← Product Code [BSP](/productcode/BSP) · K023218

# MODIFICATION TO PLEXOLONG SETS (K023218)

_Pajunk GmbH · BSP · Mar 24, 2003 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K023218

## Device Facts

- **Applicant:** Pajunk GmbH
- **Product Code:** [BSP](/productcode/BSP.md)
- **Decision Date:** Mar 24, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 868.5150
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic

## Indications for Use

The PAJUNK Plexolong sets consist of Pajunk Unipolar conduction needle with a plastic cannula placed over the needle, or a Tuohy conduction needle. They are intended for delivery of continuous conduction anesthesia of peripheral nerves and plexus for up to 72 hours. Continuous delivery for up to 72 hours is accomplished using the Polyamide indwelling catheter. An electrical stimulus may be applied to the conduction needle via a cable and connector to assist the physician pinpoint the area of application.

## Device Story

Plexolong sets consist of sterile, single-use Tuohy or Unipolar conduction needles, Polyamide indwelling catheters, tubing, adapters, and filters. Used for continuous peripheral nerve and plexus anesthesia delivery for up to 72 hours. Physician applies electrical stimulus via cable/connector to conduction needle to assist in pinpointing application site. Output is physical delivery of anesthetic agent via catheter. Used in clinical settings by physicians. Benefits include localized, continuous pain management.

## Clinical Evidence

Bench testing only. Device performance verified through conformance to recognized consensus standards including DIN 13090/ISO 594 (Luer fittings), DIN 17442/ISO 9626 (steel), DIN EN 550/556 (sterilization), and DIN EN 1441 (risk analysis). No clinical data provided.

## Technological Characteristics

Components: stainless steel needles, Polyamide catheters, polypropylene containers. Sensing: electrical nerve stimulation via cable/connector. Connectivity: none. Sterilization: ETO. Standards: DIN 13090/ISO 594, DIN 17442/ISO 9626, DIN EN 550, DIN EN 556, DIN EN 1707, DIN EN 1441.

## Regulatory Identification

An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.

## Predicate Devices

- Plexolong Sets ([K013041](/device/K013041.md))
- B Braun Touhy needle ([K813186](/device/K813186.md))

## Reference Devices

- Pajunk Anesthesia Sets ([K000722](/device/K000722.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(k) Summary of Safety and Effectiveness

#### Submitter Information:

PAJUNK GmbH AM Holzplatz 5-7 78187 Geisingen Germany

USA Contact:Mr. Burk A. Brandt CE Consultancy, Inc. 5010 NW Crescent Vly Dr Corvallis, OR 97330 - USA

Phone: (541) 752-3953 Fax : (541) 752-3972

#### Device Name:

Trade Name: Plexolong Sets Common Name: Anesthesia Conduction Kit Classification Name: Anesthesia Conduction Kit (Reference, 21CFR, 868.5140, April 1, 2000)

#### Predicate Device:

These Plexus Anesthesia sets consist of a Paiunk Touby needle (with nerve stimulus connector and Luer lock), tubing (with Luer locks)}, catheter (with Teflon coated stylet), catheter adapter.and filter. The set, including the packaging is identical to the Plexolong Sets cleared for market under 510(k) number K013041. except for the Touhy needles with nerve stimulus connector. The Touhy needle is identical to the Unipolar needle, included in these sets, except for the needle tip. The Touhy needle tip configuration is same as the B Braun Touhy needle, cleared for market under 510(k) number K813186 as referenced in their submission number K013610.

The intended use statement, except for a reference to the Tuohy needle, is identical to the intended use statement in K013041. Except for the shape of the needle tip, the materials, construction and manufacturing processes are also identical to the cleared Pajunk Anesthesia Sets.

The contract sterilizer and sterilizing process are identical to those used for the Pajunk Plexolong sets. The packaging materials are also the same as those used to package the Pajunk previously cleared Plexolong sets.

#### Device Description:

The PAJUNK Plexolong sets contain single use sterile and non-pyrogenic needles with tubing. with or without a plastic cannula and catheters intended for delivery of continuous conduction anesthesia of peripheral nerves and plexus for up to 72 hours. Continuous delivery is accomplished using the catheter. An electrical stimulus may be applied to the conduction needle via a cable and connector to assist the physician pinpoint the area of application.

## Intended Use:

The PAJUNK Plexolong sets consist of Pajunk Unipolar conduction needle with a plastic cannula placed over the needle, or a Tuohy conduction needle. They are intended for delivery of continuous conduction anesthesia of peripheral nerves and plexus for up to 72 hours. Continuous delivery for up to 72 hours is accomplished using the Polyamide indwelling catheter. An electrical stimulus may be applied to the conduction needle via a cable and connector to assist the physician pinpoint the area of application.

