, Light Industrial Area,<br>Jodhpur 342003 (INDIA)<br>PHONE: +91-291-2741183 (F), 2430028 (R) FAX: +91-291-2742289<br>E-mail: joharicare@sify.com | | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---|----------------------------------------------------------------------------------|--|
| | | "Special 510(k) Summary" | |
| Submitter's name | : | JOHARI DIGITAL HEALTHCARE LTD. | |
| Address | : | EC-1-2-3, Electronics Complex<br>Light Industrial Area<br>Jodhpur - 342003 INDIA | |
| Phone | : | +91-291-741183 | |
| FAX | : | +91-291-2742289 | |
| E-mail | : | joharicare@sify.com | |
| Contact Person | : | Nisha Johari | |
| Date of Summary<br>is submitted | : | 27.11.2002 | |
| Resubmitting on | : | N.A. | |
8.1
Sales & Service Office : DELHI : 101-A, M.C. Bhawan, 11/56, D. B. Gupta Marg, Karol Bagh, Delhi-05, Tel. # 3624527, 3549454 MUMBAI : 317, Veena Beena Shopping Centre, Opp. Bandra Stn. (V) Tel. # 022-6409446 Pager # 9602-157374 Fax : 6514955 CHENNAI : No. 16, 1st Floor, Duraiswamy Road, T. Nagar Tel. # 044-4310159 Pager # 9632-739632 INDORE : 14, R.K. Puram Colony, A.B. Road Tel. # 0731-263075 KANPUR : Tel. # 0512-610371
Better life for Better Today
{1}------------------------------------------------
## ANNEXURE 'VIII'
## MODIFIED DEVICE FOR WHICH SUBMITTING
| Trade Name : | Combo Stimulator IF-5000 |
|-----------------------|-----------------------------------------------------------------------------------------|
| Common Name : | Electrical Nerve & Muscle Stimulator |
| Classification Name : | Physical Medicine (Per 21 CFR Section 890) and Neurology (Per 21 CFR Section 882-5890). |
. .
## LEGALLY MARKETED DEVICE
| Analgesic Pulsar AP-439 | : | Nerve & Interferential Stimulator<br>510(k) Number K993229 |
|-------------------------|---|------------------------------------------------------------------------------------|
| Manufacturer | : | JOHARI DIGITAL HEALTHCARE LTD |
| Address | : | EC-1-2-3, ELECTRONIC COMPLEX,<br>LIGHT INDUSTRIAL AREA,<br>JODHPUR – 342 003 INDIA |
| Tel. | : | +91-291-2741183 |
+91-291-2741183
{2}------------------------------------------------
### ANNEXURE 'VIII'
### DESCRIPTION OF MODIFIED DEVICE IF-5000
Johari 'Combo Stimulator' is a Neuromuscular Electrical Stimulation System, consists of programmable controller just like the predicate device Analgesic Pulsar AP-439. This unit Combo Stimulator allows the treatment specifically tailored to the precise needs of the patient in the clinic and for subsequent treatment off site. The clinicians can set the treatment parameters and lock it, so patient can repeatedly use the unit as prescribed by the clinician. The stimulation is most comfortable and the waveforms are unique.
It has been the clinician requirement many a times where they wanted to have a unit that can perform for muscle stimulation followed by interferential, so that, the painful muscle condition can be treated. Usually they have to turn the machine On and Off for different stimulation waveforms. The unique feature of Combo Stimulator is that in such above conditions, the clinicians can set the unit to perform in such a way that desired muscle stimulation can be followed by desired Interferential stimulation just by pressing one key "Combo", otherwise in itself the unit can perform muscle stimulation through all the outputs or Interferential stimulation through all the output.
The ease and simplicity by which the programming can be done is not more than touch of the key on the panel and LCD display. All the parameters remain in the system control except the amplitude & time which can be set by the patient himself. This unit is ergonomically designed, so that, it is comfortable to hold in hands.
