← Product Code [ITI](/productcode/ITI) · K023426

# TRX VERTICAL POSITIONER (K023426)

_Motion Concepts · ITI · Nov 21, 2002 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K023426

## Device Facts

- **Applicant:** Motion Concepts
- **Product Code:** [ITI](/productcode/ITI.md)
- **Decision Date:** Nov 21, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3860
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The TRX Vertical Positioner is appropriate for use by any individual who drives a power wheelchair and who desires or requires a change of position without having to utilize the services of an attendant. Needs for position changes include: - All positioning benefits associated with the tilt/recline product -. > Comfort As with any individual able-bodied or disabled changes in position are necessary to maintain a state of comfort. > Repositioning Individuals without adequate upper-body stability can be tilted to allow gravity to hold them in position. > Pressure relief Individuals who wish, from time to time, to redistribute pressures from one area of the body to another, can do so by tilting, reclining or vertical positioning. By changing the individual's orientation in space, pressures caused by gravity will shift. - Positioning/Versatility Individuals are able to reach higher elevations, increasing their . range of motion and accessibility. The TRX Vertical Positioner is used to add the vertical positioning function to the tilt, recline, and elevating functions already available in with the TRX Power Positioning System defined in K021264.

## Device Story

TRX Vertical Positioner is an accessory for power wheelchairs; adds vertical positioning to existing tilt, recline, and elevating functions. Input: user-operated switches (push buttons/toggles) or wheelchair joystick. Operation: coordinates elevating and reclining motion with anterior tilt to transition user from seated to vertical position. Output: physical change in user orientation. Used by wheelchair-bound individuals in various settings. Safety features: drive lock-out (prevents driving if tilted >20° or elevated >1/2"), front anti-tipper extensions, and active track suspension lockout. Benefits: independent pressure relief, improved comfort, and increased reach/accessibility. Installed by qualified dealers.

## Clinical Evidence

No clinical data. Bench testing only performed to ensure stability and safety of the power wheelchair system following the addition of the vertical positioner.

## Technological Characteristics

Materials: steel and aluminum, powder coated. Energy: 24V electrical system with fuses and current limiters. Actuation: electromechanical, controlled via switches or joystick. Features: drive lock-out, anti-tipper extensions, suspension lockout. Connectivity: none. Sterilization: N/A.

## Regulatory Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- Levo Mobile LCM ([K963817](/device/K963817.md))

## Reference Devices

- TRX Power Positioning System ([K021264](/device/K021264.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KOZ344

Image /page/0/Picture/1 description: The image shows a logo with a circular design on the upper left and the word "Motion" in bold, larger font below it. To the right and slightly above the word "Motion" is the word "Concepts" in a smaller, slanted font. The logo appears to be for a company or product named "Motion Concepts."

NOV 21 2002

# 510(k) Summary

Submitter: Motion Concepts

- 84 Citation Dr. Address: Concord, ON L4K 3C1 Canada
- Phone: (905) 695-0134
- Contact: David Ciolfe
- dciolfe @ motionconcepts.com E-mail:
- October 1, 2002 Date Prepared:
- Device Trade Name: TRX Vertical Positioner
- Common Name: Power Positioning System
- Classification Name: Powered Wheelchair
- ldentification of Predicate Device: Levo Mobile LCM (K963817)

{1}------------------------------------------------

### Description of Device:

The TRX Vertical Positioner is used to add the vertical positioning function to the tilt, recline. and elevating functions already available in with the TRX Power Positioning System defined in K021264. The Vertical Positioner functions by coordinating the elevating and reclining motion with an anterior tilt motion to change the user from an upright seated position to a vertical position and back again.

The TRX Vertical Positioner is assembled using primarily steel and aluminum, powder coated parts.

The vertical positioning is be activated using TRx switches consisting of either push buttons or togale switches, or through the wheelchair joystick.

Additional safety features include a drive lock-out which prevents the user from driving the power chair while tilted beyond a pre-set limit of 20° from the vertical or elevated beyond 1/2". The system also includes extensions for the existing front anti-tipper wheels to provide additional front stability, and a lockout for the active track suspension in the vertical positioning mode. Electrical components are a maximum of 24 volts with a fuse between the batteries and the relay box and current limiter built into the relay box. Stability of the TRX Vertical Positioner was tested in our facility to ensure that the safety of the power wheelchair was not compromised by the addition of the system.

