← Product Code [MOS](/productcode/MOS) · K023982

# LEO III TX/RX QUADRATURE KNEE COIL; ALPHA III TX/RX QUADRATURE WRIST COIL (K023982)

_Usa Instruments, Inc. · MOS · Feb 21, 2003 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K023982

## Device Facts

- **Applicant:** Usa Instruments, Inc.
- **Product Code:** [MOS](/productcode/MOS.md)
- **Decision Date:** Feb 21, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Leo III Tx/Rx Quadrature Knee Coil is a transmit-receive quadrature RF coil, used for obtaining diagnostic images of the knee and surrounding anatomy in Magnetic Resonance Imaging Systems. The Leo III Tx/Rx Quadrature Knee Coil is designed for use with the Magnetom Trio 3.0T MRI system manufactured by Siemens Medical Solutions, Inc. The indications for use are the same as for standard MR Imaging. The Alpha III Tx/Rx Quadrature Wrist Coil is a transmit-receive quadrature RF coil, used for obtaining diagnostic images of the wrist and hand in Magnetic Resonance Imaging Systems. The Alpha III T/R Quadrature Wrist Coil is designed for use with the Magnetom Trio 3.0T MRI system manufactured by Siemens Medical Solutions, Inc. The indications for use are the same as for standard MR Imaging.

## Device Story

Leo III and Alpha III are multi-element quadrature transmit-receive RF coils for 3.0T MRI systems; designed for knee/surrounding anatomy and wrist/hand imaging respectively. Coils function as passive accessories to the MRI scanner; they receive RF signals from hydrogen nuclei and transmit RF pulses. Enclosed in rigid, fire-rated plastic housings with open windows for patient comfort/positioning. Safety features include PIN diodes for transmit/receive channel isolation, active RF blowing fuses to prevent resonant loops, and non-conductive housing materials. Power deposition is managed by the host MRI system's SAR algorithm. Used in clinical imaging environments by trained MRI technologists/radiologists; output is diagnostic-quality MR images interpreted by physicians to aid clinical diagnosis.

## Clinical Evidence

Bench testing only. No clinical data provided. Safety and effectiveness supported by comparison of technical specifications, materials, and design features to legally marketed predicate devices.

## Technological Characteristics

Multi-element quadrature transmit-receive RF coils. Materials: Flame retardant polyurethane, polycarbonate, Kydex Acrylic/PVC, Delrin Acetal. Leo III: 12 z-inductor design. Alpha III: 2-element saddle design. Connectivity: Direct interface to Siemens Magnetom Trio 3.0T MRI system. Safety: PIN diode switching, active RF blowing fuses, non-conductive housing. SAR limited by host system algorithm.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Allegra 3.0Tesla TxRx Head coil ([K002179](/device/K002179.md))
- Leo 7000 Knee Coil ([K971246](/device/K971246.md))
- Alpha 7000 Wrist Coil ([K972205](/device/K972205.md))

## Reference Devices

- Hi-Res 9000 Phased Array Carotid Coil ([K001210](/device/K001210.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# SUMMARY OF SAFETY AND EFFECTIVENESS

# K023982

| 1. Device Name:                   | Magnetic Resonance Imaging Accessory                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Proprietary Name:              | Leo III Tx/Rx Quadrature Knee Coil                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| 3. Classification:                | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| 4. Establishment Registration #:  | 1529041                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| 5. Manufacture Facility Location: | USA Instruments, Inc.,<br>1515 Danner Drive<br>Aurora, Ohio 44202, USA<br>Telephone: 330-562-1000; Fax: 330-562-1422.                                                                                                                                                                                                                                                                                                                                                                |
| 6. Performance Standard:          | No applicable performance standards have been<br>issued under Section 514 of the Food, Drug and<br>Cosmetic Act.                                                                                                                                                                                                                                                                                                                                                                     |
| 7. Intended Use:                  | The Leo III Tx/Rx Quadrature Knee Coil is a<br>transmit-receive quadrature RF coil, used for<br>obtaining diagnostic images of the knee and<br>surrounding anatomy in Magnetic Resonance<br>Imaging Systems. The Leo III Tx/Rx Quadrature<br>Knee Coil is designed for use with the Magnetom<br>Trio 3.0T MRI system manufactured by Siemens<br>Medical Solutions, Inc. The indications for use are<br>the same as for standard MR Imaging.                                          |
| 8. Device Description:            | The Leo III Tx/Rx Quadrature Knee Coil is a multi-<br>element quadrature transmit-receive coil. The coil<br>has a split top design with a latching device and a<br>large open viewing window on the top. The open,<br>patient friendly design eases patient handling and<br>positioning and maximizes patient comfort. The coil<br>elements and accessory electronics are enclosed in<br>a rigid plastic housing, which is fire rated and has a<br>high impact and tensile strength. |

