← Product Code [LZA](/productcode/LZA) · K023517

# DERMAGRIP-P POWDER FREE PINK NITRILE EXAMINATION GLOVES, NON-STERILE (K023517)

_Wrp Asia Pacific Sdn. Bhd. · LZA · Dec 2, 2002 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K023517

## Device Facts

- **Applicant:** Wrp Asia Pacific Sdn. Bhd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Dec 2, 2002
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The Powder Free Pink Nitrile Examination Glove, Non-Sterile is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## Device Story

Disposable nitrile examination glove; worn on examiner's hand or finger; prevents cross-contamination between patient and examiner. Device is non-sterile; powder-free. Modification from predicate involves change in color additive to pink. Used in clinical settings by healthcare professionals. Provides barrier protection.

## Clinical Evidence

Bench testing only. Device performance verified against ASTM D 6319-00aE3 for dimensions and physical properties, and ASTM D 6124-01 for powder residual. Freedom from pinholes verified per FDA 21 CFR 800.20.

## Technological Characteristics

Nitrile examination glove; powder-free. Meets ASTM D 6319-00aE3 (dimensions, physical properties) and ASTM D 6124-01 (powder residual < 2 mg/glove). Non-sterile. Color additive: pink.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Powder Free Blue Nitrile Examination Gloves ([K022904](/device/K022904.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K023517

Image /page/0/Picture/1 description: The image shows the logo for WRP Asia Pacific Sdn Bhd. The logo consists of the letters "WRP" in a bold, sans-serif font, enclosed in a rounded, black shape. To the right of the logo is the text "WRP Asia Pacific Sdn Bhd" in a similar font. Below the text is the number "1 4 7 8 1 7 V".

DEC 0 2 2002

### SPECIAL 510(k) SUMMARY

#### 1.0 Submitter:

| Name:      | WRP Asia Pacific Sdn Bhd                                                                                    |
|------------|-------------------------------------------------------------------------------------------------------------|
| Address:   | Lot 1, Jalan 3, Kawasan Perushaan Bandar Baru Salak Tinggi,<br>43900 Sepang, Selangor Darul Ehsan, MALAYSIA |
| Phone No.: | +60 3 8706 1486                                                                                             |
| Fax No.:   | +60 3 8706 1485                                                                                             |

17 OCT 2002 Date of Summary Prepared:

#### 2.0 Contact Person:

| Name:      | Mr. Terence Lim |
|------------|-----------------|
| Phone No.: | +60 3 8706 1486 |
| Fax No.:   | +60 3 8706 1485 |

#### 3.0 Modified Device Identification:

| Trade Name:          | 1) Dermagrip-P, and                                      |
|----------------------|----------------------------------------------------------|
|                      | 2) Multiple or Customers' Trade Name                     |
| Device Name:         | Powder Free Pink Nitrile Examination Gloves, Non-Sterile |
| Common Name:         | Examination Gloves                                       |
| Classification Name: | Patient Examination Gloves (per 21 CFR 880.6250)         |

### Identification of the Legally Marketed Device: 4.0

Class I patient examination gloves, 80LZA, powder free, that meets all the
requirements of ASTM standard D 6319 - 00a53 and FDA 21 CFR 800.20.

#### 5.0 Description of Device Modification:

The Powder Free Pink Nitrile Examination Gloves, Non Sterile is equivalent to the existing model, i.e. Powder Free Blue Nitrile Examination Gloves, Non Sterile which had submitted and cleared under 510(k) number K022904.

{1}------------------------------------------------

11023517

Image /page/1/Picture/1 description: The image shows the logo for WRP Asia Pacific Sdn Bhd. The logo consists of the letters "WRP" in a bold, sans-serif font, enclosed in a black oval shape. To the right of the logo is the company name, "WRP Asia Pacific Sdn Bhd," also in a bold, sans-serif font. Below the company name is the number 147017V.

The difference in this submission is:

- Change of colour additive used, i.e. Pink a)
The modification of colour does not affect the intended use of the device as well as it does not affect its safety and effectiveness. The indication for use and proposed labeling for the device are illustrated in subsequent sections.

The Powder Free Pink Nitrile Examination Gloves, Non-Sterile meets all the requirements of ASTM standard D 6319 - 00a23 and FDA 21 CFR 800,20.

#### 4,0 Intended Use of the Device:

The Powder Free Pink Nitrile Examination Gloves, Non-Sterile is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

#### 5.0 Summary of Technological Characteristics for the Modified Device:

The Powder Free Pink Nitrile Examination Gloves, Non-Sterile are summarized with the following technological characteristics compared to ASTM or equivalent standards.

| CHARACTERISTICS       | STANDARDS                                | DEVICE<br>PERFORMANCE |
|-----------------------|------------------------------------------|-----------------------|
| Dimensions            | ASTM D 6319 - 00aE3                      | Meets                 |
| Physical Properties   | ASTM D 6319 - 00aE3                      | Meets                 |
| Freedom from pinholes | ASTM D 6319 - 00aE3<br>FDA 21 CFR 800,20 | Meets                 |
| Powder Residual       | ASTM D 6124 - 01                         | Meets<br>< 2 mg/glove |

#### 6.0 Conclusion:

The Powder Free Pink Nitrile Examination Gloves, Non-Sterile will perform according to the glove performance standards referenced in section 7 above and meet ASTM standards, and FDA requirements for waterleak test on pinhole AQL. Consequently, this device is substantially equivalent to currently marketed devices.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# DEC 02 2002

Mr. Terence Lim Associate Manager, RA/QA WRP Asia Pacific Sdn. Bhd. Lot 1, Jalan 3, Kawasan Perusahaan Bandar Baru Salak Tinggi, 43900 Sepang, Selangor Darul Ehsan, MALAYSIA

Re: K023517

Trade/Device Name: DermaGrip-P Powder Free Pink Nitrile Examination Gloves. Non-Sterile Regulation Number: 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LZA Dated: November 7, 2002 Received: November 12, 2002

Dear Mr. Lim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 -- Mr. Lim

 Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618 . Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Susan Runner

Timothy A. Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo with the letters "WRP" in white against a black, rounded rectangle. The letters are bold and sans-serif. The logo has a simple, clean design. The letters are slightly distorted.

WRP Asia Pacific Sdn Bhd

1 4 7 8 1 7 V

## INDICATIONS FOR USE

Applicant:

WRP Asia Pacific Sdn Bhd

510(k) Number (if known):

\$23517________________________________________________________________________________________________________________________________________________________________________

Device Name:

DERMAGRIP-P POWDER FREE PINK NITRILE EXAMINATION GLOVES, NON-STERILE

Page 1 of 1

Indications For Use:

The Powder Free Pink Nitrile Examination Glove, Non-Sterile is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use     | OR | Over-The-Counter | <div> <div></div> </div> |
|----------------------|----|------------------|--------------------------|
| (Per 21 CFR 801.109) |    |                  |                          |

  

 Susan Runor
 
  

 (Division Sign-Off)
 
 Division of Anesthesiology, General Hospital,
   

 Infection Control, Dental Devices
 

| 510(k) Number | K023517 |
|---------------|---------|
|---------------|---------|

---

**Source:** [https://fda-staging.innolitics.com/device/K023517](https://fda-staging.innolitics.com/device/K023517)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).