The GMRS Adapter is intended to mate the Proximal Femoral Components and Femoral Bodies of the MRS or Proximal Femoral Body or Extension Pieces of the GMRS to the distal femoral stems of the Restoration™M Modular Hip System. The GMRS adapter is intended for use with the above-noted components in proximal femoral replacement indicated in Oncology cases where radical resection and replacement of bone is required, and in limb salvage procedures where radical resection and replacement of bone is required. Adequate bone stock must be present to allow the use of the Restoration™ Modular Distal Femoral Stems.
Device Story
GMRS Adapter is a two-piece orthopedic implant (adapter and locking screw) designed to connect proximal femoral components of the MRS/GMRS system to Restoration Modular distal femoral stems. Fabricated from titanium alloy; available in four sizes (standard, +10mm, +20mm, +30mm). Features male/female tapers for modular assembly; includes scribe marks for anteversion orientation. Locking screw augments taper lock. Used by orthopedic surgeons in clinical settings for limb salvage and oncology-related proximal femoral replacement. Device facilitates press-fit fixation. Benefits include enabling modular reconstruction of bone defects following radical resection.
Indicated for proximal femoral replacement in oncology cases requiring radical bone resection/replacement and limb salvage procedures (severe trauma, failed previous prosthesis, oncology). Contraindicated in patients with local/distant tumor spread, pathological fracture, overt infection, poor biopsy incision placement, rapid disease progression, mental/neuromuscular disorders risking instability/fixation failure, or inadequate bone stock for press-fit stems.
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum. This generic type of device includes designs which are intended to be fixed to the bone with bone cement (§ 888.3027) as well as designs which have large window-like holes in the stem of the device and which are intended for use without bone cement. However, in these latter designs, fixation of the device is not achieved by means of bone ingrowth.
Predicate Devices
Femoral Body Segments of the MRS System (Howmedica Osteonics Corp.)
T3 Femoral Component (Howmedica Osteonics Corp.)
Submission Summary (Full Text)
{0}------------------------------------------------
K023692 GMRS Adapter Response to Questions February 25, 2003
Submission Information
### 510(k) Summary
page 1 of 3
2 0 2003
| Name and Address of Sponsor: | Howmedica Osteonics Corp.<br>59 Route 17<br>Allendale, NJ 07401 | |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------|--|
| For Information contact: | Margaret F. Crowe<br>Regulatory Affairs Consultant<br>Howmedica Osteonics Corp.<br>59 Route 17<br>Allendale, NJ 07401 | |
| Device Identification | | |
| Proprietary Name: | Global Modular Replacement System Adapter | |
| Common Name: | Proximal Femoral Replacement | |
| Classification Name and Reference: | Prosthesis, Hip, Semi-constrained Metal/Polymer<br>Uncemented | |
| Proposed Regulatory Class: | Class II | |
| Device Product Code: | OR(87) LWJ | |
### Intended Use
The GMRS Adapter is intended to mate the Proximal Femoral Components and Femoral Bodies of the MRS or Proximal Femoral Body or Extension Pieces of the GMRS to the distal femoral stems of the Restoration™M Modular Hip System. The GMRS adapter is intended for use with the above-noted components in proximal femoral replacement indicated in Oncology cases where radical resection and replacement of bone is required, and in limb salvage procedures where radical resection and replacement of bone is required. Adequate bone stock must be present to allow the use of the Restoration™ Modular Distal Femoral Stems Indications and contraindications of the GMRS Adapter
{1}------------------------------------------------
K023692 GMRS Adapter Response to Questions February 25, 2003 when used with the Proximal Femoral Bodies of the MRS and GMRS femoral stems of the Restoration™ Modular Hip System follow:
# Indications
Proximal femoral replacement in Oncology cases where radical resection and replacement of bone is required, and in limb salvage procedures where radical resection and replacement of the bone is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous prosthesis, and/or Oncology indications.
## Contraindications
#### As related to Bone Tumors A.
Not all bone tumors may be treated successfully by segmental resection. Any condition that may have already resulted in either local or distant spread of the tumor may be a contraindication. Examples of such conditions include:
- pathological fracture; .
- . overt infection;
- . inopportune placement of biopsy incision; and,
- rapid disease progression beyond a respectable margin. .
- C. As related to Failed Previous Prosthesis and Trauma
- Any active or suspected latent infection in or about the hip joint. .
- . Any mental or neuromuscular disorder which would create an unacceptable risk of prosthesis instability, prosthesis fixation failure, or complication in postoperative care.
- Bone stock compromised by disease, infection, or prior implantation that cannot . provide adequate support and fixation of the prosthesis.
For the use of GMRS Adaper with the Distal Stems of the Restoration Modular Hip System, the following additional contraindication should be noted:
- Inadequate bone stock to allow the use of a press fit stem .
