The Leforte System Screws are intended for fracture fixation as well as reconstruction and stabilization of small bones (i.e. toe, foot, finger, hand, etc) and the mandibulofacial (i.e. mandible and maxilla) bones.
Device Story
Leforte System Bone Screws are metallic orthopedic implants used for fracture fixation, reconstruction, and stabilization of small bones and mandibulofacial structures. The system includes common, self-drilling, and dual-head screws available in micro, mid, mini, and maxi classes. Manufactured from titanium alloy (ASTM F136-98), these devices are provided non-sterile and require steam sterilization by the user prior to surgical implantation. The screws are intended for use by surgeons in clinical settings to provide mechanical fixation of bone segments. By securing bone fragments, the device facilitates healing and stabilization of fractures or reconstructive sites. The device is designed to be functionally equivalent to existing titanium alloy bone screw systems.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing, material composition (ASTM F136-98), and comparison of design characteristics to legally marketed predicate devices.
Technological Characteristics
Material: Titanium alloy (ASTM F136-98). Types: Common, self-drilling, and dual-head screws. Classes: Micro, mid, mini, and maxi. Sterilization: Non-sterile, requires steam sterilization. Design: Metallic bone fixation fastener.
Indications for Use
Indicated for fracture fixation, reconstruction, and stabilization of small bones (toe, foot, finger, hand) and mandibulofacial bones (mandible, maxilla).
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
## KC23365 Page V3
## EXHIBIT 2. 510(K) SUMMARY
APR 1 8 2003
Jeil Medical Corporation 808, ACE Techno-Tower, 3rd 197-48, Kuro 3Dong Kuro-ku, Seoul, KOREA Telephone : 82-2-2109-6781~9 Fax: 82-2-2109-6791 January 30, 2002 Contact: N.K.Kim, R&D Director 510(k) Summary
- 1. Identification of the Device: Proprietary-Trade Name: Leforte System Bone Screw Classification Name: Screw, Fixation, Bone, Product Code HWC Common/Usual Name: Bone Screw
- 2. Equivalent legally marketed device: Twist off™ Screw Medinov-AMP, K971069, Aesculap Titanium Alloy Bone Screws Aesculp, Inc., K970549, K3 Bone Screw System, Kinetikos Medical Inc., K960533
- 3. Indications for Use (intended use) . The Leforte System Screws are intended for fracture fixation as well as reconstruction and stabilization of small bones (i.e. toe, foot, finger, hand, etc) and the mandibulofacial (i.e. mandible and maxilla) bones
- 4. Description of the Device: Leforte system bone screw is made of Titanium alloy (ASTM F136) and consists of common bone screw, self-drilling screw (auto screw) and dual head screw. The common bone screws are classified to 4 classes; Micro screw, Mid screw, Mini screw, Maxi screw. This device is manufactured and intended for small bone (toe, foot, finger, hand etc.) and mandibulofacial (mandible & Maxilla) fracture fixation. Also This device is intended for reconstruction & stabilization of small bone and mandibulofacial (mandible & Maxilla) Jeil Medical Corporation utilizes the state of the art technology and apply the essential requirement of MDD (93/42/EEC) and ISO 14630, 1997 from the device design to manufacturing and QC ...
- 5. Potential Adverse Affects and Complications: (Common to all devices of this type)
{1}------------------------------------------------
- Poor bone formation, Osteoporosis, Osteomyelitis, inhibited revascularization, . or infection can cause loosening, bending, cracking or fracture of the device or premature loss of fixation with the bone, leading to nonunion.
- Nonunion or delayed union which may lead to breakage of the implant. ●
- Migration, bending, fracture or loosening of the implant. .
- . Metal sensitivity, or allergic reaction to a foreign body:
- Decrease in bone density due to stress shielding. .
- Pain, discomfort, or abnormal sensation due to the presence of the device. .
- Increased fibrous tissue response around the fracture site and/or the implan". ●
- Necrosis of bone. ●
- Inadequate healing.
Apart from these adverse effects there are always possible complications of any surgical procedure such as, but not limited to, infection, nerve damage, and pain which may not be related to the implant
| Device Name | Twist off™<br>Screw | Aesculap<br>Titanium<br>Alloy Bone<br>Screws | K3 Bone Screw<br>System | Le forte System<br>(Bone Screw) |
|----------------------------------|----------------------------------------------------------|----------------------------------------------|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Device<br>Classification<br>Name | No Comment | Screw, Fixation,<br>Bone<br>888.3030 | Smooth or Threaded<br>Bone Fixation<br>Fastener | Screw, Fixation, Bone<br>888.3040 |
| Applicant | Medinov-AMP | Aesculp, Inc. | KMI<br>(Kinetikos Medical<br>Inc.) | Jeil Medical<br>Corporation |
| 510(K) Number | K971069 | K970549 | K960533 | (This submission) |
| Material | Titanium Alloy | Titanium Alloy | Titanium Alloy<br>(Ti-6Al-4V ELI<br>Alloy<br>: ASTM F136-84) | Titanium Alloy<br>(ASTM F 136-98) |
| Intended use | Fixation of<br>fracture, fusion<br>of a joint or<br>Bone | Long & small<br>bone fracture<br>fixation | Fixation/stabilization<br>of small bones in<br>hand or small bones<br>in mid & forefoot | Small bone and<br>mandibulofacial<br>(mandible & Maxilla)<br>fracture fixation |
- 6. Comparisons to predicate devices:
{2}------------------------------------------------
| | reconstruction of<br>the carpals,<br>metacarpals &<br>phalanges of the<br>hand | | fractures | reconstruction &<br>stabilization of small<br>bone and<br>mandibulofacial<br>(mandible & Maxilla) |
|-------------------|--------------------------------------------------------------------------------|------------|------------|---------------------------------------------------------------------------------------------------|
| Supplied Sterile? | No Comment | No Comment | No comment | Non sterile, steam<br>sterilize before use |
- 7. Conclusion: In all respects, the Leforte System Bone Screws are the equivalent of currently marked devices. They are made of the same materials and have similar dimensions and characteristics. Potential adverse effects are identical to those of predicate devices. This device is manufactured from material of titanium alloy (ASTM 136-97) that is used generally in this kind of bone screw. Similar devices made from titanium alloy (ASTM 136-97) to this device are manufactured and sold around the world. This device, Le forte system bone screw is substantially equivalent in design, material, intended use and function to the products on the table above. These devices are certificated by notified body for CE.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the department's name around the perimeter. In the center of the seal is a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it and a pair of wings at the top.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20:350
APR 1 8 2003
Jeil Medical Corporation c/o Mr. Daniel Kamm, P.E. Regulatory Engineer Kamm & Associates P.O. Box 7007 Deerfield, Illinois 60015
Re: K023365
Trade Name: LeForte System Screws Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: HWC Dated: February 3, 2003 Received: February 4, 2003
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
## Page 2 - Mr. Daniel Kamm, P.E.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Sincerely yours,
Mark N Millman
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
j) Indications for Use
K023365 510(k) Number_
Device Name: LeForte System Screws (various models)
"The Leforte System Screws are intended for fracture fixation as well as reconstruction and stabilization of small bones (i.e. toe, foot, finger, hand, etc) and the mandibulofacial (i.e. mandible and maxilla) bones."
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use × OR (Per 21 CFR 801.109)
Over the Counter Use _________________________________________________________________________________________________________________________________________________________
Mark N Mlkern
for Division Sign Off
(Division Sign-Off) Division of General, Restorative and Neurological Devices
10(k) Number K023365
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.