X O ODONTOSURGE 4

K023672 · Xo Care A/S · EKZ · Apr 29, 2003 · Dental

Device Facts

Record IDK023672
Device NameX O ODONTOSURGE 4
ApplicantXo Care A/S
Product CodeEKZ · Dental
Decision DateApr 29, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4920
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The X O Odontosurge 4 is intended for use in cutting (removing) soft tissue and controlling bleeding in the oral cavity during surgical procedures in all phases of dentistry, including prosthodontics, periodontics, endodontics, orthodontics, oral surgery, and routine restorative dentistry.

Device Story

High-frequency electrosurgical unit; components include control box, handpiece, cord, and seven electrodes. Used in dental clinics by practitioners to cut and coagulate soft tissue. Device transforms electrical energy to perform tissue resection and hemostasis. No software utilized. Benefits include tissue removal and bleeding control during various dental surgical procedures.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

High-frequency electrosurgical unit; reusable; includes control box, handpiece, and seven electrodes. No software. Sterilization method not specified.

Indications for Use

Indicated for removal of soft tissue and hemostasis in the oral cavity for patients undergoing dental procedures across all specialties, including prosthodontics, periodontics, endodontics, pedodontics, orthodontics, oral surgery, and restorative dentistry.

Regulatory Classification

Identification

A dental electrosurgical unit and accessories is an AC-powered device consisting of a controlled power source and a set of cutting and coagulating electrodes. This device is intended to cut or remove soft tissue or to control bleeding during surgical procedures in the oral cavity. An electrical current passes through the tip of the electrode into the tissue and, depending upon the operating mode selected, cuts through soft tissue or coagulates the tissue.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Premarket Notification 510 (k) Summary for the Odontosurge 4 # Submitter: APR 2 9 2003 X O Care James Dieroff, Official United States Agent 1100 S. Coast Highway, Ste 204 Laguna Beach, CA 92651 Date Summary was prepared: October 29, 2002 ### Name of the device: X O Odontosurge 4 #### ldentification of Predicate device: ArthroCare Dental Electrosurgery System ArthroCare Corporation K962445 ### Description of the device: The X O Odontosurge 4 is a high frequency electrosurgery unit that is comprised of the following three principal components: Control Box, Handpiece and Cord, and seven different Electrodes contained in a separate case. The X O Odontosurge 4 is compact and easily portable. The X O Odontosurge 4 is a reusable electrosurgical unit that is used to cut and to coagulate soft tissue during procedures in all disciplines of dentistry. There is no software utilized in the operation of the X O Odontosurge 4. ### Intended use: The X O Odontosurge 4 is intended for use in cutting (removing) soft tissue and controlling bleeding in the oral cavity during surgical procedures in all phases of dentistry, including prosthodontics, periodontics, endodontics, orthodontics, oral surgery, and routine restorative dentistry. ### Comparison of device characteristics to predicate: The intended use, sterilization method, and mode of operation of the X O Odontosurge 4 are the same as the ArthroCare Dental Electrosurgery System. The difference between the two devices is the frequency of operation. However, X O Care believes that this operating frequency offers many convenience and safety advantages. # Conclusion: The intended use, general design, materials of fabrication, and performance of the X O Odontosurge 4 are the same as the predicate device, ArthroCare Dental Electrosurgery System and devices already on the market. Therefore, the X O Odontosurge 4 that is the subject of this 510 (k) is substantially equivalent to dental Electrosurgical units in interstate commerce prior to May 28, 1976. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling a symbol. The symbol consists of three stylized human profiles facing right, with flowing lines suggesting movement or progress. The overall design is simple and conveys a sense of health and well-being. Public Health Service APR 2 9 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 X O Care A/S C/O Mr. James Dieroff Odonto-Wave 1100 South Coast Highway, Suite 204 Laguna Beach, California 92651 Re: K023672 Trade/Device Name: X O Odontosurge 4 Regulation Number: 21 CFR 872.4920 Regulation Name: Dental Electrosurgical Unit and Accessories Regulatory Class: II Product Code: EKZ Dated: April 22, 2003 Received: April 23, 2003 Dear Mr. Dieroff: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ #### Page 2 - Mr. Dieroff Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market. If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Susan Runse Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Appendix G 510 (k) Number: K023672 None assigned as of this time #### Device Name: X O Odontosurge 4 # Indications for Use: The X O Odontosurge 4 is intended for use in removing soft tissue and controlling bleeding in the oral cavity in all phases of dentistry, including prosthodontics, periodontics, endodontics, pedodontics, orthodontics, oral surgery, and routine restorative dentistry. Concurrence of CDRH, Office of Device Evaluation (ODE) Image /page/3/Picture/10 description: The image contains a large, bold 'X' mark. The 'X' is drawn with thick lines, indicating it was likely created with a marker or pen. A horizontal line is drawn underneath the 'X', emphasizing the mark. Prescription Use (per 21 CFR 801.109) Over-the-counter (OTC) Use Kein Muluy Son MSR General Hospital. 510(k) Number: K023672
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...