{2}------------------------------------------------
K023692 GMRS Adapter Response to Questions February 25, 2003 The GMRS Adapter is intended to be used in a press fit mode.
023692
page 3 of 3
# Device Description
The GMRS Adapter is a single use, two piece design consisting of the Adapter and the Adapter Locking Screw The GMRS Adapter is fabricated from titanium alloy (titanium -6 Aluminum-4 Vanadium) and is available in four sizes: standard, +10mm, +20mm, and +30mm. The proximal portion of the GMRS Adapter serves as the male portion of a taper lock that is assembled to the female taper in the distal end of the MRS or GMRS Proximal Femoral Component or Extension Pieces. The distal portion of the GMRS Adapter contains a female taper that accepts the male taper of the Restoration™ Modular Distal Femoral Stems. There are scribe marks on the GMRS Adapter to allow the surgeon to orient the amount of anteversion of the MRS or GMRS Proximal Femoral Component. The proximal portion of the male taper of the GMRS Adapter contains a female thread that serves two purposes: 1) to accept the GMRS Adapter Locking Screw after the taper lock has been engaged; and 2) for mating with a separation instrument should the surgeon need to reposition the adapter after it has been assembled to the Distal Stems of the Restoration™ Modular Total Hip System.
The GMRS Adapter Locking Screw is fabricated from titanium alloy (titanium- 6 Aluminum -4 Vanadium) and is available in four sizes that correspond to the sizes of the Adapter: standard; +10mm, +20mm, and +30mm. The GMRS Adapter Locking Screw is used to augment the taper lock of the GMRS Adapter into the Restoration™ Modular Distal Femoral Stems. The GMRS Adapter Locking Screw is packaged with the corresponding size GMRS Adapter.
Equivalent products include:
- Femoral Body Segments of the MRS System Howmedica Osteonics Corp. 1.
- T3 Femoral Component Howmedica Osteonics Corp. 2.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or a stylized human figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 0 2003
Ms. Margaret Crowe Regulatory Affairs Consultant Howmedica Osteonics Corp. 59 Route 17 South Allendale, NJ 07401
Re: K023692
Trade/Device Name: Global Modular Replacement System (GMRS) Adapter Regulation Number: 888.3350 Regulation Name: Hip joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: II Product Code: LWJ Dated: February 10, 2003 Received: February 12, 2003
Dear Ms. Crowe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 -- Ms. Margaret Crowe
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Mark McMillan
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K023692 GMRS Adapter Response to Questions February 25, 2003
510(k) Number (if known): K023692
page 1 of 2
Device: GMRS Adapter
The GMRS Adapter is intended to mate the Proximal Femoral Components and Femoral Bodies of the MRS or Proximal Femoral Body or Extension Pieces of the GMRS to the distal femoral stems of the Restoration™M Modular Hip System. The GMRS adapter is intended for use with the above-noted components in proximal femoral replacement indicated in Oncology cases where radical resection and replacement of bone is required, and in limb salvage procedures where radical resection and replacement of bone is required. Adequate bone stock must be present to allow the use of the Restoration™ Modular Distal Femoral Stems. Indications and contraindications of the GMRS Adapter when used with the Proximal Femoral Bodies of the MRS and GMRS, and the distal femoral stems of the Restoration™ Modular Hip System follow:
# Indications
Proximal femoral replacement in Oncology cases where radical resection and replacement of bone is required, and in limb salvage procedures where radical resection and replacement of the bone is required. Limb salvage procedures would include surgical intervention for severe trauma, failed previous prosthesis, and/or Oncology indications.
## Contraindications
- A. As related to Bone Tumors
Not all bone tumors may be treated successfully by segmental resection. Any condition that may have already resulted in either local or distant spread of the tumor m contraindication. Examples of such conditions include:
- . pathological fracture;
- . overt infection;
- . inopportune placement of biopsy incision; and,
- rapid disease progression beyond a respectable margin. 10(k) Number 1 .
- As related to Failed Previous Prosthesis and Trauma B.
4
ral, Restorative
{6}------------------------------------------------
K023692 GMRS Adapter Response to Questions February 25, 2003
023692
page 2 of 2
- Any active or suspected latent infection in or about the hip joint. ●
- Any mental or neuromuscular disorder which would create an unacceptable risk of . prosthesis instability, prosthesis fixation failure, or complication in postoperative care.
- Bone stock compromised by disease, infection, or prior implantation that cannot . provide adequate support and fixation of the prosthesis.
For the use of GMRS Adaper with the Distal Stems of the Restoration Modular Hip System, the following additional contraindication should be noted:
- Inadequate bone stock to allow the use of a press fit stem .
The GMRS Adapter is intended to be used in a press fit mode.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use OR 801.109)
Over-the Counter-Use (per 21 CFR
Mark n Milke
Division Sign-Off
Division of General, Restorative
and Neurological Devices
K023692
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.