PROPOXYPHENE ENZYME IMMUNOASSAY, CATALOG #0120 (500 TEST KIT), #0121 (5000 TEST KIT)

K023795 · Lin-Zhi International, Inc. · JXN · Jan 21, 2003 · Clinical Toxicology

Device Facts

Record IDK023795
Device NamePROPOXYPHENE ENZYME IMMUNOASSAY, CATALOG #0120 (500 TEST KIT), #0121 (5000 TEST KIT)
ApplicantLin-Zhi International, Inc.
Product CodeJXN · Clinical Toxicology
Decision DateJan 21, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3700
Device ClassClass 2

Indications for Use

The Propoxyphene Enzyme Immunoassay is a homogeneous enzyme immunoassay with a 300 ng/mL cutoff. The assay is intended for use in the qualitative and semi-quantitative analyses of propoxyphene in human urine. The assay is designed for professional use with a number of automated clinical chemistry analyzers. The Propoxyphene Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgement should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

Device Story

Ready-to-use liquid reagent homogeneous enzyme immunoassay; detects propoxyphene in human urine. Principle: competition between G6PDH-labeled propoxyphene and free drug in sample for specific antibody binding sites; absence of drug allows antibody binding to labeled enzyme, decreasing enzyme activity. Activity measured spectrophotometrically at 340 nm via conversion of NAD to NADH. Used in clinical laboratories with automated chemistry analyzers; operated by laboratory technicians/professionals. Provides preliminary qualitative/semi-quantitative results; requires GC/MS confirmation. Assists clinicians in drug-of-abuse screening.

Clinical Evidence

Bench testing only. Evaluated precision (within-run and run-to-run), sensitivity (7.5 ng/mL), accuracy, analytical recovery, and specificity. Accuracy demonstrated 100% agreement with GC/MS for positive samples. Semi-quantitative recovery within ±10% of nominal concentration between 30-900 ng/mL.

Technological Characteristics

Homogeneous enzyme immunoassay; G6PDH enzyme label; spectrophotometric detection at 340 nm. Liquid reagent format. Designed for use on automated clinical chemistry analyzers. No specific materials or software architecture details provided.

Indications for Use

Indicated for qualitative and semi-quantitative analysis of propoxyphene in human urine for professional use in clinical settings. Provides preliminary analytical results requiring confirmation by GC/MS.

