← Product Code [BTM](/productcode/BTM) · K023494

# NELLCOR INDGO MANUAL RESUSCITATOR (K023494)

_Nellcor Puritan Bennett, Inc. · BTM · Feb 7, 2003 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K023494

## Device Facts

- **Applicant:** Nellcor Puritan Bennett, Inc.
- **Product Code:** [BTM](/productcode/BTM.md)
- **Decision Date:** Feb 7, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 868.5915
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Nellcor INdGO Manual Resuscitator is a portable, nonsterile, single-patient use device intended for use on patients requiring manual ventilatory support. During artificial ventilation, the resuscitator can be operated from ambient air or provide oxygen-enriched air using the oxygen accumulator and connecting the supply tubing to a metered oxygen source. The Nellcor INdGO Manual Resuscitator is available with or without an integrated CO2 detector. The integrated CO2 detector can be used to assist the verification of tube placement during endotracheal or nasotracheal intubation. The integrated CO2 detector detects approximate ranges of CO2 by color comparison. The Nellcor INdGO Manual Resuscitator is available in two sizes and is for use on the following patient populations: Adult size is for use on adults > 30kg. Child size is for use on children 10 30 kg. The Nellcor INdGO Manual Resuscitator is intended for use by qualified healthcare professionals in any environment where pulmonary support resuscitation is indicated, such as hospital, transport, mobile, and home. This device is for prescription use only.

## Device Story

Bag-valve-mask manual resuscitator; delivers ambient or oxygen-enriched air to patients requiring ventilatory support. Optional integrated INdCAP CO2 detector cartridge attached to expiratory port; provides visual color-comparison CO2 range detection to verify airway tube placement. Single-patient use; non-sterile; latex-free; available in adult (>30kg) and child (10-30kg) sizes. Operated by qualified healthcare professionals in hospital, transport, mobile, or home settings. Provides manual pulmonary support; assists clinicians in confirming correct endotracheal/nasotracheal tube placement via colorimetric CO2 monitoring. Benefits include portable respiratory support and immediate visual feedback on tube positioning.

## Clinical Evidence

Bench testing only. Performance consistent with ASTM F920 and ISO 8382 standards.

## Technological Characteristics

Bag-valve-mask manual resuscitator; non-sterile; latex-free. Includes double-swivel elbow, oxygen accumulator, and optional colorimetric CO2 detector cartridge. Complies with ASTM F920 and ISO 8382. Mechanical operation; no software or electronic components.

## Regulatory Identification

A manual emergency ventilator is a device, usually incorporating a bag and valve, intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.

## Predicate Devices

- DMR²Plus™ Disposable Manual Resuscitator with Integrated CO2 Detection ([K973419](/device/K973419.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K023494

easanton, CA 9

Tele: 925 463-4427
Fax: 925 463-4020

tuco Healthcare

Nellcor

# 510(k) Summary

FEB 0 7 2003

| Submitted by:                             | Nellcor Puritan Bennett, Incorporated<br>4280 Hacienda Drive<br>Pleasanton, CA 94588                                         |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Company Contact:                          | Gina To<br>Senior Regulatory Affairs Project Manager<br>(925) 463-4427<br>(925) 463-4020 - FAX                               |
| Date Summary<br>Prepared:                 | February 5, 2003                                                                                                             |
| Trade Name:                               | Nellcor INdGO Manual Resuscitator                                                                                            |
| Common/Usual Name:                        | Disposable Manual Resuscitator                                                                                               |
| Classification Name:                      | Manual Emergency Ventilator, BTM per<br>21CFR§868.5915                                                                       |
| Legally Marketed<br>(Unmodified) Devices: | DMR²Plus™ Disposable Manual Resuscitator with<br>Integrated CO2 Detection, Nellcor Puritan Bennett, Inc.<br>510(k) # K973419 |

# Device Description

Nellcor INdGO Manual Resuscitator is a bag-valve-mask device with the capability of delivering supplemental oxygen. During artificial ventilation, the resuscitator can be operated from ambient air or provide oxygen-enriched air by connecting the supply tubing to a metered oxygen source.