It is portable and has two output channels, four channels works on rechargeable battery system. The unit can perform offsite wherein the hospital room, clinic or home while the predict device Analgesic Pulsar could not run on rechargeable battery system and so limited its use near to the AC outlet. This has always limited the clinicians to perform on their patient. Also many times, the patient may be in need to have a repeated treatment and this modified device "Combo Stimulator" can work in such cases. The Combo Stimulator can provide True Interferential with desired frequency, sweep selection and muscle stimulation of desired mode similar to Analgesic Pulsar.
Combo Stimulator IF-5000 comes complete with all the necessary components to perform Pain Relief & Electrical Stimulation. Below is a list of items that are included:
| 1. | Combo Stimulator IF-5000 unit | 01 |
|----|-----------------------------------------------|----|
| 2. | Electrode Cable (4 Core) | 02 |
| 3. | Electrode 2.75" Round Diameter(Self Adhesive) | 08 |
| 4. | Adapter | 01 |
| 5. | Rechargeable batteries | 04 |
| 5. | Manual | 01 |
| 6. | Carry Bag | 01 |
The only difference in these two devices is that:
- Predicate Device AP-439 is a bigger in size. a)
- Predicate Device AP-439 uses direct AC mains whereas Combo Stimulator IF-5000 uses b) rechargeable battery system and makes it much more safe and effective.
- Both the units have two outputs with four channels (8 electrodes). c)
- Both are Micro-computer controlled and digital displays guide through the programming. d)
- Both are ABS molded ergonomically designed but since IF-5000 is small in size, which can e) be used repeatedly and at home also.
{3}------------------------------------------------
### INTENDED USE OF MODIFIED DEVICE IF-5000
Modified unit "Combo IF 5000" has similar intended use as predicate device Analgesic Pulsar AP-439. Both the electrical stimulators are to be used in Physical Medicine by Clinicians and Therapists.
### INTENDED USE
- Interferential Current Stimulation is indicated for : A)
- 1. Symptomatic relief and management of chronic and intractable pain.
- Adjunctive treatment in the management of post-surgical and post-traumatic, acute pain 2. conditions.
- B) Electrical Muscle Stimulation (EMS or Russian) is indicated for:
- 1 . Relaxation of Muscle spasm.
- 2. Prevention or Retardation of disuse atrophy.
- 3. Increasing local blood circulation.
- 4. Muscle re-education.
- న్. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis.
- 6. Maintaining or increasing range of motion.
### INTENDED USE OF ANALGESIC PULSAR AP-439
### (A) Interferential current stimulation , Premodulated Bipolar Mode and Faradic Stimulation is indicated for :
- 1. Symptomatic relief and management of chronic, intractable pain.
- 2. Adjunctive treatment in the management of post-surgical and post-traumatic, acute pain conditions.
### (B) Electrical Muscle Stimulation ( Russian and Galvanic-interrupted) is indicated for:
- 1. Relaxation of Muscle spasm.
- 2. Prevention or Retardation of disuse atrophy.
- 3. Increasing local blood circulation.
- 4. Muscle re-education.
- 5. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis.
- 6. Maintaining or increasing range of motion.
### (C) Galvanic-DC Continuous mode is indicated for:-
- 1. Relaxation of muscle spasm.
{4}------------------------------------------------
# TECHNICAL SPECIFICATIONS OF THE MODIFIED DEVICE IF-5000
IF-5000 has basically same Technical Specification as predicate Analgesic Pulsar AP-439.
