VACUETTE TRACE ELEMENTS TUBES

K023971 · Greiner Bio-One Vacuette North America · JKA · Jan 27, 2003 · Clinical Chemistry

Device Facts

Record IDK023971
Device NameVACUETTE TRACE ELEMENTS TUBES
ApplicantGreiner Bio-One Vacuette North America
Product CodeJKA · Clinical Chemistry
Decision DateJan 27, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1675
Device ClassClass 2

Indications for Use

VACUETTE® Tubes, Holders and Needles are used together as a system for the collection of venous blood. The Greiner VACUETTE® Trace Elements tubes with sodium heparin or no additive are used to collect, transport and process blood for testing plasma, serum, or whole blood for trace elements in the clinical laboratory.

Device Story

VACUETTE® Trace Elements Tubes are evacuated blood collection tubes designed for clinical laboratory use. Tubes are composed of clear plastic with royal blue plastic/rubber caps and black inner cap rings; dimensions 13 x 75 mm with 6mL draw. Used with a vacuum tube holder to facilitate venous blood collection. Available in two configurations: with sodium heparin (for plasma/whole blood) or no additive (for serum). Tubes serve as containers for blood transport and processing prior to trace element testing. No electronic or algorithmic components involved.

Clinical Evidence

Bench testing only. Two studies conducted: 1) testing for background trace element presence in tubes using deionized water; 2) comparative performance study using blood samples from 40 donors. Results demonstrated no background trace elements and equivalent performance to the predicate.

Technological Characteristics

Clear plastic tubes; royal blue plastic/rubber caps; black plastic inner cap rings. Dimensions: 13 x 75 mm; 6mL draw. Vacuum-based collection system. No electronic components, software, or energy sources.

Indications for Use

Indicated for the collection, transport, and processing of venous blood samples (plasma, serum, or whole blood) for trace element analysis in clinical laboratory settings.

Regulatory Classification

Identification

A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # JAN 2 7 2003 November 27, 2002 ## 510(K) SUMMARY # Greiner VACUETTE® Trace Elements Tubes Greiner VACUETTE® North America. Inc. P.O Box 1026 Monroe, NC 28111 For information regarding this 510(k) Summary, please contact Greiner VACUETTE® North America, Douglas L. Harris. Device Names: Proprietary Name: VACUETTE® Trace Elements Tubes Common Name: Blood Collection Tubes Classification Name: Tubes, Vials, Systems, Serum Separators, Blood Collection #### Device Description: VACUETTE® Tubes are used to collect, transport and process blood for testing serum, plasma or whole blood in the clinical laboratory. The VACUETTE® Trace Elements tube with sodium heparin may be used to collect a whole blood/plasma sample. The VACUETTE® Trace Elements tube with no additive may be used to collect a serum sample. The tubes are composed of clear plastic. The caps are roval blue and made of plastic and rubber: the inner cap rings are black and made of plastic. The tubes' size is 13 x 75 mm. 6mL draw. The tubes are equipped with a vacuum tube holder to assist in positioning the product when obtaining blood samples. The vacuum tube holder is composed of plastic. ### Intended Use: VACUETTE® Tubes, Holders and Needles are used together as a system for the collection of venous blood. The Greiner VACUETTE® Trace Elements tubes with sodium heparin or no additive are used to collect, transport and process blood for testing plasma, serum, or whole blood for trace elements in the clinical laboratory. ### Substantial Equivalence: The Greiner VACUETTE® Trace Elements tubes are substantially equivalent to the Becton Dickinson Vacutainer® Trace Elements tube with sodium heparin (pre-amendment). The blood collection tubes have the same intended use and contain the same tube material and stopper material; the VACUETTE® Trace Elements tube containing sodium heparin contains the same anticoagulant. Two studies were conducted on the tubes – testing for presence of trace elements in the tubes using deionized water and comparison testing using blood collected from 40 blood donors. Test results for trace elements to be claimed showed no background presence of trace elements in the tubes and equivalent performance with blood samples. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is facing left. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Greiner Bio-One Vacuette® North America c/o Ms. Judi Smith Principal Sienna Partners, L.L.C. P.O. Box 103 Baldwin, MD 21013 JAN 2 7 2003 k023971 Re: Trade/Device Name: Greiner VACUETTE® Trace Elements Tubes Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA Dated: November 27, 2002 Received: November 29, 2002 Dear Ms. Smith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). {2}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours. Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Ko2397/ 510(k) Number (if known): Device Name: Greiner VACUETTE® Trace Elements Evacuated Blood Collection Tubes with sodium heparin or no additive Indications For Use: Greiner VACUETTE® Trace Elements Evacuated Blood Collection Tubes with sodium heparin or no additive are used to collect, transport and process blood for testing plasma, serum, or whole blood for trace elements in the clinical laboratory. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Lan V.A. La De Cooper (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number K093471 Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use_ (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...