GC OIA

K023947 · Thermo Biostar, Inc. · LIR · Apr 17, 2003 · Microbiology

Device Facts

Record IDK023947
Device NameGC OIA
ApplicantThermo Biostar, Inc.
Product CodeLIR · Microbiology
Decision DateApr 17, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3390
Device ClassClass 2

Indications for Use

The Thermo BioStar® GC OIA assay is an Optical ImmunoAssay test for the rapid qualitative detection of gonococcal antigen (L7/L12 ribosomal protein) in female endocervical swab and male urine specimens. Urine specimens must be prepared using an accessory Urine Filtration Device (UFD) for concentration and extraction. This test is intended for in vitro diagnostic use as an aid in identifying the presence of Neisseria gonorrhoeae antigen. The assay is intended for use with symptomatic females and males, in populations at risk for sexually transmitted diseases.

Device Story

GC OIA is an in vitro diagnostic test for rapid qualitative detection of N. gonorrhoeae antigen. Input: female endocervical swabs or male urine (processed via Urine Filtration Device). Principle: Optical ImmunoAssay (OIA) technology; antigen-antibody binding on silicon wafer surface; antibody-enzyme conjugate added to form immune complex sandwich; substrate addition increases molecular thin film thickness (mass enhancement). Output: visual color change (purple spot on gold background) indicating presence/absence of antigen. Used in clinical settings/physician offices; performed by laboratory personnel. Results provide rapid (under 30 minutes) identification of gonococcal antigen to aid clinical diagnosis of symptomatic patients.

Clinical Evidence

Multicenter study (n=904) comparing GC OIA to culture with LCR confirmation. Male urine: 93.2% sensitivity, 97.5% specificity. Female endocervical swabs: 70.7% sensitivity, 99.4% specificity. Reproducibility: 98.1% for urine, 88.9% for swabs.

Technological Characteristics

Optical ImmunoAssay (OIA) technology; silicon wafer optical surface; mass enhancement via antibody-enzyme conjugate and substrate; visual colorimetric readout; qualitative detection of L7/L12 ribosomal protein antigen.

Indications for Use

Indicated for symptomatic males and females at risk for sexually transmitted diseases to detect Neisseria gonorrhoeae antigen (L7/L12 ribosomal protein) in female endocervical swabs and male urine specimens.

