← Product Code [GCT](/productcode/GCT) · K023468

# ENDOSCOPIC LIGHT SOURCE XL180/L3 (K023468)

_World of Medicine Lemke GmbH · GCT · Jan 14, 2003 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K023468

## Device Facts

- **Applicant:** World of Medicine Lemke GmbH
- **Product Code:** [GCT](/productcode/GCT.md)
- **Decision Date:** Jan 14, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Endoscopic Light Source XL180/L3 is intended to be used with fiber optic endoscopes to provide illumination of body cavities, hollow organs and canals during endoscopic procedures.

## Device Story

Endoscopic Light Source XL180/L3 provides illumination for endoscopic diagnostic and surgical procedures; utilizes 180 W xenon lamp (6000 °K color temperature); connects to fiber optic endoscopes via fiber optic cable; brightness adjusted manually by clinician; used in clinical settings; benefits patient by enabling visualization of body cavities, hollow organs, and canals.

## Clinical Evidence

Bench testing only. Device complies with IEC 60601-1, IEC 60601-1-2, and UL2601-1 standards.

## Technological Characteristics

180 W xenon lamp light source; 6000 °K color temperature; ~500-hour lamp life; manual brightness adjustment; fiber optic cable interface; electrical safety per IEC 60601-1, IEC 60601-1-2, and UL2601-1.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Endoscopic Light Source XL202/L3 ([K020889](/device/K020889.md))
- Endoscopic Light Source XL300/L5 ([K021717](/device/K021717.md))
- Karl Storz Xenon Light Source Model 201320-20 ([K934559](/device/K934559.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K023 468

Pg 1 of 2

### 510(K) SUMMARY Endoscopic Light Source XL180/L3

JAN 1 4 2003

#### I. Submitter:

WORLD OF MEDICINE Lemke GmbH Danziger Strasse 21 82194 Gröbenzell Germany

#### II. Device Names:

- Classification Name: 1.
- 2. Common or Usual Name:
- Proprietary Name: 3.

Accessory to an Endoscope Endoscopic Light Source Endoscopic Light Source

XL180/L3

#### III. Classification:

Class II. The device is described in 21 C.F.R. § 876.1500. The product code for the device is GCT.

#### IV. Predicate Device:

- Endoscopic Light Source XL202/L3 (K020889) manufactured by ● WORLD OF MEDICINE Lemke GmbH
- Endoscopic Light Source XL300/L5 (K021717) manufactured by . WORLD OF MEDICINE Lemke GmbH
- Karl Storz Xenon Light Source Model 201320-20 (K934559) ● manufactured by Karl Storz Endoscopy Imanging, Inc.

#### V. Intended Use:

The Endoscopic Light Source XL180/L3 is intended to be used with fiber optic endoscopes to provide illumination of body cavities, hollow organs and canals during endoscopic procedures.

#### VI. Device Description:

The Endoscopic Light Source XL180/L3 uses a 180 W xenon lamp to provide illumination during endoscopic diagnostic and surgical procedures through a fiber optic cable, which is connected to the device. Brightness can be adjusted manually. The color temperature of the xenon lamp is approximately 6000 °K and the lamp life is approximately 500 hours.

{1}------------------------------------------------

#### VII. Substantial Equivalence:

The Endoscopic Light Source XL180/L3 described in this notification is similar in design and technological characteristics to the Endoscopic Light Source XL202/L3 (K020889) and to the Endoscopic Light Source XL300/L5 (K021717) both manufactured by WORLD OF MEDICINE Lemke GmbH and the Karl Storz Xenon Light Source Model 201320-20 (K934559) manufactured by Karl Storz Endoscopy Imanging, Inc.

Both the Endoscopic Light Source XL180/L3 and the predicate devices are intended to provide illumination of body cavities, hollow organs and canals during endoscopic diagnostic and surgical procedures. In addition, both the Endoscopic Light Source XL180/L3 and the predicate device Endoscopic Light Source XL202/L3 (K020889) are designed to be used with a 180 W Xenon lamp.

The differences between the Endoscopic Light Source XL180/L3 and the predicate devices are minor and raise no new questions of safety and effectiveness. Accordingly, WORLD OF MEDICINE Lemke GmbH believes that the Endoscopic Light Source XL180/L3 is substantially equivalent to the predicate devices currently on the market.

#### VIII. Performance Data:

The Endoscopic Light Source XL180/L3 complies with the International Standard IEC 60601-1. IEC 60601-1-2 and conforms to the Medical Device Directive 93/42/EEC. In addition, the device will meet the requirements of the Underwriters Laboratories Standard UL2601-1.

Signed:

S. Reads

Susanne Raab Official Correspondent

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three tail feathers. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## WORLD OF MEDICINE LEMKE GmbH c/o Ms. Susanne Raab Official Correspondent 91 Trowbridge Street CAMBRIDGE MA 02138

JAN 1 4 2003

Re: K023468

Trade/Device Name: Endoscopic Light Source XL 180/L3 Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: 78 GCT Dated: September 24, 2002 Received: October 16, 2002

Dear Ms. Raab:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K023468

# STATEMENT OF INDICATIONS FOR USE

K023468

APPLICANT:

WORLD OF MEDICINE Lemke GmbH

510(K) NUMBER (if known):

DEVICE NAME:

Endoscopic Light Source XL180/L3

INDICATIONS FOR USE:

The Endoscopic Light Source XL180/L3 is intended to be used with fiber optic endoscopes to provide illumination of body cavities, hollow organs and canals during endoscopic procedures.

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Per 21 C.F.R. § 801.109)

(Optional Format 1-2-96)

David A. Ingram

Prescription Use . (Per 21 CFR 801.109)

---

**Source:** [https://fda-staging.innolitics.com/device/K023468](https://fda-staging.innolitics.com/device/K023468)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).