← Product Code [LLZ](/productcode/LLZ) · K024028

# SHARP VIEW IMAGE ENHANCEMENT SYSTEM (K024028)

_Context Vision AB · LLZ · Jan 24, 2003 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K024028

## Device Facts

- **Applicant:** Context Vision AB
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Jan 24, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The SharpView Image Enhancement System is intended for use by a qualified/trained technologist for transfer, storage, enhancement, and viewing of multi-modality images from a variety of diagnosis imaging systems.

## Device Story

SharpView is an image enhancement system consisting of software and a hardware image processing board (Javelin PCI-bus) installed on a PC. It receives digital medical images in DICOM 3 format over a network from various diagnostic imaging systems. The system performs image enhancement using the GOP Enhancement software core. Enhanced and original images can be viewed, stored locally, or transmitted via network. Used in clinical settings by trained technologists to assist in diagnostic workflows. The device facilitates image visualization, potentially aiding clinical decision-making by improving image quality for the interpreting physician.

## Clinical Evidence

Bench testing and user testing were performed to validate the system. No clinical data was required or provided.

## Technological Characteristics

PC-based system; Javelin PCI-bus image processing board; GOP Enhancement software core; DICOM 3 connectivity; compatible with Windows 98, NT 4.0, 2000, and XP; storage via hard disk, optical, CD-ROM, or tape.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- ContextVision SharpView Image Enhancement System ([K993802](/device/K993802.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## EXHIBIT 2

JAN 2 4 2003

ContextVision AB Storgatan 24 SE-582 23 Linköping SWEDEN Phone: +46 (0)13 35 85 61 Fax: +46 (0)13 10 42 82 CONTACT: Kent Strandlund

November 29, 2002

## 510(k) Summary of Safety and Effectiveness

- 1. Identification of the device: Proprietary-Trade Name: SharpView Classification Name: Image Processing, System Product code: 90LLZ Common/Usual name: SharpView Image Enhancement System
- Equivalent legally marketed devices: This product is similar in design and function to 2. the ContextVision SharpView Image Enhancement System, K993802
- Indications for Use (intended use); The SharpView Image Enhancement System is 3. intended for use by a qualified/trained technologist for transfer, storage, enhancement, and viewing of multi-modality images from a variety of diagnosis imaging systems.
- 4. Description of the device: The product is a software or a kit containing software and hardware (Image processing board), which is intended to be installed on a personal computer. Typically the personal computer receives digital medical images in DICOM 3 format over a network connection. The enhanced and original image can be sent in DICOM 3 format over the network connection. The original file and the enhanced file can be kept locally if selected.
- 5. Safety and Effectiveness, comparison to predicate device. The result of bench and user testing indicates that the modified device is as safe and effective as the predicate device.

{1}------------------------------------------------

- 6. Substantial Equivalence Chart

| Characteristics:             | SharpView Image<br>Enhancement System                                                                                                                                    | SharpView Image<br>Enhancement System, modified                                                                                                                                      |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use:                | The Image<br>Enhancement System is<br>intended for use by a<br>qualified/trained<br>technologist for transfer,<br>storage, enhancement,<br>and viewing of MRI<br>images. | The Image<br>Enhancement System is<br>intended for use by a<br>qualified/trained<br>technologist for transfer,<br>storage, enhancement,<br>and viewing of multi-<br>modality images. |
| Physical<br>characteristics: |                                                                                                                                                                          |                                                                                                                                                                                      |
| Computer                     | PC compatible                                                                                                                                                            | Same                                                                                                                                                                                 |
| Operating system             | Windows 98, NT 4.0                                                                                                                                                       | Windows 98, NT 4.0,<br>2000 and XP                                                                                                                                                   |
| Storage                      | Hard disk or any<br>compatible PC method:<br>Optical, CDROM, Tape                                                                                                        | Same                                                                                                                                                                                 |
| Image processing<br>board    | MIP-PCI                                                                                                                                                                  | Javelin (PCI-bus)                                                                                                                                                                    |
| Software core                | GOP® Enhancement<br>software                                                                                                                                             | Same<br>(Trademark is the<br>property of<br>ContextVision)                                                                                                                           |
| Image input                  | DICOM 3                                                                                                                                                                  | Same                                                                                                                                                                                 |

## 7. Conclusion

After analysing both bench and user testing data, it is the conclusion of ContextVision AB that the multi-modality SharpView Image Enhancement System is as safe and effective as the predicate device, has few technological differences, and only a minor change to the indications for use compared with the predicate device, thus rendering it substantially equivalent to the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human figure, composed of three curved lines.

JAN 2 4 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Kent Strandlund Quality & Regulatory Affairs ContextVision AB Storgatan 24 SE-582 23 Linköping SWEDEN

Re: K024028 Trade/Device Name: SharpView Image Enhancement System Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: November 29, 2002 Received: December 6, 2002

Dear Mr. Strandlund:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that vour device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## j) Indications for Use

510(k) Number_Ko24028

Device Name: SharpView Image Enhancement System

Indications for Use: The SharpView Image Enhancement System is intended for use by a qualified/trained technologist for transfer, storage, enhancement, and viewing of multi-modality images from a variety of diagnosis imaging systems.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use OR

Over the Counter Use_ (Per 21 CFR801.109)

Nancy C. Brogdon
(Division Sign-Off)

Division of Rons and Radiologi 510(k) Nul

---

**Source:** [https://fda-staging.innolitics.com/device/K024028](https://fda-staging.innolitics.com/device/K024028)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