{2}------------------------------------------------

## Statement of Indications for Use:

The TRX Vertical Positioner is appropriate for use by any individual who drives a power wheelchair and who desires or requires a change of position without having to utilize the services of an attendant. Needs for position changes include:

- All positioning benefits associated with the tilt/recline product -.
	- > Comfort As with any individual able-bodied or disabled changes in position are necessary to maintain a state of comfort.
	- > Repositioning Individuals without adequate upper-body stability can be tilted to allow gravity to hold them in position.
	- > Pressure relief Individuals who wish, from time to time, to redistribute pressures from one area of the body to another, can do so by tilting, reclining or vertical positioning. By changing the individual's orientation in space, pressures caused by gravity will shift.
- Positioning/Versatility Individuals are able to reach higher elevations, increasing their . range of motion and accessibility.

The TRX Vertical Positioner is used to add the vertical positioning function to the tilt, recline, and elevating functions already available in with the TRX Power Positioning System defined in K021264.

Motion Concepts makes no claims as to the therapeutic effectiveness of the products. Our only claims relate to the ability of the products to provide safe and reliable powered repositioning on the equipment onto which they are installed. TRX Vertical Positioners are to be installed ONLY by qualified Dealers.

## Substantial Equivalence Comparison:

1100315 Ontario Limited claims that our product is substantially equivalent to the Mobile LCM as manufactured by Levo. The pre-existing device is classified as Class I as listed in 21 CFR Parts 890.3860. The Levo 510(k) document control number is K963817.

Both the TRx Vertical Positioner and the Levo Mobile LCM systems are devices that are used on power wheelchairs in order to provide the user with the capability to change his/her position from a seated position to a vertical position and back. Both products use similar components. Both are fabricated from steel and aluminum and have a generally similar mechanical operation. Both systems combine the anterior tilting and reclining action to achieve a vertical position.

The following is a comparison of the Mobile LCM and the TRx Vertical Positioner.

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is a stylized representation of three human profiles facing to the right, with flowing lines suggesting movement or progress. The overall design is simple and professional, reflecting the department's role in public health and human services.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 2.1 2002

Motion Concepts David Ciolfe Operations Manager 84 Citation Drive Concord, Ontario L4K 3C1 Canada

Re: K023426

Trade/Device Name: TRX Vertical Positioner Regulation Number: 890.3860 Regulation Name: Powered Wheelchair (accessory) Regulatory Class: Class II Product Code: ITI Dated: November 4, 2002 Received: November 8, 2002

Dear Mr. Ciolfe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the

{4}------------------------------------------------

quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Muriam C. Provost

fel Celia M. Witten, Ph.D., MD Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known):__KO23426

Device Name: TRx Vertical Positioner

#### Statement of Indications for Use:

The TRX Vertical Positioner is appropriate for use by any individual who drives a power wheelchair and who desires or requires a change of position without having to utilize the services of an attendant. Needs for position changes include:

- All positioning benefits associated with the tilt/recline product -●
	- Comfort As with any individual able-bodied or disabled changes in position are necessary A to maintain a state of comfort.
	- > Repositioning - Individuals without adequate upper-body stability can be tilted to allow gravity to hold them in position.
	- Pressure relief Individuals who wish, from time to time, to redistribute pressures from one A area of the body to another, can do so by tilting, reclining or vertical positioning. By changing the individual's orientation in space, pressures caused by gravity will shift.
- . Positioning/Versatility -- Individuals are able to reach higher elevations, increasing their range of motion and accessibility.

The TRX Vertical Positioner is used to add the vertical position to the tilt, recline, and elevating functions already available in with the TRX Power Positioning System defined in K021264.

Motion Concepts makes no claims as to the therapeutic effectiveness of the products. Our only claims relate to the ability of the products to provide safe and reliable powered repositioning on the equipment onto which they are installed. TRX Vertical Positioners are to be installed ONLY by qualified Dealers.

The above indications for use are identical to those of the Levo Mobile LCM defined in K963817 to which we are claiming substantial equivalence.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

Miriam C. Provost

Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number K023426

---

**Source:** [https://fda-staging.innolitics.com/device/K023426](https://fda-staging.innolitics.com/device/K023426)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).