Regulatory Classification

Identification

Neisseria spp. direct serological test reagents are devices that consist of antigens and antisera used in serological tests to identify Neisseria spp. from cultured isolates. Additionally, some of these reagents consist of Neisseria spp. antisera conjugated with a fluorescent dye (immunofluorescent reagents) which may be used to detect the presence of Neisseria spp. directly from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genus Neisseria, such as epidemic cerebrospinal meningitis, meningococcal disease, and gonorrhea, and also provides epidemiological information on diseases caused by these microorganisms. The device does not include products for the detection of gonorrhea in humans by indirect methods, such as detection of antibodies or of oxidase produced by gonococcal organisms.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 510(k) SUMMARY (page 1 of 3) 8.0 APR 1 7 ZUU3 # 510(k) SUMMARY This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: K023947 A. Submitter's information required per [§807.92(a)(1)]: - SUBMITTER'S NAME: Thermo BioStar, Inc. . - ADDRESS: 331 South 104th Street, Louisville CO 800271 ● - TELEPHONE: (303) 530-3888 ext. 612 - FAX: (303) 581-6405 - CONTACT PERSON: John G. Adams . - DATE 510(k) SUMMARY PREPARED: April, 2003 ● B. Device information required per [§807.92(a)(2)]: - TRADE OR PROPRIETARY NAME: GC OIA® - COMMON NAME: Neisseria gonorrhoeae antigen assay - CLASSIFICATION NAME: Antigen, Enzyme Linked Immunoabsorbent Assay, Neisseria gonorrhoeae ● - C. Identification of legally marketed device to which we are comparing performance. Historical Reference Method - Neisseria gonorrhoeae culture . Device Technology: | Trade or Proprietary Name: | Chlamydia OIA Assay | |----------------------------|---------------------| | Regulatory Class: | I | | Manufacturer: | Thermo BioStar | | 510(k) Number: | K951010 | - D. Intended use of device [§807.92(a)(5)]: The Thermo BioStar® GC OIA assay is an Optical ImmunoAssay test for the rapid qualitative detection of gonococcal antigen (L7/L12 ribosomal protein) in female endocervical swab and male urine specimens. Urine specimens must be prepared using an accessory Urine Filtration Device (UFD) for concentration and extraction. This test is intended for in vitro diagnostic use as an aid in identifying the presence of Neisseria gonorrhoeae antigen. The assay is intended for use with symptomatic females and males, in populations at risk for sexually transmitted diseases. - E. Description of device [§807.92(a)(4)]: Principle of the Test: The GC OIA test involves the qualitative extraction and detection of an antigen unique to N. gonorrhoeae. The Optical ImmunoAssay technology enables the direct visual detection of a physical change in the optical thickness of molecular thin films. This change is a result of antigen-antibody binding on an optical surface (silicon wafer). When an extracted specimen is placed directly on the optical surface, the immobilized specific antibodies capture the antibody-enzyme conjugate is then added for form an immune complex "sandwich" of immobilized antibody-sample-antibody HRP on the surface. After washing, the {1}------------------------------------------------ ### 8.0 510(k) SUMMARY (page 2 of 3) substrate is added, increasing the thickness (mass enhancement) of the molecular thin film. This change in thickness alters the reflected light path and is visually perceived as a color change. Slight changes in optical thickness produce a distinct, visible color change. A positive result appears as a purple spot on the predominant gold background. When antigen is not present in the specimen, no binding takes place. Therefore, the optical thickness remains unchanged and the surface retains the original gold color indicating a negative result. #### Device Comparison [§807.92(a)(6)]: F. ### Comparison to Historical reference method The GC OIA assay is similar to culture methods in that: - Both assays are used to detect and identify N. gonorrhoeae ; ● - . Both assays detect N. gonorrhoeae in endocervical swabs and male urine specimens. The GC OIA assay differs from traditional culture methods in that: - GC OIA assay detects an antigen unique to the N. gonorrhoeae organism, while the traditional . culture methods detect the whole living organism; - The GC OIA assay can provide results in less than 30 minutes, in contrast to culture methods that ● can take up to 72 hours. ### Comparison to existing device technology The GC OIA assay is similar to the Chlamydia OIA assay in that: - Both assays have the same Optical ImmunoAssay (OIA) technology; ● - . Both assays detect antigen in endocervical swab specimens; - . The intended use for both assays is to evaluate symptomatic patients for a sexually transmitted disease infection; - Both assays are qualitative; - . Both assays utilize liquid reagents The GC OIA assay differs with the Chlamydia assay in that: - GC OIA is testing for the presence of a different STD, Neisseria gonorrhoeae; - . GC OIA detects antigens in symptomatic male urine specimens, in addition to endocervical swabs ## SUMMARY OF PERFORMANCE DATA: #### CLINICAL STUDIES Performance characteristics for the GC OIA assay were initially established in a multicenter study at four geographically diverse clinical sites. - G. Summary of clinical testing [§807.92(b)(2)]: #### Reproducibility Urine: Reproducibility testing was conducted at four clinical sites and two point of care settings including physician offices. Urine testing was performed with 2 mL samples on three days with nine blinded specimens each day. This testing resulted in an overall reproducibility of 98.1% (95%CI = 93.8-99.3). Swab: Reproducibility testing was conducted at three physician office laboratories. Swab testing was performed on three days with nine blinded samples each day. The negative swabs contained no GC cells. Overall reproducibility of the testing using the swab panels was 88.9% at the three sites. (95% CI = 93.8-99.3) {2}------------------------------------------------ ### 8.0 510(k) SUMMARY (page 3 of 3) ### Clinical Sensitivity and Specificity A study comparing the GC OLA test to commercial culture media, with secondary confirmation testing by LCR (Ligase Chain Reaction). A total of 904 valid results were obtained from symptomatic patient specimens collected at four clinical trial locations. Sensitivity and specificity for symptomatic male unnes was 93.2 % and 97.5 % respectively. Sensitivity and specificity for symptomatic female endocervical swabs was 70.7% and 99.4% respectively. Overall PPV for males were 94.0 % and 97.2 % respectively; and for females were 90.6% and 97.5% respectively. - H. Conclusions from nonclinical / clinical testing [§807.92(b)(3)]: The results of the above described internal and external studies demonstrate that the GC OIA test is as safe and effective as the comparative device. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES U.S.A." arranged in a circular pattern around the symbol. The logo is black and white. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 APR 1 7 2003 Mr. John G. Adams Regulatory Affairs Manager Thermo BioStar, Inc. 331 South 104th Street Louisville, CO 80027 k023947 Re: Trade/Device Name: GC OIA® Test Kit Regulation Number: 21 CFR 866.3390 Regulation Name: Neisseria spp. Direct Serological Test Reagents Regulatory Class: Class II Product Code: LIR Dated: February 18, 2003 Received: February 19, 2003 Dear Mr. Adams: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {4}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### INDICATIONS FOR USE STATEMENT 12.0 # INDICATIONS FOR USE STATEMENT 510(k) Number (if known): K 023947 GC OIA® Device Name: Indications For Use: The Thermo BioStar® GC OIA assay is an Optical ImmunoAssay test for the rapid qualitative detection of gonococcal antigen (L7/L12 ribosomal protein) in female endocervical swab and male urine specimens. Urine specimens must be prepared using an accessory Urine Filtration Device (UFD) for concentration and extraction. This test is intended for in vitro diagnostic use as an aid in identifying the presence of Neisseria gonorrhoeae antigen. The assay is intended for use with symptomatic females and males, in populations at risk for sexually transmitted diseases. # (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) ### Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use the text is the following: Division of Clinical Laboratory Devices 510(k) Number K023947
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...