UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM, MODEL 973100
K023764 · Beckman Coulter, Inc. · JJE · Jan 28, 2003 · Clinical Chemistry
Device Facts
Record ID
K023764
Device Name
UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM, MODEL 973100
Applicant
Beckman Coulter, Inc.
Product Code
JJE · Clinical Chemistry
Decision Date
Jan 28, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.2160
Device Class
Class 1
Indications for Use
The UniCel Dxl 800 Access Immunoassay System is a microcomputer controlled, random and continuous access analyzer that includes an external computer. This computer stores the system user interface (UI) software and allows the operator to interface with and direct the instrument software. The UniCel Dxl 800 System uses enzyme immunoassays (utilizing paramagnetic particle solid phase and chemiluminescent detection) for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. The UniCel Dxl 800 System is an in vitro diagnostic device for use in the clinical laboratory.
Device Story
Floor-model, microcomputer-controlled, random/continuous access analyzer; performs enzyme immunoassays (EIA) using paramagnetic particle solid phase and chemiluminescent detection. Input: human body fluid samples (serum, plasma, urine). Operation: automated reagent pipetting, fluid handling, incubation, and signal detection. Output: quantitative/qualitative analyte concentrations. Used in clinical laboratories by trained personnel. Modifications from predicate (Access 2) include increased throughput, on-board process monitoring, improved vessel handling, and refrigerated on-board sample aliquot storage. System utilizes external computer for UI and instrument direction. Healthcare providers use results for clinical decision-making regarding diagnosis and treatment of various diseases.
Clinical Evidence
Bench testing only. Method comparison, precision, and analytical sensitivity studies conducted using five representative assays. Studies demonstrated good correlation between the UniCel Dxl 800 and the Access 2 predicate system. All five assays met established acceptance criteria for precision and analytical sensitivity.
Technological Characteristics
Floor-model, microcomputer-controlled analyzer. Uses paramagnetic particle solid phase and chemiluminescent detection. Features include refrigerated on-board sample aliquot storage, automated vessel handling, and sensors/encoders for process monitoring. Connectivity via external computer for UI and instrument control. No changes to reagent chemistry, incubation, or signal detection compared to predicate.
Indications for Use
Indicated for in vitro diagnostic use in clinical laboratories to perform quantitative, semi-quantitative, or qualitative determination of various analytes in human body fluids (serum, plasma, urine) to aid in the diagnosis, treatment, and monitoring of various clinical conditions, including thyroid disease, cardiac muscle damage, and reproductive status.
Regulatory Classification
Identification
A discrete photometric chemistry analyzer for clinical use is a device intended to duplicate manual analytical procedures by performing automatically various steps such as pipetting, preparing filtrates, heating, and measuring color intensity. This device is intended for use in conjunction with certain materials to measure a variety of analytes. Different models of the device incorporate various instrumentation such as micro analysis apparatus, double beam, single, or dual channel photometers, and bichromatic 2-wavelength photometers. Some models of the device may include reagent-containing components that may also serve as reaction units.
{0}------------------------------------------------
JAN 2 8 2003
Confidential
#### 9 510(K) SUMMARY
(Page 1 of 2)
#### 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
510(k) Number _
Date Prepared: November 8, 2002
| Submitter | Contact Person |
|---------------------------|-------------------------------------|
| Beckman Coulter, Inc | Lynn S. Weist |
| Diagnostics Division | Staff Regulatory Affairs Specialist |
| 1000 Lake Hazeltine Drive | Phone: 952-368-1271 |
| Chaska, MN 55318 | Fax: 952-368-7610 |
#### General Information
| Trade Name | UniCel™ DxI 800 Access® Immunoassay System |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Discrete photometric chemistry analyzer for clinical use |
| Classification Name | Discrete photometric chemistry analyzer for clinical use |
| Legally Marketed (Predicate)<br>Device | Access® 2 Immunoassay System, Manufactured by<br>Beckman Coulter, Inc., (K922823 "Add to File" Letter,<br>dated May 18, 2001 and PMA Supplement<br>P980041/S001, cleared October 1, 2002) |
## Device Description
The UniCel Dxl 800 Access Immunoassay System is a floor model, microcomputer controlled, random and continuous access analyzer that performs enzyme immunoassays (EIA) utilizing a paramagnetic particle solid phase and chemiluminescent detection.
#### Intended Use
The UniCel Dxl 800 Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.
#### Substantial Equivalence Comparison
The UniCel Dxl 800 system represents an evolutionary change to the Access 2 Immunoassay System design. The modifications to the basic Access 2 platform support at least three major improvements desirable to the customer with medium-high testing volume. These are: 1) increased throughput,
{1}------------------------------------------------
(Page 2 of 2)
2) on-board process monitoring and 3) release of the primary sample tubes back to the laboratory workflow as early as possible.
The main modifications to the design include an obstruction feature, larger volume bulk supplies to accommodate higher throughput, the method of vessel handling, incorporation of sensors and encoders at key points for process monitoring and the addition of a sample pipettor and refrigerated on-board sample aliquot storage.
Reagent pipetting, fluid handling, reagent chemistry, assay incubation temperature and timing, signal detection and assay data processing are all unchanged from the Access 2 design. The Dxl uses the same Access Immunoassay System assays and calibrators, packaged the same as for Access 2. The formulations of the substrate and wash buffer used with the Access assays are unchanged. There are no changes to the intended uses, technical specifications or final performance specifications and claims for the assays.
