← Product Code [OTN](/productcode/OTN) · K023516

# MONARC SLING SYSTEM (K023516)

_American Medical Systems, Inc. · OTN · Nov 19, 2002 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K023516

## Device Facts

- **Applicant:** American Medical Systems, Inc.
- **Product Code:** [OTN](/productcode/OTN.md)
- **Decision Date:** Nov 19, 2002
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The MONARC Sling System is intended for the placement of a pubourethral sling for the treatment of female stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency.

## Device Story

MONARC Sling System is a sterile, single-use surgical kit for treating female stress urinary incontinence. Components include two stainless steel 22cm spiral needle passers and one polypropylene sling mesh with attached dilating connectors. A fixed absorbable PGA tensioning suture is integrated into the mesh, protected by two overlapping plastic sheaths. During the procedure, the surgeon attaches the dilating connectors to the keyed ends of the spiral needle passers to facilitate placement of the pubourethral sling. The spiral needle design distinguishes it from the predicate curved needle. The device is used in a clinical/surgical setting by a physician to provide mechanical support to the urethra, potentially benefiting patients by correcting urethral hypermobility or intrinsic sphincter deficiency.

## Clinical Evidence

No clinical data provided; substantial equivalence is supported by risk analysis and bench testing.

## Technological Characteristics

Materials: Polypropylene (sling mesh), stainless steel (needle passers), absorbable PGA (tensioning suture). Form factor: 22cm spiral needle passers. Energy source: None (manual surgical instrument). Sterilization: Sterile, single-use.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- SPARC Sling System ([K011251](/device/K011251.md), [K013355](/device/K013355.md), [K020663](/device/K020663.md), [K021263](/device/K021263.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a black and white drawing of an abstract shape. The shape is composed of a triangle on the left and a series of vertical lines on the right. The triangle is filled with diagonal lines, and the vertical lines are of varying lengths.

AMERICAN MEDICAL SYSTEMS

# 510(k) SUMMARY

| Submitter's Name:            | American Medical Systems, Inc.                              |
|------------------------------|-------------------------------------------------------------|
| Address:                     | 10700 Bren Road West<br>Minnetonka, MN 55343                |
| Tel:                         | 952-933-4666                                                |
| Fax:                         | 952-930-6496                                                |
| Contact Person:              | David Worrell                                               |
| Date of Summary Preparation: | October 15th, 2002                                          |
| Device Common Name:          | Surgical Mesh, Sling, Urethral Sling                        |
| Device Trade Name:           | MONARC™ Sling System                                        |
| Device Classification Name:  | Surgical Mesh, polymeric                                    |
| Predicate Device:            | SPARC™ Sling System - K011251, K013355,<br>K020663, K021263 |

### Device Description

The MONARC Sling System is a sterile, single use procedure kit consisting of two stainless steel, 22cm spiral needle passers (also called insertion tools) and one AMS polypropylene sling mesh with attached dilating connectors. The dilating connectors attach to the keyed ends of the MONARC needle passers during the procedure to facilitate sling placement. A fixed absorbable PGA tensioning suture runs through the middle of the sling mesh. Two plastic sheaths overlap in the center of the sling mesh and protect it during placement.

### Indications for Use

The MONARC Sling System is intended for the placement of a pubourethral sling for the treatment of female stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency.

## Comparison to Predicate Device

The fundamental scientific technology of the MONARC device is unchanged from the predicate device. The primary change to the device is the shape of the needle. The needle will have a spiral shape instead of a curved shape. The sling mesh and absorbable suture remain the same. The devices that were the subject of K020663 and K021263 will continue to be marketed.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a black and white drawing of a building. The building has a slanted roof on the left side and a flat roof on the right side. The building is made of vertical lines and horizontal lines.

#### AMERICAN MEDICAL SYSTEMS

## Supporting Information

A risk analysis for the MONARC device and the verification and validation activity reported in this 510(k) application substantiate equivalence to the predicate and did not raise any new questions of safety or effectiveness.

## Conclusion

The MONARC device is substantially equivalent to the predicate with respect to intended use and technological characteristics.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

American Medical Systems, Inc. Mr. David WorreIl Sr. Regulatory Affairs Specialist 10700 Bren Road, West MINNETONKA MN 55343

SEP 2 8 2012

Re: K023516 Trade/Device Name: MonarcTM Sling System Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: OTN Dated: October 17, 2002 Received: October 21,2002

Dear Mr. Worrell:

This letter corrects our substantially equivalent letter of November 19, 2002.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 -

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Sincerely yours,

Benjamin R. Fuchs

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# INDICATIONS FOR USE ENCLOSURE

| 510(k) Number: |  |
|----------------|--|
|----------------|--|

Ko2.356.

Device Name:

MONARC™ Sling System

Indications for Use:

The MONARC™ Sling System is intended for the placement of a pubourethral sling for the treatment of female stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency.

### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-off) Division of General and Restorative Devices

510(k) Number K023516

Prescription Use
(Per 21 CFR801.109)

OR

Over the Counter Use

Muriam C. Picont

(Division Sign-Off) Division of General, Restorative and Neurological Devices

- 1 ()(k) Number _____________________________________________________________________________________________________________________________________________________________

3

---

**Source:** [https://fda-staging.innolitics.com/device/K023516](https://fda-staging.innolitics.com/device/K023516)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com