Regulatory Classification

Identification

A propoxyphene test system is a device intended to measure propoxyphene, a pain-relieving drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of propoxyphene use or overdose or in monitoring levels of propoxyphene to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). A propoxyphene test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K023795 JAN 21 2003 ## 510(k) Summary of Safety and Effectiveness This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. ### Introduction According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. #### Submitter name, Address, and Contact Lin-Zhi International, Inc. 2391 Zanker Road, Suite 340 San Jose, CA 95131-1124 Phone: (408) 944-0360 (408) 944-0359 Fax: Chiu Chin Chang, Ph.D. Contact: VP, R&D ### Device Name and Classification | Classification Name: | Enzyme Immunoassay, Propoxyphene, Class II, JXN (91 Toxicology), 21CFR 862.3700 | |----------------------|--------------------------------------------------------------------------------------| | Common Name: | Homogeneous enzyme immunoassay for the determination of propoxyphene level in urine. | | Proprietary Name: | None | ### Legally Marketed Predicate Device(s) Lin-Zhi International, Inc.' Propoxyphene Enzyme Immunoassay is substantially equivalent to the Propoxyphene Enzyme Immunoassay (By DRI/Microgenics Corp.), cleared under premarket notification K943414. LZI's Propoxyphene Enzyme Immunoassay is identical or similar to its predicate in terms of intended use, method principle, device components, and clinical performance. ### Device Description LZI's Propoxyphene Enzyme Immunoassay is a ready-to-use, liquid reagent, homogeneous enzyme immunoassay. The assay uses specific antibody that can detect propoxyphene in human urine with minimal cross-reactivity to various, common prescription drugs and abused drugs. {1}------------------------------------------------ The assay is based on competition between propoxyphene labeled with glucose-6-phosphate dehydrogenase (G6PDH) enzyme, and free drug from the urine sample for a fixed amount of specific antibody. In the absence of free drug from the urine sample the specific antibody binds to the drug labeled with G6PDH enzyme causing a decrease in enzyme activity. The G6PDH enzyme activity is determined spectrophotometrically at 340 nm by measuring its ability to convert nicotinamide adenine dinucleotide (NAD) to NADH. ### Intended Use The Propoxyphene Enzyme Immunoassay is a homogeneous enzyme immunoassay with a 300 ng/mL cutoff. The assay is intended for use in the qualitative and semi-quantitative analyses of propoxyphene in human urine. ## Comparison to Predicate Device LZI's Propoxyphene Enzyme Immunoassay is substantially equivalent to other products in commercially distribution intended for similar use. Most notably it is substantially equivalent to the currently, commercially marketed Propoxyphene Enzyme Immunoassay (K943414) by Diagnostic Reagents, Inc. (DRI, now Microgenics Corporation) The following table compares LZI's Propoxyphene Enzyme Immunoassay with the predicate device, DRI's Propoxyphene Enzyme Immunoassay: Similarities: - Both assays are for qualitative and semi-quantitative determination of . propoxyphene in human urine. - Both assays use the same method principle, and device components. . - Both assays use 300 ng/mL as cutoff level. . Differences: - · LZI's Propoxyphene Enzyme Immunoassay uses 5 calibrators for the semiquantitative analysis of propoxyphene concentration in urine. DRI's Propoxyphene EIA used 3 calibrators previously. A total of 5 calibrators are available now from DRI in the "Multi-drug Urine Calibrators and Controls" product. {2}------------------------------------------------ # (Comparison to Predicate Device, continued) ## Performance Characteristics | Feature | DRI's Propoxyphene EIA | | | LZI's Propoxyphene EIA | | | | | |-----------------------|--------------------------------------------------------------------|-----|------|-------------------------------------------------------------------------------------------|------------|-------|------|------| | Within Run Precision: | | | | | | | | | | Qualitative: | Mean Rate | SD | % CV | Mean Rate | SD | % CV | | | | | Negative | 153 | - | 1.1 | Negative | 117.4 | 0.5 | 0.47 | | | 225 ng/mL | - | - | - | 225 ng/mL | 225.1 | 1.3 | 0.59 | | | 300 ng/mL | 265 | - | 1.0 | 300 ng/mL | 261.3 | 1.6 | 0.61 | | | 375 ng/mL | - | - | - | 375 ng/mL | 287.7 | 1.5 | 0.51 | | | 1000 ng/mL | 322 | - | 0.9 | 1000 ng/mL | 350.0 | 1.4 | 0.39 | | Semi-quantitative: | No data available. | | | | Mean Conc. | SD | % CV | | | | | | | 225 ng/mL | 231.3 | 3.1 | 1.34 | | | | | | | 300 ng/mL | 299.6 | 5.8 | 1.92 | | | | | | | 375 ng/mL | 379.7 | 5.6 | 1.46 | | | Run-To-Run Precision: | | | | | | | | | | Qualitative: | Mean Rate | SD | % CV | Mean Rate | SD | % CV | | | | | Negative | 154 | - | 1.0 | Negative | 116.8 | 1.0 | 0.88 | | | 225 ng/mL | - | - | - | 225 ng/mL | 220.8 | 2.4 | 1.07 | | | 300 ng/mL | 265 | - | 0.8 | 300 ng/mL | 255.9 | 2.1 | 0.81 | | | 375 ng/mL | - | - | - | 375 ng/mL | 285.1 | 2.2 | 0.76 | | | 1000 ng/mL | 324 | - | 0.9 | 1000 ng/mL | 349.5 | 1.9 | 0.55 | | Semi-quantitative: | No data available. | | | | Mean Conc. | SD | % CV | | | | | | | 225 ng/mL | 232.6 | 3.0 | 1.27 | | | | | | | 300 ng/mL | 298.7 | 4.7 | 1.56 | | | | | | | 375 ng/mL | 378.0 | 7.4 | 1.97 | | | Sensitivity: | 15 ng/mL | | | 7.5 ng/mL | | | | | | Accuracy: | Vs. a commercial EIA: | | | Vs. GC/MS | | | | | | | 116/126 in agreement | | | 100 % agreement | | | | | | | All 57 positive samples by both assays<br>were confirmed by GC/MS. | | | 100 % agreement | | | | | | Analytical Recovery: | | | | | | | | | | | Qualitative: No data available | | | 100 % accuracy on positive vs. negative tests | | | | | | | Semi-quantitative: No data available | | | Quantitate within ±10% of the nominal<br>concentration between 30 ng/mL and 900<br>ng/mL. | | | | | | | | | | Average 104.9 % recovery at 225 ng/mL<br>level (Cutoff -25%) | | | | | | | | | | Average 103.8 % recovery at 375 ng/mL<br>level (Cutoff + 25%) | | | | | | Specificity: | See attached DRI's Propoxyphene EIA<br>package insert | | | Comparable to the predicate device. | | | | | {3}------------------------------------------------ ### Conclusion LZI's Propoxyphene Enzyme Immunoassay was evaluated for several performance characteristics including precision, sensitivity, accuracy, analytical recovery, and specificity. All the studies showed acceptable results when compared to the predicate device. We trust the information provided in this Premarket Notification [510(k)] submission will support a determination of substantial equivalence of the LZI's Propoxyphene Enzyme Immunoassay to other propoxyphene test systems currently marketed in the United States. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the bird symbol. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JAN 2 1 2003 Chiu Chin Chang, Ph.D. VP. R&D Lin-Zhi International, Inc. 687 North Pastoria Avenue Sunnyvale, CA 94085-2917 Re: k023795 Trade/Device Name: Propoxyphene Enzyme Immunoassay Regulation Number: 21 CFR 862.3700 Regulation Name: Propoxyphene test system Regulatory Class: Class II Product Code: JXN Dated: November 11, 2002 Received: November 13, 2002 Dear Dr. Chang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {5}------------------------------------------------ Page 2 – This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## Premarket Notification ## Indications for Use Statement 510(k) Number (if known): ## Device Name: Propoxyphene Enzyme Immunoassay ### Indications for Use: The Propoxyphene Enzyme Immunoassay is a homogeneous enzyme immunoassay with a 300 ng/mL cutoff. The assay is intended for use in the qualitative and semi-quantitative analyses of propoxyphene in human urine. The assay is designed for professional use with a number of automated clinical chemistry analyzers. The Propoxyphene Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgement should be applied to any drug-ofabuse test result, particularly when preliminary positive results are used. Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use Sean Cooper (Optional Format 1Quession Sign-C visión of Clinical I 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...