K023606 · Tisport · IOR · Nov 19, 2002 · Physical Medicine
Device Facts
Record ID
K023606
Device Name
TILITE YG AND TILITE YGS
Applicant
Tisport
Product Code
IOR · Physical Medicine
Decision Date
Nov 19, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3850
Device Class
Class 1
Attributes
Pediatric
Indications for Use
The intended use of this device (growable, manual, rigid wheelchair) is the same as the predicate device, the Zippie® (growable, manual, rigid wheelchair) manufactured by Sunrise Medical Inc. The intended use for the growable, manual, rigid wheelchair is to provide mobility to physically impaired children. The manual wheelchair is intended for ongoing daily use and is designed so that the wheelchair can be adjusted as the user grows.
Device Story
TiLite YG and YGS are rigid, manual, growable wheelchairs for children. Device provides mobility for daily use; adjustable frame accommodates user growth. Constructed from titanium; offers superior strength-to-weight ratio and vibration absorption compared to aluminum predicates. Supports up to 250-lb passengers. Used in clinical or home settings; operated by patient or caregiver. Customization options allow for precise clinical fitting, enhancing safety and access to accessories.
Clinical Evidence
Bench testing only. Device meets requirements of ISO 7176 Parts 1, 3, 5, 7, and 8.
Technological Characteristics
Rigid, manual, growable wheelchair frame. Material: Titanium. Sensing/Actuation: Mechanical. Connectivity: None. Sterilization: Not applicable. Standards: ISO 7176 Parts 1, 3, 5, 7, and 8.
Indications for Use
Indicated for physically impaired children requiring mobility assistance for daily use. Conditions include: amputee, arthritis, arthrogryposis, cerebral palsy, geriatric conditions, head injury/trauma, hemiplegia, Multiple Sclerosis, Muscular Dystrophy, paraplegia, Polio, quadriplegia, Spina Bifida, Stroke/CVA, tetraplegia, and other immobilizing/debilitating conditions including spinal cord injuries and extremity paralysis.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
23606
NOV 1 9 2002
T H E ULTI M T E
#### 501(k) SUMMARY 9.
#### 9.1. Date:
October 25, 2002
#### 9.2. Submitted By:
TiSport, LLC 1426 East Third Avenue Kennewick, WA 99337 Phone: 509-586-6117 Fax: 509-586-2413
#### 9.3. Contact Person:
Richard S. Forman Vice President - Legal and Business Affairs TiSport, LLC 1426 East Third Avenue Kennewick, WA 99337 Phone: 509-586-6117, Ext. 238 Fax: 509-586-2413
#### Trade/Proprietary Name of Device: 9.4.
TiLite YG and TiLite YGS
#### Common Name of Device: 9.5.
Growable, Manual, Rigid Wheelchair
#### 9.6. Classification Name of Device:
Wheelchair, Mechanical (per 21 CFR section 890.3850)
#### Classification of Device: 9.7.
Class I
#### 9.8. Panel:
# Physical Medicine - Prosthetic Devices Subpart D - 890
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with the letters 'Ti' in white against a black square background. The 'T' is capitalized and the 'i' is lowercase, with a dot above it. The letters are bold and sans-serif. The logo is simple and modern.
### Product Code: THE 9.9.
89IOR
#### Legally Marketed Predicate Device For Claimed Equivalence: 9.10.
Zippie® Rigid Manual Wheelchair (K890050)
#### 9.11. Description of Device:
The TiLite YG and YGS model wheelchairs are rigid manual titanium wheelchairs.
#### Intended Use of the Device: 9.12.
The intended use of this device (growable, manual, rigid wheelchair) is the same as the predicate device. The intended use for the manual wheelchair is to provide mobility to physically impaired children. The manual wheelchair is intended for ongoing daily use and is designed so that the wheelchair can be adjusted as the user grows.
#### 9.13. Target Population:
This device is indicated for individuals with the specific medical conditions listed, but the indications are not necessarily limited to such conditions:
- Amputee:
- Arthritis;
- Arthrogriposis;
- Cerebral Palsy;
- geriatric conditions;
- head injury or trauma;
- hemiplegic:
- Multiple Sclerosis;
- Muscular Dystrophy;
- paraplegic;
- Polio;
- quadraplegic;
- Spina Bifida;
- Stoke/CVA;
- tetraplegic; and
- other immobilizing or debilitating conditions, including spinal cord injuries and other lower and upper extremity paralysis
#### Testing Results: 9.14.
Meets the requirements of the ISO 7176 Parts 1, 3, 5, 7,and 8.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with the letters 'Ti' in white against a black square background. Above the 'i' is a white circle. Below the 'Ti' is the word 'LITE' in smaller white letters. The logo appears to be for a product or company named 'Ti Lite'.
#### 9.15. Device Comparison:
The principal differences between the submitted device and the predicate device are that TiLite YG and YGS series wheel chairs are the materials used in the manufacturing of the frames. The TiLite YG and YGS wheelchairs are made with titanium whereas the predicate device is made with aluminum. TiSport believes that titanium frames provide certain benefits when compared with aluminum frames from a safety perspective and a clinical perspective because of titanium's proven superior strength-to-weight ratio and its natural ability to absorb vibration. Accordingly, for example, the TiLite YG and YGS wheelchairs can accommodate at 250-lb. Passenger, whereas the predicate device can accommodate only a 150-lb. passenger.
The other difference is that the TiLite YG and YGS models offer a wider range of customization for the end user. TiSport believes that greater customization options allows for a better opportunity to properly "fit" the end user in a clinical setting as well as ensuring better safety and access to the chairs' options and accessories.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 9 2002
TiSport, LLC Richard S. Forman Vice President 1426 East Third Avenue Kennewick. Washington 99337-9669
Re: K023606
Trade/Device Name: Tilite YG and Tilite YGS Regulation Number: 890.3850 Regulation Name: Wheelchair, mechanical Regulatory Class: Class I Product Code: IOR Dated: October 25, 2002 Received: October 28, 2002
Dear Mr. Forman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that vour device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 -- Mr. Richard S. Forman
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sincerely yours,
Mark A. Mikkelsen
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
. . .
{5}------------------------------------------------
## STATEMENT OF INDICATIONS FOR USE
#### Device Name: 7.1.
TiLite YG and TiLite Y
#### 7.2. . Indication for Use:
The intended use of this device (growable, manual, rigid wheelchair) is the same as the predicate device, the Zippie® (growable, manual, rigid wheelchair) manufactured by Sunrise Medical Inc. The intended use for the growable, manual, rigid wheelchair is to provide mobility to physically impaired children. The manual wheelchair is intended for ongoing daily use and is designed so that the wheelchair can be adjusted as the user grows.
This device is indicated for individuals with the specific medical conditions listed, but the indications are not necessarily limited to such conditions:
- Amputee;
- Arthritis:
- Arthrogriposis;
- Cerebral Palsy;
- geriatric conditions;
- head injury or trauma;
- hemiplegic;
- Multiple Sclerosis;
- Muscular Dystrophy;
- paraplegic;
- Polio:
- quadraplegic;
- Spina Bifida;
- Stoke/CVA;
- tetraplegic; and
- other immobilizing or debilitating conditions, including spinal cord injuries and other lower and upper extremity paralysis
Minh N Milken
Division Sign-C Division of Gene and Neurologica®
K023606
10 - ) Number -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.