| Power Source<br>Number of outputs<br>Channels<br>Synchronous | : 1.2V x 4 Ni-Mh Rechargeable batteries<br>: Two<br>: Four<br>: Yes |
|------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Max out put current | : Select PW & Ramp Burst - 90mA pp @ 500 ohm load<br>Russian & IFT – 54mA RMS @ 500 Ohm Load |
| Max out put voltage | Select PW & Ramp Burst - 90mA vpp @ 10K ohm loac<br>Russian & IFT – 37V RMS @ 10k Ohm Load |
| Channel isolation | : Yes, confirm to ANSI 3.2.3.2, 1985. |
| Waveform | : Biphasic & Sinewave |
| Current density<br>on 2" diameter electrodes | Select PW & Ramp Burst - 0.533 mA/cm2 @ 500 ohm<br>Russian & IFT - 2.66mA/cm2 @ 500 Ohm |
| Power density<br>On 2" diameter electrodes | Select PW & Ramp Burst – 0.024 W/cm2 @ 500 ohm<br>Russian & IFT – 0.072 W/cm2 @ 500 Ohm |
| Max. phase charge<br>at 500 Ohm<br>at 2 K Ohm<br>at 10 K Ohm | Select PW & Ramp Burst<br>18 micro coulomb<br>7.4 micro coulomb<br>1.8 micro coulomb |
| at 500 Ohm<br>at 2 K Ohm<br>at 10 K Ohm | Russian<br>5.4 micro coulomb<br>1.7 micro coulomb<br>0.37 micro coulomb |
| at 500 Ohm<br>at 2 K Ohm<br>at 10 K Ohm | IFT<br>2.7 micro coulomb<br>0.85 micro coulomb<br>0.18 micro coulomb |
| Modulation Options<br>Amplitude<br>Frequency<br>Mechanical Specifications<br>Pulse Width | Preset IFT & Russian only<br>IFT section only<br>Weight 1 Lb, Size: 3.9"(l) x 1.6"(h) x 4.0(d)<br>Selectable PW: 50 to 400 $\mu$ s in step of 50<br>Ramp Burst: 400 $\mu$ s<br>Russian: 200 $\mu$ s<br>IFT: 100 $\mu$ s |
{5}------------------------------------------------
## TECHNICAL SPECIFICATIONS OF THE PREDICATE DEVICE DEVICE ANALGESIC PULSER AP-439
| Power Source | : 110V AC, 50/60 Hz. |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Number of outputs | : Two |
| Channels | : Four |
| Synchronous | : Yes |
| Max out put current | : 100mA (max. @ 500 ohm loads) |
| Max out put voltage | 132V pp (max. @ 10K ohm loads) |
| Channel isolation | : Yes, confirm to ANSI 3.2.3.2 , 1985. |
| Waveform | : Monophasic, Biphasic, Sinewave & D.C. Line |
| Current density<br>on 2" diameter<br>electrodes | : 3.55mA/cm2 (max @ 500 ohm load) |
| Power density<br>on 2" diameter<br>electrodes | : 0.127 W/cm2 (max @ 500 ohm load) |
| Max. phase charge | : 40 micro coulomb (max @ 500 ohms load) |
| Modulation Options | |
| Amplitude | : Preset IFT only |
| Frequency | : IFT section only |
| Mechanical Specifications | : Weight 5.5 Lb, 12.2" (l)X 3.5" (h)X 7.5" (d) |
| Pulse width | : Faradic: 100-400 (selectable)<br>IFT: 125 µs. phase interval<br>Russian: 200 µs<br>Galvanic: D.C. |
{6}------------------------------------------------
## TECHNOLOGICAL CHARACTERISTICS
#### 1. MICRO-CONTROLLER
Combo Stimulator IF-5000 & Predicate device Analgesic Pulsar AP-439 both the units are programmable Microcomputer Controlled. HMOS Micro-controller is used in IF-5000, which makes it reliable with Electro Magnetic Compatibility while in AP-439, Microcontroller 87C51 is used on a base of oscillating crystal at frequency 12M Hz.
### 2. MULTICHANNEL
Combo Stimulator IF-5000 & Predicate device AP-439 both the units have two outputs (four channels - eight electrodes) to provide different kind of stimulation to body sites.
### MULTIWAVE SELECTIONS & PROGRAMABILITY 3.
Combo Stimulator IF-5000 produces selection of Muscle Stimulation (selectable pulse width mode, ramp burst & Russian 2500 Hz.) & Interferential Stimulation (frequency of 5000 Hz with three fixed sweep frequencies and one fixed beat frequency.) for pain relief through all the outputs. If desired, the clinicians can select the Combination, which combines EMS (for 1 minute and more) and IFT (for 1 minute and more) throughout the treatment session.