Please turn over

{1}------------------------------------------------

## 9. Safety and Effectiveness

| Leo III Tx/Rx Quadrature Knee Coil<br>Product Features                                                                                                                                                                      | Comparison to predicate device or other 510(k)<br>cleared products                                                                                                                           |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: Imaging of the knee and<br>surrounding anatomy.                                                                                                                                                               | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179)<br>-Similar to the Leo 7000 Knee Coil manufactured<br>by USA Instruments, Inc. (K971246) |
| Indications for Use: Identical to routine<br>MRI imaging.                                                                                                                                                                   | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179)<br>-Similar to the Leo 7000 Knee Coil manufactured<br>by USA Instruments, Inc. (K971246) |
| Coil Enclosure Material:<br>Flame Retardant Polyurethane<br>Flame Retardant Polycarbonate<br>Kydex Acyrlic/PVC<br>Delrin Acetal                                                                                             | - Similar to the Hi-Res 9000 Phased Array Carotid<br>Coil manufactured by USA Instruments, Inc.<br>(K001210)                                                                                 |
| Coil Design: 12 z-inductor transmit<br>receive quadrature coil.                                                                                                                                                             | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179)                                                                                          |
| Decoupling: This is a transmit and<br>receive coil. There is no decoupling, the<br>Transmit / Receive Switch is actively<br>switched during transmit and receive<br>modes.                                                  | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179)                                                                                          |
| Prevention of RF Burns: The coil's<br>transmit /receive switch uses pin diodes<br>to isolate the receive channel from the<br>transmit channel; coil elements and<br>circuitry are enclosed in a non-<br>conductive housing. | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179)                                                                                          |
| Radio Frequency Absorption: Power<br>deposition during imaging is limited by<br>SAR algorithm                                                                                                                               | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179)                                                                                          |
| Formation of Resonant Loop: Active<br>diodes and fast RF blowing fuses<br>isolate the coil elements from RF fields;<br>length of cable and stiffness does not<br>permit looping                                             | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179)                                                                                          |

{2}------------------------------------------------

## SUMMARY OF SAFETY AND EFFECTIVENESS

| 1. Device Name:                    | Magnetic Resonance Imaging Accessory                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Proprietary Name:               | Alpha III Tx/Rx Quadrature Wrist Coil                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| 3. Classification:                 | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| 4. Establishment Registration #:   | 1529041                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| 5.  Manufacture Facility Location: | USA Instruments, Inc.<br>1515 Danner Drive<br>Aurora, Ohio 44202, USA<br>Telephone: 330-562-1000; Fax: 330-562-1422.                                                                                                                                                                                                                                                                                                                                                   |
| 6. Performance Standard:           | No applicable performance standards have been<br>issued under Section 514 of the Food, Drug and<br>Cosmetic Act.                                                                                                                                                                                                                                                                                                                                                       |
| 7. Intended Use:                   | The Alpha III Tx/Rx Quadrature Wrist Coil is a<br>transmit-receive quadrature RF coil, used for<br>obtaining diagnostic images of the wrist and hand in<br>Magnetic Resonance Imaging Systems. The Alpha<br>III T/R Quadrature Wrist Coil is designed for use<br>with the Magnetom Trio 3.0T MRI system<br>manufactured by Siemens Medical Solutions, Inc.<br>The indications for use are the same as for standard<br>MR Imaging.                                      |
| 8. Device Description:             | The Alpha III T/R Quadrature Wrist Coil is a multi-<br>element quadrature transmit-receive coil. The coil<br>has a rigid enclosure with windows on top to<br>facilitate patient handling. The open, patient friendly<br>design eases patient handling and positioning and<br>maximizes patient comfort.  The coil elements and<br>accessory electronics are enclosed in a rigid plastic<br>housing, which is fire rated and has a high impact<br>and tensile strength. |

