XTJF-160AF DUODENOVIDEOSCOPE

K024033 · Olympus Optical Co., Ltd. · FDT · Dec 20, 2002 · Gastroenterology, Urology

Device Facts

Record IDK024033
Device NameXTJF-160AF DUODENOVIDEOSCOPE
ApplicantOlympus Optical Co., Ltd.
Product CodeFDT · Gastroenterology, Urology
Decision DateDec 20, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Indications for Use

The XTJF-160AF has been designed to be used with an Olympus EVIS video system center, light source, Documentation Equipment, Video Monitor, Endo-Therapy accessories (such as a Forceps, Electrosurgical Accessories, etc.), Electrosurgical Unit, and other Ancillary Equipment for endoscopy and endoscopic surgery within the duodenum. Do not use these instruments for any purpose other than their intended uses.

Device Story

XTJF-160AF is a duodenovideoscope used for endoscopic procedures and surgery within the duodenum. It functions as part of the Olympus EVIS video system, connecting to a light source, video processor, monitor, and electrosurgical unit. The device features a variable stiffness mechanism, allowing the operator to adjust the scope's flexibility during insertion. It is operated by a physician in a clinical setting. The device provides visual feedback to the physician via the video system, facilitating diagnostic and therapeutic interventions. The variable stiffness design is intended to assist in navigation, leveraging a mechanism previously cleared in colonovideoscope predicates. The device is used with standard endo-therapy accessories like forceps and electrosurgical tools to perform procedures.

Clinical Evidence

Bench testing only. The device was tested for compliance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-18. Biocompatibility of patient-contacting materials was confirmed per ISO 10993-1.

Technological Characteristics

Duodenovideoscope with variable stiffness mechanism. Materials are identical to previously cleared devices and compliant with ISO 10993-1. Complies with IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), and IEC 60601-2-18 (endoscopic equipment). System includes video processor, light source, and electrosurgical unit connectivity.

Indications for Use

Indicated for endoscopy and endoscopic surgery, including electrocautery, within the duodenum.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the circle is an abstract image of an eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Olympus Optical Co., Ltd. % Mr. Donald J. Sherratt Medical Stream Director Intertek Testing Services NA, Inc. ETL SEMKO 70 Codman Hill Road Boxborough, MA 01719 JUL 2,7,2015 Re: K024033 > Trade/Device Name: XTJF - 160AF Duodenovideoscope Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDT Dated (Date on orig SE Itr); December 5, 2002 Received (Date on orig SE ltr): December 6, 2002 Dear Mr. Sherratt, This letter corrects our substantially equivalent letter of December 20, 2002. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be {1}------------------------------------------------ Page 2 - found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ 510(k) Number(if known): _____________________________________________________________________________________________________________________________________________________ Device Name:_XTJF-160AF DUODENOVIDEOSCOPE Indications for Use: The XTJF-160AF has been designed to be used with an Olympus EVIS video system center, light source, Documentation Equipment, Video Monitor, Endo-Therapy accessories (such as a Forceps, Electrosurgical Accessories, etc.), Electrosurgical Unit, and other Ancillary Equipment for endoscopy and endoscopic surgery within the duodenum. Do not use these instruments for any purpose other than their intended uses. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) V Prescription Use_ ્વર Over-The-Counter Use_ (Per 21 CFR 801.109) Daniel A. Lannon (Division Sign-Off) Division of Reproductive. Abdominal and Radiological Devices 510(k) Number (Optional Format 1-2-96) {3}------------------------------------------------ KO24033 # DEC 2 0 2002 # SMDA 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR Section 807.92. # A. Submitter's Name, Address, Phone and Fax Number | 1. Applicant: | Olympus Opt-Electronics Co., Ltd.<br>(New Company Name: AIZU Olympus Co., Ltd.)<br>500 Aza Muranishi Ooaza Niidera, Monden-machi<br>Aizuwakamatsu-shi Fukushima, Japan, 965-8520<br>Registration No.: 9610595 | | | |---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--| | 2. Initial Importer | Olympus America Inc.<br>Two Corporate Center Drive<br>Melville, NY 11747-3157<br>Registration No.: 2429304 | | | | 3. Contact Person | Junko Obata | | | | 3. Contact Person | Junko Obata | |-------------------|-------------------------------------| | | Olympus Optical Co., Ltd. | | | 2951 Ishikawa-cho | | | Hachioji-shi, Tokyo 192-8507, Japan | | | TEL 81-426-42-2891 | | | FAX 81-426-42-2291 | ## B. Device Name, Common Name - 1. Trade/Proprietary Name and Common Name Trade Name: XTJF-160AF DUODENOVIDEOSCOPE Common Name: DUODENOVIDEOSCOPE - 2. Class, Classification Number and Classification Name | AER Number - -<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Classification iver<br>Name | · · Product Cope · | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|--------------------| | 876.1500 | Endoscopes and accessones | R-KOG | - 3. Identification of Legally Marketed Devices Which we Claim Substantial Equivalence The following listed devices are seem to be as predicate devices in consideration of its characteristic and the following table shows regulatory history. | Model | #510(k) | Manufacturer | Class | Product Code | |-------------------------------------|----------|----------------------|-------|--------------| | TJF-140R<br>Duodenovideoscope | #K980465 | Olympus Optical Co., | II | 78-KOG | | JF-V10<br>Videoimage Duodenoscope | #K853585 | Olympus Optical Co., | II | 78-KOG | | EVIS EXERA<br>Colonovideoendoscopes | #K001241 | Olympus Optical Co., | II | 78-KOG | {4}------------------------------------------------ ### C. Summary Description of the Device - 1. Summary This premarket notification is proposed to establish the safety and effectiveness of the variable stiffness scope that is intended for endoscopy and endoscopic surgery, including electrocautery application, within the duodenum. The intended use of the XTJF-160AF is identical to that of previously cleared device, the TJF-140R and JF-V10. The mechanical structure of variable stiffness is identical to that of another predicate device, the EVIS EXERA Colonovideoendoscopes. While the variable stiffness scope is brand new device in the duodenum, endoscopy and endoscopic surgery, including electrocautery, within the duodenum is popular and the mechanical structure of variable stiffness has been confirmed to be safe by the EVIS EXERA Colonovideoendoscopes. Any new insertion method or techniques in consequence of the added mechanical structure of variable stiffness are not considered. Therefore, this modification does not affect the safety and effectiveness of the proposed device, #### 2. Design XTJF-160AF has been designed, manufactured and tested in compliance with voluntary safety standards. It meets the requirement of IEC60601-1, IEC60601-1-2 and IEC60601-2-18. - 3. Materials There are no new patient contacting materials. All of patient contacting materials is cleared by previous 510(k). And all materials have been confirmed with ISO 10993-1. #### 4. Intended Use of the device The XTJF-160AF has been designed to be used with an OLYMPUS EVIS Video System Center, Light Source, Documentation Equipment, Video Monitor, Endo Therapy Accessories (such as a Forceps, Electrosurgical Accessories, etc.), Electorosurgical Unit, and other Ancillary Equipment for endoscopy and endoscopic surgery within the duodenum. #### 5. Summary including Conclusions drawn from Non-clinical Tests When compared to the predicate device, XTJF-160AF does not incorporate any significant changes in the intended use, method of operation, material, or design that could affect the safety or effectiveness.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...