← Product Code [LYY](/productcode/LYY) · K023590

# POWDERED NATURAL AND/OR COLORED LATEX EXAM GLOVES (K023590)

_Yty Industry (Manjung) Sdn Bhd · LYY · Nov 8, 2002 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K023590

## Device Facts

- **Applicant:** Yty Industry (Manjung) Sdn Bhd
- **Product Code:** [LYY](/productcode/LYY.md)
- **Decision Date:** Nov 8, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## Device Story

Disposable powdered natural latex examination gloves; available in natural, pink, or blue colors. Worn on examiner's hand or finger to provide a physical barrier; prevents cross-contamination between patient and healthcare professional. Used in hospitals, clinics, dental offices, and laboratories. Device functions as a passive protective barrier; no active components or electronic processing. Benefits include reduction of infection transmission risk during patient examinations.

## Clinical Evidence

Bench testing only. Device met or exceeded ASTM D3578-01 standards for watertight integrity (1000ml test), dimensions, physical properties (tensile strength, elongation), and powder content. Biocompatibility testing confirmed dermal sensitization and primary skin irritation safety.

## Technological Characteristics

Material: Natural rubber latex. Standards: ASTM D3578-01, ASTM D5712-99 (protein), ASTM D6124-01 (powder), ASTM D5151-99 (watertight), ISO 2859. Form factor: Disposable, powdered, ambidextrous gloves in various sizes. Energy source: None. Sterilization: Not specified.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Shield Gloves Manufacturer ([K943807](/device/K943807.md))
- YTY Industry (Manjung) SDN BHD ([K974191](/device/K974191.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

8 2002 NOV

|                          | K023590                                                                                                                                                                                                                                                                                                                 |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) SUMMARY           |                                                                                                                                                                                                                                                                                                                         |
| Submitted For:           | YTY INDUSTRY (MANJUNG) SDN BHD                                                                                                                                                                                                                                                                                          |
| Submitted By:            | TUCKER & ASSOCIATES<br>Official Correspondent for YTY INDUSTRY<br>(MANJUNG) SDN BHD<br>JANNA P. TUCKER, President-CEO<br>198 Avenue de la D'emerald<br>Sparks, NV 89434-9550                                                                                                                                            |
| Phone:                   | 775-342-2612                                                                                                                                                                                                                                                                                                            |
| Fax:                     | 775-342-2613                                                                                                                                                                                                                                                                                                            |
| E-Mail:                  | Tuckerjan@aol.com                                                                                                                                                                                                                                                                                                       |
| Date of Submission:      | 23 October 2002                                                                                                                                                                                                                                                                                                         |
| Device Name:             | POWDERED NATURAL AND/OR COLORED<br>LATEX EXAM GLOVES<br>Class I Device, 80LYY                                                                                                                                                                                                                                           |
| Proprietary Name:        | (Multiple Private Labels)                                                                                                                                                                                                                                                                                               |
| Labels/Labeling:         | This device will be marketed to healthcare professionals at<br>Dentist and Doctor Offices, Laboratories, Clinics and<br>Hospitals through its distributors for the intended use.                                                                                                                                        |
| Intended Use:            | A patient examination glove is a disposable device intended<br>for medical purposes that is worn on the examiner's hand<br>or finger to prevent contamination between patient and<br>examiner.                                                                                                                          |
| Substantial Equivalence: | Both in its intended use and/or physical<br>characteristics, this device is equivalent to devices<br>currently marketed by U.S. companies. It is Substantially<br>Equivlalent to the devices manufactured by Shield Gloves<br>Manufacturer (M) K943807, and YTY Industry (Manjung)<br>SDN BHD, K974191 (natural color). |

EXHIBIT N
Page 101 of 103

{1}------------------------------------------------

#### 7.0 Summary of Performance Data:

Performance data of gloves based on ASTM D3578-01 and FDA 1000ML watertight test.

| TEST                                                             | ASTM D3578-01/AE2                              | EVERGREEN POWDERED<br>LATEX EXAM. GLOVES   |
|------------------------------------------------------------------|------------------------------------------------|--------------------------------------------|
| 1. Watertight (1000ml)                                           | Multiple Normal<br>GI<br>AQL = 2.5             | Pass GI<br>AQL = 2.5                       |
| 2. Length (mm)<br>Size<br>XS<br>S<br>M<br>L<br>XL                | Min 220<br>Min 220<br>Min 230<br>Min 230<br>-  | 240 mm minimum for all sizes               |
| 3. Palm width (mm)<br>Size<br>XS<br>S<br>M<br>L<br>XL            | 70 ± 10<br>80 ± 10<br>95 ± 10<br>111 ± 10<br>- | 73 - 78<br>83 - 88<br>93 - 98<br>103 - 107 |
| 4. Thickness (mm)<br>(Single Layer)                              |                                                |                                            |
| Finger<br>Palm                                                   | Min 0.08<br>Min 0.08                           | Min 0.10<br>Min 0.10                       |
| 5. Physical Properties<br>Before Aging<br>Tensile Strength (MPa) | Min 18                                         | 23 - 27                                    |
| Ultimate Elongation (%)                                          | Min 650                                        | 820 - 880                                  |
| Stress at 500% Elongation                                        | Max 5.5                                        | 2.9-3.6                                    |
| After Aging<br>Tensile Strength (MPa)                            | Min 14                                         | 20 - 27                                    |
| Ultimate Elongation (%)                                          | Min 500                                        | 780 - 860                                  |
| 6. Powder Content                                                | 10mg per square decimeter max                  | Below 10 mg per square decimeter           |

EXHIBIT N
Page 102 of 103

Page 2 of 3

{2}------------------------------------------------

510k) Summary page 2.

Test Results (Means and/or Successful Results:

This device has met or exceeded the following standards and/or tests:

ASTM D 5712-99 ASTM D 3578-01aE2 ASTM D 6124-01 ASTM D 5151-99 ISO 2859

Bio-Compatibility: Dermal Sensitization Primary Skin Irritation

Conclusion:

This device is substantially equivalent to the devices approved as K943807 and K974191.

EXHIBIT N
Page 103 of 103

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract image of a stylized caduceus, a symbol often associated with medicine and healthcare. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

8 2002 NOV

YTY Industry (Manjung) Sdn Bhd C/O Ms. Janna P. Tucker Tucker & Associates 198 Avenue De La D'emerald Sparks, Nevada 89434-9550

Re: K023590

Trade/Device Name: Powdered, Natural, Pink and Blue Colored Latex Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LYY Dated: October 23, 2002 Received: October 25, 2002

Dear Ms. Tucker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition; FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 -- Ms. Tucker

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Susan Dunn

Timothy A. Ulatowski Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

### Enclosure

{5}------------------------------------------------

# INDICATIONS FOR USE

## APPLICANT: YTY INDUSTRY (MANJUNG) SDN. BHD.

510(k) NUMBER: KO23590

# POWDERED, NATURAL AND/OR (Pink, Buce) DEVICE NAME: COLORED LATEX EXAM GLOVES

A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801

OR

Over-The-Counter Use

(Optional Format 1-2-96)

Qlin S. Lin

sion Sign-Off) on of Anesthesiology, General Hospital, on Control. Dental Device

EXHIBIT B
Page 2 of 103

510(k) Number: K 923080

---

**Source:** [https://fda-staging.innolitics.com/device/K023590](https://fda-staging.innolitics.com/device/K023590)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com