Please turn over

{3}------------------------------------------------

### 9. Safety and Effectiveness

| Alpha III Tx/Rx Quadrature Wrist Coil<br>Product Features                                                                                                                                                                                        | Comparison to predicate device or other 510(k)<br>cleared products                                                                                                                                         |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: Imaging of the wrist and<br>hand                                                                                                                                                                                                   | -Similar to the Alpha 7000 Wrist Coil manufactured<br>by USA Instruments, Inc. (K972205)                                                                                                                   |
| Indications for Use: Identical to routine<br>MRI imaging.                                                                                                                                                                                        | -Similar to the Alpha 7000 Wrist Coil manufactured<br>by USA Instruments, Inc. (K972205)<br>-Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179)             |
| Coil Enclosure Material:<br>Flame Retardant Polyurethane<br>Flame Retardant Polycarbonate<br>Kydex Acrylic/PVC<br>Delrin Acetal                                                                                                                  | - Similar to the Hi-Res 9000 Phased Array Carotid<br>Coil manufactured by USA Instruments, Inc.<br>(K001210)                                                                                               |
| Coil Design: 2 element saddle transmit<br>receive quadrature coil.<br>Decoupling: This is a transmit and<br>receive coil. There is no decoupling, the<br>Transmit / Receive Switch is actively<br>switched during transmit and receive<br>modes. | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179)<br>- Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179) |
| Prevention of RF Burns: The coil's<br>transmit /receive switch uses pin diodes<br>to isolate the receive channel from the<br>transmit channel; coil elements and<br>circuitry are enclosed in a non-<br>conductive housing.                      | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179)                                                                                                        |
| Radio Frequency Absorption: Power<br>deposition during imaging is limited by<br>SAR algorithm                                                                                                                                                    | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179)                                                                                                        |
| Formation of Resonant Loop: Active<br>diodes and fast RF blowing fuses<br>isolate the coil elements from RF fields;<br>length of cable and stiffness does not<br>permit looping                                                                  | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179)                                                                                                        |

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Christie Shumaker Manager OA and Regulatory USA Instruments, Inc. 1515 Danner Drive AURORA OH 44202

FEB 2 1 2003

Re: K023982 Trade/Device Name: Leo III Tx/Rx Quadrature Knee Coil and Alpha III Tx/Rx Quadrature Wrist Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 MOS Dated: November 27, 2002 Received: December 2, 2002

Dear Ms. Shumaker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Page 1 of 1

510(k) Number (if known):__ <○23982

Device Name: Leo III Tx/Rx Quadrature Knee Coil Alpha III Tx/Rx Quadrature Wrist Coil

Indications for Use: The Leo III Tx/Rx Quadrature Knee Coil is designed to provide Magnetic Resonance Images of the knee and surrounding anatomy. The Alpha III Tx/Rx Quadrature Wrist Coil is designed to provide Magnetic Resonance Images of the wrist and hand. The Leo III Tx/Rx Quadrature Knee Coil and the Alpha III Tx/Rx Quadrature Wrist Coil are designed for use with the Siemens Magnetom Trio 3.0T scanner manufactured by Siemens Medical Solutions.

> Anatomic Regions: Leo III Tx/Rx Quadrature Knee Coil: Knee and surrounding Anatomy Alpha IIII Tx/Rx Quadrature Wrist Coil: Wrist and Hand Nuclei Excited: Hydrogen

The indications for use are the same as for standard imaging:

The 3.0T MRI system is indicated for use as an NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR signal, (2) depend upon NMR parameters (proton density, spin lattice relaxation time T1, spin-spin relaxation time T2) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images vield information that can be useful in the determination of a diagnosis.

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE)        |               |
|---------------------------------------------------------------|---------------|
| Prescription Use (Per 21 CFR 801.109)                         | <span></span> |
|                                                               | OR            |
| Over-The-Counter Use (Optional Format 1-2-96)                 | <span></span> |
| (Division Sign-Off)                                           | <span></span> |
| Division of Reproductive, Abdominal, and Radiological Devices | <span></span> |
| 510(k) Number                                                 | K023982       |

---

**Source:** [https://fda-staging.innolitics.com/device/K023982](https://fda-staging.innolitics.com/device/K023982)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).