K023658 · Carr Metal Products, Inc. · FRG · Mar 25, 2003 · General Hospital
Device Facts
Record ID
K023658
Device Name
STERILIZATION CASES AND TRAYS
Applicant
Carr Metal Products, Inc.
Product Code
FRG · General Hospital
Decision Date
Mar 25, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
Medical device instrumentation cases, trays, and cassettes intended to hold instruments and accessories in place throughout the entire instrument use, cleaning, and sterilization cycle. These cases, trays, and cassettes are suitable for Pulsing High Vacuum (pre-vac) steam sterilization at 132 °C for a 4-minute minimum plus a minimum of 15 minutes drying time.
Device Story
Sterilization cases, trays, and cassettes designed to secure medical and dental instrumentation during cleaning and sterilization. Containers feature lids and internal holding mechanisms for instruments. Constructed from metal and plastic. Used in hospitals, ORs, and dental offices. Compatible with Pulsing High Vacuum (pre-vac) steam sterilization (132°C, 4-minute cycle, 15-minute drying). Facilitates organized instrument handling throughout the sterilization workflow.
Clinical Evidence
Bench testing only. Performance validated according to AAMI TIR No. 12-1994 for Pulsing High Vacuum (pre-vacuum) steam sterilization. Validation included half-cycle testing to challenge the sterilization method.
Technological Characteristics
Materials: Aluminum, stainless steel, plastic. Design: Container with lid, internal instrument holding features, optional latch system. Dimensions: 10.375" x 21.75" x 2.5" to 10.375" x 21.75" x 5". Sterilization: Pulsing High Vacuum steam (132°C, 4-min exposure, 15-min dry).
Indications for Use
Indicated for holding medical device instrumentation in place during use, cleaning, and sterilization cycles in hospital, OR, physician, and dental office settings.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Symmetry - PolyVac surgical instrument delivery system (K012105)
Submission Summary (Full Text)
{0}------------------------------------------------
K023658
## 510(k) Summary of Safety and Effectiveness
## Non-Confidential Summary of Safety and Effectiveness
Page 1 of 2 January 10, 2003
| Carr Medical Products, Inc.<br>3735 N. Arlington Ave.<br>Indianapolis, IN 46218 | Tel - (317) 542-0691<br>Fax - (317) 542-0694 |
|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Alan L. Booker - Operations Manager |
| Proprietary or Trade Name: | Sterilization Cases, Trays, and Cassettes |
| Common/Usual Name: | Sterilization cases, trays, and cassettes |
| Classification Name: | Sterilization Wrapper Pack, Bag, and Accessories |
| Predicate Devices: | C/T Med-Systems Cassette system (Carr Medical<br>Products) - K980065<br>Symmetry - PolyVac surgical instrument delivery<br>system - K012105 |
## Device Description:
Sterilization cases, trays, and cassettes designed to hold various general dental, medical device instrumentation during the cleaning, use, and sterilization process. The design is a container (case) with separate lid, which has various methods of holding the instruments in place. These are trays and cassettes, which are made of metal and plastic. They are available in various sizes ranging from [width x length x height (depth)] 10.375 " x 21.75" x 21.75" x 21.75" x 5" with trays and cassettes of similar size, which are stacked inside the case.
## Indicated Use:
Medical device instrumentation cases, trays, and cassettes intended to hold instruments and accessories in place throughout the entire instrument use, cleaning, and sterilization cycle.
These cases, trays, and cassettes are suitable for Pulsing High Vacuum (pre-vac) steam sterilization at 132 ℃ for a 4-minute minimum with a 15-minute drying time.
## Environment of Use:
Hospital, Operating Room (OR), physician and dental office or places where instruments are sterilized.
{1}------------------------------------------------
## Non-Confidential Summary of Safety and Effectiveness Page 2 of 2 January 10, 2003
#### Summary of Performance testing:
The Carr Metal Products Sterilization Cases, Trays, and Cassettes were independently tested according to AAMI TIR No. 12-1994 for their performance with the Pulsing High Vacuum (prevacuum) Steam sterilization method.
#### General Technical Characteristics
| Attribute | Proposed devices |
|---------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | Indicated for holding medical device instrumentation in<br>place throughout entire instrument use, cleaning, and<br>sterilization cycle |
| Sterilization Method | Pulsing High Vacuum (pre-vacuum) steam sterilization<br>at 132°C 4 minute minimum cycle with a 15-minute<br>minimum drying time. |
| Intended to be reused | Yes |
| Intended Environment of Use | Hospital, Operating Room (OR), physician and dental<br>office or places where instruments are sterilized |
| Design | |
| Various sizes of cases and lids and associated trays<br>and cassettes offered | 10.375 " x 21.75" x 2.5" to 10.375" x 21.75" x 5" with<br>trays and cassettes of similar size, which are stacked<br>inside the case |
| Utilizes various methods of holding instruments in<br>place | Yes |
| May incorporate latch system to hold lid in place | Yes |
| Materials | |
| Aluminum, stainless steel, plastic | Yes |
| Performance Standards | |
| None under Section 514 | Yes |
| Tested in accordance to AAMI TIR No. 12-1994 | Yes |
| Validation study performed with half cycles to<br>challenge sterilization method used | Yes |
## Differences between Other Legally Marketed Predicate Devices
There are no significant differences between the intended device and the predicates, C/T Med-Systems – K980065 other than larger sizes and Symmetry – PolyVac surgical instrument delivery system - K012105.
{2}------------------------------------------------
MAR 2 5 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Carr Metal Products, Incorporated C/O Mr. Paul Dryden ProMedic, Incorporated 6329 West Waterview Court McCordsville, Indiana 46055-9501
Re: K023658
Trade/Device Name: Carr Sterilization Cases, Trays and Cassettes Regulation Number: 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: FRG Dated: January 10, 2003 Received: January 13, 2003
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Susan Kuono
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
# Page 1 of 1
| 510(k) Number: | K023658 (To be assigned) |
|----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Carr sterilization cases, trays, and cassettes |
| Intended Use: | Medical device instrumentation cases, trays, and cassettes<br>intended to hold instruments and accessories in place<br>throughout the entire instrument use, cleaning, and<br>sterilization cycle. |
| | These cases, trays, and cassettes are suitable for Pulsing<br>High Vacuum (pre-vac) steam sterilization at 132 °C for a<br>4-minute minimum plus a minimum of 15 minutes drying<br>time. |
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per CFR 801.109)
or
Over-the-counter use _________________________________________________________________________________________________________________________________________________________
Clim S. Lim
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital, Division Control, Der e 40/6) Niimb
Page 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.