The Nellcor INdGO Manual Resuscitator is available with and without integrated breath-to-breath CO2 detection. The integrated CO2 detector model incorporates an integral CO2 detector cartridge called INdCAP™ attached to the expiratory port of the resuscitator.

The Nellcor INdGO Manual Resuscitator is single-patient use, non-sterile, latex-free, and available in two sizes: adult and child. It has a double-swivel elbow, three oxygen accumulator styles, and comes in a variety of configurations.

# Indications for Use

The Nellcor INdGO Manual Resuscitator is a portable, nonsterile, single-patient use device intended for use on patients requiring manual ventilatory support. During artificial ventilation, the resuscitator can be operated from ambient air or provide oxygen-enriched air using the oxygen accumulator and connecting the supply tubing to a metered oxygen source. The Nellcor INdGO Manual Resuscitator is available with or without an integrated CO2 detector. The 

{1}------------------------------------------------

integrated CO2 detector can be used to assist the verification of tube placement during endotracheal or nasotracheal intubation. The integrated CO2 detector detects approximate ranges of CO2 by color comparison.

The Nellcor INdGO Manual Resuscitator is available in two sizes and is for use on the following patient populations:

- Adult size is for use on adults > 30kg. .
- Child size is for use on children 10 30 kg. .

The Nellcor INdGO Manual Resuscitator is intended for use by qualified healthcare professionals in any environment where pulmonary support resuscitation is indicated, such as hospital, transport, mobile, and home. This device is for prescription use only.

# Summary of Technological Characteristics of the Device Compared to the Legally Marketed (Unmodified) Device

The method of operation, technological characteristics, and CO2 detector technology of the Nellcor INdGO Manual Resuscitator remain the same as the above referenced predicate device. The only modifications relate to the elbow, NRV housing, check valve cartridge, and canister in the CO2 cartridge.

# Testing

Performance and testing are consistent with the requirements for this device type specified by ASTM F920 and ISO 8382.

# Conclusions

The design modifications and results of testing do not raise new questions of safety or effectiveness when compared to the legally marketed predicate device.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings.

FEB 0 7 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Gina To Nellcor Puritan Bennett, Incorporated 4280 Hacienda Drive Pleasanton, California 94588

Re: K023494

Trade/Device Name: Nellcor® INdGO™ Manual Resuscitator Regulation Number: 21 CFR 868.5915 Regulation Name: Manual Emergency Resuscitator Regulatory Class: II Product Code: BTM Dated: January 15, 2003 Received: January 16, 2003

Dear Ms. To:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice

{3}------------------------------------------------

Page 2 -- Ms. Gina To

requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Susan Runner

Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): K023494

Nellcor INdGO Manual Resuscitator Device Name:

#### Indications For Use:

The Nellcor INdGO Manual Resuscitator is a portable, nonsterile, single-patient use device intended for use on patients requiring manual ventilatory support. During artificial ventilation, the resuscitator can be operated from ambient air or provide oxygen-enriched air using the oxygen accumulator and connecting the supply tubing to a metered oxygen source. The Nellcor INdGO Manual Resuscitator is available with or without an integrated CO2 detector. The integrated CO2 detector can be used to assist the verification of tube placement during endotracheal or nasotracheal intubation. The integrated CO2 detector detects approximate ranges of CO2 by color comparison.

The Nellcor INdGO Manual Resuscitator is available in two sizes and is for use on the following patient populations:

- Adult size is for use on adults > 30kg. .
- Child size is for use on children 10 30 kg. .

The Nellcor INdGO Manual Resuscitator is intended for use by qualified healthcare professionals in any environment where pulmonary support resuscitation is indicated, such as hospital, transport, mobile, and home. This device is for prescription use only.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

|                | <div> <i>(Division Sign-Off)</i> </div>                                            |
|----------------|------------------------------------------------------------------------------------|
|                | Division of Anesthesiology, General Hospital,<br>Infection Control, Dental Devices |
| 510(k) Number: | K023494                                                                            |

(Optional Format 3-10-98)

Prescription Use _ (Per 21 CFR 801.109) ﺬ

---

**Source:** [https://fda-staging.innolitics.com/device/K023494](https://fda-staging.innolitics.com/device/K023494)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
