# ECHOPULSE MUSCLE STIMULATOR SYSTEM, MODEL 800 (K023230)

_Western Clinical Technology Systems, Inc. · IPF · May 5, 2003 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K023230

## Device Facts

- **Applicant:** Western Clinical Technology Systems, Inc.
- **Product Code:** [IPF](/productcode/IPF.md)
- **Decision Date:** May 5, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5850
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

Relaxation of muscle spasms

## Device Story

EchoPulse Muscle Stimulator System (Models 800, 400, 100) is an eight-channel, synchronous, biphasic, powered muscle stimulator. Device delivers electrical stimulation to patient skin via 3-inch diameter Carbonflex electrodes and disposable sponge fabric wetable pads. System produces maximum output of 105 V and 210 mA at 500 ohms with a pulse width of 180 usec. Device is rack-mountable, weighing 27.25 pounds. Intended for clinical use to relax muscle spasms. Healthcare providers apply electrodes to the patient and operate the device to deliver stimulation, which assists in muscle relaxation. Benefit is the therapeutic management of muscle spasms.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Powered muscle stimulator; 8-channel, synchronous, biphasic output. Max output: 105 V, 210 mA at 500 ohms; pulse width 180 usec. Components: 3-inch Carbonflex electrodes, disposable sponge fabric wetable pads. Dimensions: 19"W x 5.25"H x 12.5"D. Rack-mountable. Class II device (21 CFR 890.5850).

## Regulatory Identification

A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

## Predicate Devices

- Bio-Stym 250, Microvas Technologies, Inc. ([K891987](/device/K891987.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

\$\kappa023230\$

MAY 0 5 2003

## 510(k) SUMMARY

This summary is provided in accordance with the Safe Medical Device Act of 1990 (SMDA). The information provided in the 510(k) Premarket Notification was in accordance with 21 CFR 807.87.

- 1. Applicant, Official Correspondence and Owner of 510(k)

| Applicant/Owner:        | Western Clinical Technology Systems, Inc.<br>6737 Owen Hill Road<br>College Grove, TN 37046<br>(615) 368-2426                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Correspondent:          | W.T. Workman, MS, CHT<br>18111 Copper Ridge Drive<br>San Antonio, TX 78259<br>(210) 490-6999                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| 2. Trade Name:          | EchoPulse Muscle Stimulator System, Model 800                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| 3. Common Name:         | Powered muscle stimulator (89IPF)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| 4. Classification:      | 21 CFR 890.5850 Powered Muscle Stimulator, Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| 5. Predicate Device:    | Bio-Stym 250, Microvas Technologies, Inc. (K891987)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| 6. Device Description:  | The EchoPulse 800 is an eight channel, synchronous,<br>biphasic, powered muscle stimulator. It produces a<br>maximum output voltage of 105 V at 500 ohm, a maximum<br>output current of 210 mA at 500 ohm and has a pulse width<br>of 180 usec. The 3 pin connector leads are each 72 inches<br>long. The electrodes are 3 inch diameter Carbonflex<br>electrodes and use 3 inch disposable sponge fabric wetable<br>pads for application on the skin. The EchoPulse 800<br>measures 19 inches wide, 5 1/4 inches tall, and 12 1/2 inches<br>deep. It weighs 27 1/4 pounds and is rack mountable. |
| 7. Indications for Use: | Relaxation of muscle spasms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate: Boulevard Rockville MD 20850

Western Clinical Technology Systems, Inc. c/o Mr. W.T. Workman Workman Hyperbaric Services, Inc. 18111 Copper Ridge Drive San Antonio, Texas 78259-3612

MAY 0 5 2003

Re: K023230

Trade/Device Name: EchoPulse Muscle Stimulator System - Models 800, 400 and 100 Regulation Numbers: 21 CFR 890.5850 Regulation Names: Powered muscle stimulator Regulatory Class: Class II Product Codes: IPF Dated: February 3, 2003 Received: February 4, 2003

Dear Mr. Workman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to, devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{2}------------------------------------------------

Page 2 - Mr. W.T. Workman

CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659_ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address

http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Mark N. Melkerson

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Page _1_of _1_

KOZ3230 510(k) Number (if known):

Device Name: EchoPulse Powered Muscle Stimulator, Models 800, 400, 100

Indications for Use:

1. Relaxation of muscle spasms

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IS NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescripton Use (Per 21 CFR 801.109)

OR
Over-the-Counter Use
(Optional Format 1-2-96)

for
(Division Sign-Off)

Division of General, Restorative and Neurological Devices

510(k) Number K023230

---

**Source:** [https://fda-staging.innolitics.com/device/K023230](https://fda-staging.innolitics.com/device/K023230)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).