## Supporting Data
In order to demonstrate that the Dxl system is substantially equivalent to the Access 2, method comparison, precision and analytical sensitivity studies were conducted using five representative assays. Method correlation studies demonstrated good correlation between the Access 2 and Dxl systems on each of the five assays. Each of the five assays met its established acceptance criteria for precision and analytical sensitivity.
## Conclusion
The information provided in this submission supports a substantial equivalence determination, and therefore 510(k) premarket notification clearance of the UniCel Dxl 800 Access Immunoassay System.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three stylized human profiles.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Lynn S. Weist Staff Regulatory Affairs Specialist Beckman Coulter, Inc. 1000 Lake Hazeltine Drive Chaska, MN 55318-1084
JAN 28 2003
Re: k023764 Trade/Device Name: UniCel™ Dxl 800 Access® Immunoassay System Regulation Number: 21 CFR 862.1215 Regulation Name: Creatine phosphokinase/creatine kinase or isoenzymes test system Regulatory Class: Class II Product Code: MMI, JJE, CEC, CGR, CHP, JLW Dated: November 8, 2002 Received: November 12, 2002
Dear Ms. Weist:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
#### Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## ATTACHMENT B: INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K003 76
PAGE 1 OF 1
Device Name: UniCel™ Dxl 800 Access® Immunoassay System
Indications for Use:
The UniCel Dxl 800 Access Immunoassay System is a microcomputer controlled, random and continuous access analyzer that includes an external computer. This computer stores the system user interface (UI) software and allows the operator to interface with and direct the instrument software. The UniCel Dxl 800 System uses enzyme immunoassays (utilizing paramagnetic particle solid phase and chemiluminescent detection) for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. The UniCel Dxl 800 System is an in vitro diagnostic device for use in the clinical laboratory.
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Veronican R. Celuin por Dr. Jean Cooper (
(Optional Format 3-10-98)
Division of Clinical Laboratory De 510(k) Number
{5}------------------------------------------------
510(k) Number (if known):
## Device Name: AccuTnl™ and AccuTnl Calibrators on the Access® Immunoassay Systems
#### Indications For Use:
The Access AccuTnl assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cardiac troponin 1 (cTnl) levels in human serum and plasma using the Access Immunoassay Systems to aid in the diagnosis and treatment of myocardial infarction and cardiac muscle damage.
Cardiac troponin I determination aids in the risk stratification of patients with unstable angina or non-ST seqment elevation acute coronary syndromes with respect to relative risk of mortality, myocardial infarction, or increased probability of ischemic events requiring urgent revascularization procedures.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Clinical Laboratory Devices
| 510(k) Number | 1023764 |
|---------------|---------|
|---------------|---------|
| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The Counter Use |
|------------------------------------------|----|----------------------|
|------------------------------------------|----|----------------------|
(Optional Format 1-2-96)
{6}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Access® Cortisol on the Access® Immunoassay Systems
## Indications For Use:
The Access® Cortisol Assay is a paramagnetic particle chemiluminescent immunoassay for the quantitative determination of Cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Clinical Laborat
| 510(k) Number | K023764 |
|---------------|---------|
|---------------|---------|
| Prescription Use (Per 21 CFR 801.109) | <img alt="check mark" src="check_mark.png"/> | OR | Over-The Counter Use ______ |
|---------------------------------------|----------------------------------------------|----|-----------------------------|
|---------------------------------------|----------------------------------------------|----|-----------------------------|
(Optional Format 1-2-96)
{7}------------------------------------------------
Page 1 of 1
510(k) Number (if known):
Device Name: Access® Estradiol on the Access® Immunoassay Systems
#### Indications For Use:
The Access Estradiol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of Estradiol levels in human serum, using the Access Immunoassay Systems.
Levels of estradiol are useful in monitoring ovulatory status. Because Estradiol levels reflect follicular maturation, the measurement of estradiol is a valuable tool in the assessment of sexual development, etiology of amenorrhea, causes of infertility and menopause. Abnormally high levels in males are indicative of feminizing syndromes such as gynecomastia.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Melvin for Jean Cooper
(Division Sign-Off)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The Counter Use
Division of Clinical Laboratory Devices
510(k) Number
(Optional Format 1-2-96)
{8}------------------------------------------------
510(k) Number (if known):
Device Name: Access® Free T4 on the Access® Immunoassay Systems
#### Indications For Use:
The Access Free T4 assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of free thyroxine levels in human serum and plasma (heparin) using the Access Immunoassay Systems.
A free thyroxine test system is a device intended to measure free (not protein bound) thyroxine (thyroid hormone) in human serum or plasma (heparin). Levels of free thyroxine in plasma are thought to reflect the amount of thvroxine hormone available to the cells and may therefore determine the clinical metabolic status of thyroxine. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Division of Clinical Laboratory Deview 510(k) Number OR Over-The Counter Use Prescription Use (Per 21 CFR 801.109) (Optional Format 1-2-96)
{9}------------------------------------------------
Page 1 of 1
510(k) Number (if known):
## Device Name: Access® HYPERsensitive hTSH on the Access® Immunoassay Systems
#### Indications For Use:
The Access HYPERsensitive hTSH assay provides in vitro quantitative measurement of human thyroid stimulating hormone (hTSH) in human serum or plasma. The Access HYPERsensitive hTSH assay is indicated for use with patients where an evaluation of their thyroid status is desired.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Vp Calun for Jean Cooper
(Division Sign-Off)
Division of Clinical Laboratory Devi
Clinical Laboratory Devices
510(k) Number K023764
Prescription Use (Per 21 CFR 801.109) Over-The Counter Use
(Optional Format 1-2-96)
26
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.