{1}------------------------------------------------

## Warning:

The Pajunk GmbH needles and puncture sets are not intended for RF ablation or any other type of ablation procedure.

## Technology Characteristics:

The PAJUNK Unipolar needles, which include the physical dimensions, coating, connector, tubing, metal and plastics, have been cleared under 510(k) number K000722. The Touhy needle is identical to the Unipolar needle, included in these sets, except for the shape of the needle tip. The Touhy needle tip configuration is same as the B Braun Touhy needle, cleared for market under 510(k) number K813186 as referenced in their submission number K013610. The Plexolong Sets are supplied in sealed polypropylene containers or polypropylene and styrol paper envelopes that are sealed to assure sterility.

{2}------------------------------------------------

## Summary of Performance Testing:

The PAJUNK Plexolong Sets were designed to conform to the applicable sections of the following recognized consensus standards. The testing included verifying conformance to these standards.

| Standard           | Issue Date | Title                                                                                                                                                                   |
|--------------------|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| DIN 13090/ISO 594  | 08.1984    | Luer fittings w/wo locking feature                                                                                                                                      |
| DIN 13097 Part 1   | 01.1980    | Medical injection cannula                                                                                                                                               |
| DIN 13097 Part 3   | 11.1979    | Medical cannula                                                                                                                                                         |
| DIN 17442/ISO 9626 | 10.1977    | Steel for medical instruments                                                                                                                                           |
| DIN EN 550         | 07.1993    | Sterilization of med. Prod.; Validation & routine controls for<br>sterilization with ETO                                                                                |
| DIN EN 556         | 01.1995    | Sterilization of medical products, requirements for medical<br>products that are labeled "sterile"                                                                      |
| DIN EN 724         | 12.1994    | Guidance on the application of EN29001 and EN46001 for<br>non-active medical products                                                                                   |
| PrEN 868-1         | 10.1996    | Packaging materials for the sterilization of packaged goods.<br>Part 1: general requirements for the validation of the<br>packaging of sterilized end-packaged products |
| DIN EN 868-2       | 03.1993    | Packaging materials for the sterilization of packaged goods.<br>Part 2: sterilization packaging, requirements and tests.                                                |
| DIN EN 980         | 08.1996    | Graphic symbols for marking medical products                                                                                                                            |
| DIN EN 1441        | 08.1994    | Risk analysis for medical products                                                                                                                                      |
| DIN EN 1707        | 01.1997    | 6% Luer cone connections for injection cannula and<br>particular medical equipment                                                                                      |
| DIN EN/ISO 9626    | 06.1995    | Cannula tube of non-rusting steel (SS) for the manufacture of<br>medical products                                                                                       |
| DIN en 30993-1     | 12.1994    | Biological evaluation of medical products - instructions for<br>selection of tests                                                                                      |
| DIN EN 46001       | 12.1993    | Particular requirements for medical products                                                                                                                            |
| DIN 17440          | 09.1996    | Stainless Steels                                                                                                                                                        |
| BS 4843            |            | Single entry IV cannula                                                                                                                                                 |

## Conclusion:

The PAJUNK Plexolong Sets are as safe and effective as the predicate devices when used according to the instructions in the directions for use supplied with the devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, with stylized lines representing its body and wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# MAR 2 4 2003

PAJUNK GmbH C/O Mr. Burk A. Brandt CE Consultancy, Incorporated 5010 NW Crescent Valley Drive Corvallis, Oregon 97330

Re: K023218

Trade/Device Name: Pajunk Plexolong Sets Regulation Number: 868.5140 Regulation Name: Anesthesia Conduction Kit Regulatory Class: II Product Code: CAZ, BSP Dated: February 25, 2003 Received: February 27, 2003

Dear Mr. Brandt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Brandt

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Susan Runner

Susan Runner, DDS, MA Interim Director Division of Anesthesiology. General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# INDICATIONS FOR USE

510(k) Number: ____K023218____________________________________________________________________________________________________________________________________________________

Device Name: Pajunk Plexolong Sets

Indications for use:

The Pajunk Plexolong sets are intended for delivery of continuous conduction anesthesia of peripheral nerves and plexus for up to 72 hours. Continuous delivery for up to 72 hours is accomplished using the Polyamide indwelling catheter. An electrical stimulus may be applied to the conduction needle via a cable and connectors to assist the physician pinpoint the area of application.

## Warning:

The Pajunk GmbH needles and puncture sets are not intended for RF ablation or any other type of ablation procedure.

(Please do not write below this line - continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation

V Prescription Use or Over-The- Counter _________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109)

S.H. Ward

nesthesiology. General Hospital.

510(k) Number:

---

**Source:** [https://fda-staging.innolitics.com/device/K023218](https://fda-staging.innolitics.com/device/K023218)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com