NESSY Ω™ (Omega) disposable split return electrodes are intended for dispersing and monitoring HF current in electrosurgical interventions.
Device Story
NESSY Ω™ (Omega) is a disposable, split-design patient return electrode used during electrosurgical procedures. The device consists of a flexible pad featuring a conductive aluminum layer, hydrogel, and a Scantape carrier with acrylic adhesive. It is applied to the patient's skin and connected to an electrosurgical unit (ESU) via a cable. The device's circular, equipotential omega-shaped design promotes uniform distribution of HF current, reducing pad warming and eliminating the need for specific orientation relative to the operative site. During surgery, the ESU uses the split electrode to monitor current flow; if the generator detects discrepancies, it alerts the clinician. The device is used in clinical settings by healthcare providers following standard AORN guidelines for site preparation. It benefits patients by safely dispersing HF current, thereby preventing thermal injury at the return electrode site.
Clinical Evidence
Bench testing only. Thermographic testing demonstrated the omega-shaped design provides more uniform HF current distribution compared to the predicate. Biocompatibility testing was conducted according to EN ISO 10993-1. No clinical data provided.
Indicated for dispersing and monitoring high-frequency (HF) current during electrosurgical interventions in both pediatric and adult patient populations.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
NOV 2 9 2002
| | 510K SUMMARY |
|--------------------------|--------------------------------------------------------------|
| | K023812 |
| Submitted By: | ERBE USA, Inc. |
| | 2225 Northwest Parkway |
| | Marietta, GA 30067 |
| | Tel: 770-955-4400 |
| | Fax: 770-955-2577 |
| Contact Person: | John Tartal |
| Date Prepared: | 11/14/02 |
| Common Name: | Patient Plate/Pad, Dispersive Electrode, or Return Electrode |
| Trade/Proprietary Name: | NESSY Ω <sup>TH</sup> (OMEGA) Disposable split Return Ele |
| Classification Name: | Electrosurgical cutting and coagulation device and |
| | accessories (21CFR878.4400) |
| Product Code: | 79GEI |
| Legally Marketed Device: | ERBE Disposable Patient Return Electrodes, 510(k) |
| | Number: K972269 |
#### Device Description:
There are two models of the NESSY Ω™ (Omega) Disposable Split Return Electrode, One with a connecting cable (P/N: 20193-084) and the other (P/N: 20193-085) without the cable {Note: A reusable patient return electrode cable submitted in a previous ERBE 510(k) [K960569] is used with this plate/pad}. The cable of the P/N: 20193-084 is made of Polyvinvlcholoride (PVC). Copper Wiring, and Polyethylene (PE) Foam. It is flexible, 4 m (13.2 ft.) long, and has a standard connection for fitting into an ESU. The pad portion of each is made of Siliconized Polyethylene Terephthalate (PET). Scantape coated with Acrylic Adhesive, Aluminum, and Hydrogel. There is a protective film cover to keep the pad from drying out prior to being used. The Return Electrode has a peel off tab that makes handling it easier. The plate also is flexible and molds well to the skin. The relatively round external shape and geometrical internal omega design (i.e., the equipotential ring) of the pad more uniformly distributes the High Frequency (HF) current. This was demonstrated through thermographic testing. The design also eliminates the need for orientating the pad to the operative site (i.e., there is no long side of the pad). The hydrogel layer overlaps all the aluminum edges, which reduces warming of the pad during use. The Patient Plate is suitable for use on children and adults. The Electrodes are each packaged in an aluminum foil pouch with a 2-year expiration date. Also, they are provided non-sterile and are disposable.
{1}------------------------------------------------
# 510K SUMMAR
Prior to electorsurgery, the NESSY Ω™ (Omega) Disposable Split Return Electrode is placed on the patients skin. AORN (Association of Operating Room Nurses) guidelines are to be followed for site selection, preparation of the site, etc. The Patient Plate is then attached via a connecting cable to an Electrosurgical Unit (ESU) or Generator. When the ESU is activated, the pad safely returns (disperses) the HF current. If the Generator has a safety monitoring system, the ESU measures the current going into and exiting the pad. In this manner, the Generator monitors the HF current and alerts the clinician to any differences.