Predicate device AP-439 produces selection of Interferential (4000 Hz.-Vector & Premod with programmable sweep frequency), Russian (2500 Hz), Galvanic & Faradic (6 preset programs with selectable pulse width and pulse rate).
### 4. DISPLAY
Combo Stimulator IF-5000 has LCD (Liquid Crystal Display) for showing Treatment Mode, Contraction period. Relaxation period. Treatment Time and Intensity in both the outputs. Predicate device AP-439 has LED display to show Percentage Power, Cyclic On /Off Time, Treatment Time and balance Treatment Time.
#### 5. POWER
Combo Stimulator IF-5000 works on rechargeable battery back up, so the treatment will not be limited to site near AC outlet and can be carried along. This makes the unit further safe. While Predicate device Analgesic Pulsar AP-439 works on 110V 60Hz AC mains only.
#### TACTILE TOUCH CONTROL 6.
Both the units are beautifully designed with the Membrane Panel, provides easy touch entry for treatment selection. The audible note with each key entry ensures clinician that selection is made.
#### 7. CASING
Both the units are attractive and fitted in an ABS body enclosure. They are strong and sturdy. IF-5000 is handy, smaller and more ergonomic design, so that can be held in hand while AP-439 is a desktop unit.
#### 8. WEIGHT
Combo Stimulator IF-5000 - Weigh 1.0 Lb., Predicate device AP-439 - weigh 5.5 lb.
{7}------------------------------------------------
### ANNEXURE 'VIII'
### SAFETY:
- 1. From circuitry to output again there is an insulation through transformers, thereby provision of double separations between mains & the human body (where output is to be applied)
- 2. Membrane key boards are provided which are known to withstand operation of key pressing much more than other mechanical methods.
- 3. Circuitry efficacy is good as matrix circuitry is used in keyboard & amplification circuitry is simple.
- 4. In case failure of any IC or component, the intensity will automatically come down to zero in
- 5. IF-5000 & AP-439 can be carried along for easy use at home or clinics.
### SUMMARY
Combo Stimulator IF-5000 is not designed to be worm by patient.
IF-5000 has a self-contained safety circuit.
Open/short circuit performance: This IF-5000 functions normally after open and short circuited conditions between output jacks, with the device operating for maximum of 15 minutes, in each condition at the maximum available setting of pulse width, pulse rate and pulse amplitude,
{8}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
**DEC 20 2002**
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Nisha Johari Johari Digital Healthcare Ltd. Vandana, 28 Nehru Park Jodhpur - 342003, India
## Re: K024036
Trade/Device Name: Combo Stimulator IF-5000 Regulation Number: 21 CFR 890.5850 and unclassified Regulation Name: Powered Electrical Muscle Stimulator and Interferential Current Electrical Stimulator
Regulatory Class: Class II
Product Code: IPF and LIH Dated: November 27, 2002 Received: December 6, 2002
Dear Mrs. Johari:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{9}------------------------------------------------
Page 2 -- Mrs. Nisha Johari
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincerely yours.
Miriam C. Provost
of Celia M. Witter., Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
## INDICATIONS FOR USE
Combo Stimulator IF-5000 provides combination of desired stimulating currents, which can be chosen by clinicians for the following treatment therapies:
### INTENDED USE
- Interferential Current Stimulation is indicated for: (A)
- Symptomatic relief and management of chronic, 1. intractable pain.
- 2. Adjunctive treatment in the management of postsurgical and post-traumatic, acute pain conditions.
- (B) Electrical Muscle Stimulation (EMS or Russian) is indicated for:
- Relaxation of Muscle spasm. 1 .
- Prevention or Retardation of disuse atrophy. 2.
- Increasing local blood circulation. 3.
- Muscle re-education. 4.
- 5. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis.
- 6. Maintaining or increasing range of motion.
Miriam C. Provost
ivision of General, Restorative 1d Neurological Devices
K624036
6.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.