K023890 · Appro Healthcare, Inc. · MRZ · Feb 7, 2003 · General Hospital
Device Facts
Record ID
K023890
Device Name
PREPARE CARTRIDGE
Applicant
Appro Healthcare, Inc.
Product Code
MRZ · General Hospital
Decision Date
Feb 7, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Prepare™ Cartridge is used as a drug reservoir and fluid path for administering intravenous, intra-arterial, subcutaneous, epidural, and subarachnoid drugs to a patient's access site using specified infusion pumps and extension sets. For intravenous, intra-arterial, subcutaneous, epidural and subarachnoid administration Prepare™ Cartridge must be used only with SIMS Deltec CADD®-1, CADD®-Plus, and CADD®-PCA pumps for the infusion of various medicinal solutions. Prepare™ Cartridge is indicated for use in acute and alternate care settings. Alternate care includes, but is not limited to, infusion clinics, nursing homes, and home healthcare.
Device Story
Prepare™ Cartridge serves as a 100ml drug reservoir and fluid path for infusion therapy. It interfaces with SIMS Deltec CADD®-1, CADD®-Plus, and CADD®-PCA pumps to deliver medicinal solutions to a patient's access site. Used in acute and alternate care settings, the device is operated by healthcare providers or patients/caregivers in home settings. It functions as a mechanical accessory to the infusion pump, ensuring a secure pathway for drug delivery. By providing a compatible reservoir, it enables the administration of various intravenous, intra-arterial, subcutaneous, epidural, and subarachnoid medications, facilitating therapy in diverse clinical environments.
Clinical Evidence
Bench testing only. Performance testing compared capacity, delivery accuracy, volume per pump cycle, and pump fit against the predicate. Biocompatibility testing was performed per ISO-10993-1, confirming the device is non-irritating, non-cytotoxic, non-toxic, non-hemolytic, and non-sensitizing.
Technological Characteristics
100ml reservoir capacity; fluid path interface for SIMS Deltec CADD® pumps; ethylene oxide gas sterilization; biocompatibility compliant with ISO-10993-1.
Indications for Use
Indicated for patients requiring intravenous, intra-arterial, subcutaneous, epidural, or subarachnoid drug administration in acute or alternate care settings (e.g., infusion clinics, nursing homes, home healthcare) using compatible SIMS Deltec CADD® pumps.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
CADD® 1 Medication Cassette Reservoir
Submission Summary (Full Text)
{0}------------------------------------------------
K023890
、
# 510(k) SUMMARY
| Date: | November 19, 2002<br>FEB 07 2003 |
|-------------------------|------------------------------------------------------------------------------------------------|
| Manufacturing Facility: | APPRO Healthcare, Inc.<br>847 Main Street<br>Buffalo, NY 14203 |
| Telephone: | (716) 855-1068 |
| Contact Person: | John R. Semler<br>Vice President, RD&E<br>Extension 309<br>Email: jrsemler@approhealthcare.com |
| Device Trade Name: | Prepare™ Cartridge |
| Device Common Name: | Accessory to an Infusion Pump |
| Classification Name: | Accessories, Pump, Infusion |
| Regulatory Reference: | MRZ |
| Predicate Device | CADD® 1 Medication Cassette Reservoir |
Registered trademark of SIMS Deltec, Inc.
l
{1}------------------------------------------------
## Description:
Prepare™ Cartridge is designed for use with Deltec CADD®-1, CADD®-Plus, and CADD®-PCA pumps. The device provides a reservoir for holding up to 100ml of prepared drug and interfaces with various models of SIMS Deltec infusion pumps to provide a pathway from the reservoir to the patient's access site.
## Intended Use / Indications for Use:
Prepare™ Cartridge is used as a drug reservoir and fluid path for administering intravenous, intra-arterial, subcutaneous, epidural, and subarachnoid drugs to a patient's access site using specified infusion pumps and extension sets. For intravenous, intra-arterial, subcutaneous, epidural and subarachnoid administration
Prepare™ Cartridge must be used only with SIMS Deltec CADD®-1, CADD®-Plus, and CADD®-PCA pumps for the infusion of various medicinal solutions. Prepare™ Cartridge is indicated for use in acute and alternate care settings. Alternate care includes, but is not limited to, infusion clinics, nursing homes, and home healthcare.
#### Physical/Technical Comparison
Prepare™ Cartridge can be used in place of Medication Cassette Reservoir. Physical and technical characteristics, including materials used in construction, size, intended use, and ability to interface with specified CADD® pumps are comparable.
#### Performance Summary:
The device and the predicate were subjected to various bench tests to demonstrate comparable performance characteristics. Prepare™ Cartridge capacity, accuracy of delivery, volume delivered per pump cycle, and fit to pump are substantially equivalent to the CADD® Administration Set.
#### Biocompatibility Testing:
Prepare™ Cartridge was subjected to biocompatibility testing as recommended by ISO-10993-1, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing. The device is certified as non-irritating, non-cytotoxic, non-toxic, non-hemolytic, and nonsensitizing.
#### Sterilitv:
Prepare™ Cartridge is sterilized by ethylene oxide gas in a validated sterilization process.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three stylized human profiles facing to the right.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 0 7 2003
Mr. John R. Semler Vice President, RD & E APPRO Healthcare Incorporated 847 Main Street Buffalo, New York 14203
Re: K023890
Trade/Device Name: Prepare™ Cartridge Regulation Number: 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: MRZ Dated: November 19, 2002 Received: November 21, 2002
Dear Mr. Semler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Semler
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susan Russer
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
| 510(k) Number, if known: | Not yet assigned M023890 |
|--------------------------|---------------------------------|
| Device Name: | Prepare <sup>TM</sup> Cartridge |
Prepare™ Cartridge is used as a drug reservoir and fluid path for administering intravenous, intra-arterial, subcutaneous, epidural, and subarachnoid drugs to a patient's access site using specified infusion pumps and extension sets.
For intravenous, intra-arterial, subcutaneous, epidural and subarachnoid administration
Prepare™ Cartridge must be used only with SIMS Deltec CADD®-1, CADD®-Plus, and CADD®-PCA pumps for the infusion of various medicinal solutions. Prepare™ Cartridge is indicated for use in acute and alternate care settings. Alternate care includes, but is not limited to, infusion clinics, nursing homes, and home healthcare.
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | / |
|------------------|---|
|------------------|---|
OR
Over-the-Counter Use
(Per 21CFR 801,109)
Patina Cucente
(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental D
510(k) Number: _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.