#### Intended Use:
NESSY Ω™ (Omega) Disposable Split Return Electrodes are intended for dispersing and monitoring HF current in electrosurgical interventions.
### Similarities and Differences of the Modified Device to the Current Device (Predicate Comparison/Substantial Equivalence):
#### Similarities
The modified Split Return Electrodes have intended use, performance expectations (i.e., safely dispersing/returning delivered HF current and, with the previous split return models, monitoring the current), packaging, and associated labeling as the predicate. As with the predicate, the Patient Plate/Pad comes with or without a connecting cable. Also like the predicate, they are provided non-sterile with the same expiration dating and are disposable.
Finally, the contract manufacturer for the Return Electrodes is the same (Leonhard Lang GmbH).
#### Differences
The material changes in the modified Split Return Electrode basically involve the pad's carrier being made of Scantape with Acrylic Adhesive (Note: The fabric in band aids.) which molds better to the body and an improved Hydrogel (NH-02). The materials have been found acceptable through biocompatibility testing. Tests were selected based upon EN ISO 10993-1, "Biological Evaluation of Medical Devices".
For the modified Electrode with a cable, the cord is 1 m (3.3 ft.) longer than the predicate devices that have a cable [i.e., the predicate device cable is 3 m (10 ft.) long in comparison to 4 m (13 ft.) for the modified product that has a cablel. The extra length provides the clinician with a little more latitude in regards to the position of the patient and equipment location.
The pad of the modified Split Return Electrodes, are designed with an equipotential ring. The round/omega design of the conductive films distributes HF current more evenly. As a result, only one size pad (i.e., contact surface) is needed. Also, due to the symmetrical design of the plate, there is no long side. Therefore, no orientation of the pad to the operative site is needed.
{2}------------------------------------------------
### 510K SUMMARY
## K023812
All the changes have been verified or validated in design control. Applicable sections of AAMI/ANSI HF-18, "Electrosurgical Devices"; EN/IEC 60601-1, "Medical Electrical Equipment, General Requirements for Safety, and ENVIEC 60601-2-2, "Medical Electrical Equipment, Particular Requirements for the Safety of High Frequency Surgical Equipment" standards were applied and met for the modified Split Return Electrodes.
#### Conclusion:
The modified Split Return Electrodes have the same intended use and principles of operation as the Split Return Electrodes in the previously cleared predicate device. Technologically the modified electrode's round/omega design of the pad's conductive films disperses HF current more evenly. As a result, the NESSY Ω™ (Omega) Disposable Split Return Electrodes are as safe and effective then the predicate device.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.
Food and Drug Administratio 9200 Corporate Boulevard Rockville MD 20850
# 9 9 2002
Erbe USA. Inc. John Tartal QA/RA. Manager 2225 Northwest Parkway Marietta, Georgia 30067
Re: K023812
Trade/Device Name: Nessy Ώ™ (Omega) Disposable Split Return Electrodes Regulation Number: 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: November 14, 2002 Received: November 15, 2002
Dear Mr. Tartal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the
{4}------------------------------------------------
Page 2 - Mr. John Tartal
quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
iriam C. Provost
{0' Celia M. Witten, Ph.D., MD Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) NUMBER (IF KNOWN): K023812
DEVICE NAME: NESSY ΩΤΜ (Omega) Disposable Split Return Electrodes
INDICATIONS FOR USE:
NESSY Ω™ (Omega) disposable split return electrodes are intended for dispersing and monitoring HF current in electrosurgical interventions.
Muriam C. Provost
Division Sign-Off) Division of General, Restorative and Neurological Devices
K023812
(PLEASE DO NOT MRITE BELOW THIS LINE-CONTINUE ON ANOTHER E IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODD)
Prescription Use (Per 21 CER 001.109)
OR
Over-The-Counter-(Optional Forma
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.