The Reliant BTL 5000 Puls Electrical Stimulation System, BTL 5000 Sono Ultrasound System, and BTL 5000 Combi Combination Electrical Stimulation & Ultrasound systems (and the delivery accessories that are used with them) are indicated for use in the stimulating, relaxing, and/or repeatedly contracting muscles by passing electrical currents through electrodes contacting the affected body area in the medical specialties of physical medicine, general and plastic surgery and neurology and/or for use in applying therapeutic deep heat for selected medical conditions such as relief of pain, muscle spasms, and joint contractures in the medical specialty of physical medicine. The Reliant BTL 5000 Puls Electrical Stimulation System, BTL 5000 Sono Ultrasound System, and BTL 5000 Combi Combination Electrical Stimulation & Ultrasound systems are indicated for the treatment and relaxation of muscle spasms, prevention or retardation of disuse atrophy, increasing local blood circulation, muscle re-education, maintaining or increasing range of motion, immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, symptomatic relief of chronic, intractable pain, management of pain associated with post-traumatic or post-operative conditions using electrical stimulation and are indicated for the application of therapeutic deep heat for the treatment of selected medical conditions such as relief of pain, muscle spasms and joint contractures, relief of pain, muscle spasms and joint contractures that may be associated with adhesive capsulitis, bursitis with slight calcification, myositis, soft tissue injuries, shortened tendons due to past injuries and scar tissues, relief of chronic and subchronic pain and joint contractures resulting from capsular tightness and capsular scarring using ultrasound. The combination ultrasound and electrotherapy modes deliver deep heat for pain management (ultrasound therapy) at the same time that electrical stimulation for either pain management or muscle stimulation is being delivered. Only one electrotherapy mode (TENS, High Voltage, Interferential or Premodulated) can be utilized at a time in conjunction with the ultrasound mode.
Device Story
System console with software/electronics, control panel, and delivery accessories (cables, electrodes, ultrasound sound heads). Operates by delivering electrical currents (TENS, Russian, High Voltage, Microcurrent, Interferential, Premodulated) or therapeutic ultrasound deep heat. Used in physical medicine, surgery, and neurology clinics by healthcare providers. Combination mode allows simultaneous ultrasound and one electrotherapy modality. Output provides muscle stimulation or deep heat therapy to manage pain, spasms, and contractures; improve circulation; and prevent atrophy. Benefits include non-invasive pain management and physical rehabilitation support.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via validation documentation and comparison of technical characteristics and features.
Technological Characteristics
System console with control electronics and software; delivery via electrodes/sponge covers and ultrasound sound heads. Modalities: TENS, Russian, High Voltage, Microcurrent, Interferential, Premodulated, and Ultrasound. Class II device. Connectivity: Standalone console.
Indications for Use
Indicated for patients in physical medicine, general/plastic surgery, and neurology requiring muscle stimulation, relaxation, or contraction; prevention of disuse atrophy; increased local blood circulation; muscle re-education; range of motion maintenance; post-surgical calf stimulation to prevent venous thrombosis; and symptomatic relief of chronic, intractable, post-traumatic, or post-operative pain. Also indicated for therapeutic deep heat application for pain, muscle spasms, and joint contractures associated with adhesive capsulitis, bursitis, myositis, soft tissue injuries, shortened tendons, and capsular tightness/scarring.
Regulatory Classification
Identification
An ultrasound and muscle stimulator for use in applying therapeutic deep heat for selected medical conditions is a device that applies to specific areas of the body ultrasonic energy at a frequency beyond 20 kilohertz and that is intended to generate deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies. The device also passes electrical currents through the body area to stimulate or relax muscles.
Predicate Devices
Chattanooga Group, Inc. Vectra 2C and Vectra 4C (K982317)
Chattanooga Group, Inc. Vectra Pro 2 and Vectra Pro 4 (K982324)
Dynatronics Corporation Dynatron 150 Plus Ultrasound (K935728)
Empi 300 PV Complete Electrotherapy System (K021100)
Submission Summary (Full Text)
{0}------------------------------------------------
## Attachment 15 510(k) Summary Statement for the Reliant BTL 5000 Puls Electrical Stimulation, BTL 5000 Sono Ultrasound, & BTL 5000 Combi Combination Electrical Stimulation & Ultrasound Systems
#### I. General Information
| Submitter: | Reliant Technologies, Inc.<br>260 Sheridan Avenue<br>Suite 208<br>Palo Alto, CA 94306 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------|
| Contact Person: | Anne C Worden<br>Principal Regulatory Consultant<br>AL Voss Associates<br>P.O. Box 405<br>Angels Camp, CA 95221 |
| Summary Preparation Date: | September 11, 2002 |
#### II. Names
| Device Names: | Reliant BTL 5000 Puls Electrical Stimulation System<br>Reliant BTL 5000 Sono Ultrasound System<br>Reliant BTL 5000 Combi Combination Electrical<br>Stimulation & Ultrasound System |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Classification Name: | Ultrasound and Muscle Stimulator (and Accessories) |
#### III. Predicate Devices
- · Chattanooga Group, Inc. Vectra 2C and Vectra 4C (K982317);
- · Chattanooga Group, Inc. Vectra Pro 2 and Vectra Pro 4 (K982324);
- · Dynatronics Corporation Dynatron 150 Plus Ultrasound (K935728);
- · Excel Tech XL Tek Ultra VM Ultrasound (K001166); and
- · Empi 300 PV Complete Electrotherapy System (K021100).
#### IV. Product Description
Reliant BTL 5000 Puls Electrical Stimulation System, BTL 5000 Sono Ultrasound System, and BTL 5000 Combi Combination Electrical Stimulation & Ultrasound systems are comprised of the following main components:
- A system console (including software and control electronics); .
- A control and display panel; ●
- . Delivery device accessories (patient cables and electrodes and/or ultrasound cables and sound heads.
Premarket Notification, 510(k) Submission for:
{1}------------------------------------------------
#### Indications for Use V.
The Reliant BTL 5000 Puls Electrical Stimulation System, BTL 5000 Sono Ultrasound System, and BTL 5000 Combi Combination Electrical Stimulation & Ultrasound systems (and the delivery accessories that are used with them) are indicated for use in the stimulating, relaxing, and/or repeatedly contracting muscles by passing electrical currents through electrodes contacting the affected body area in the medical specialties of physical medicine, general and plastic surgery and neurology and/or for use in applying therapeutic deep heat for selected medical conditions such as relief of pain, muscle spasms, and joint contractures in the medical specialty of physical medicine. The Reliant BTL 5000 Puls Electrical Stimulation System, BTL 5000 Sono Ultrasound System, and BTL 5000 Combi Combination Electrical Stimulation & Ultrasound systems are indicated for the treatment and relaxation of muscle spasms, prevention or retardation of disuse atrophy, increasing local blood circulation, muscle re-education, maintaining or increasing range of motion, immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, symptomatic relief of chronic, intractable pain, management of pain associated with post-traumatic or post-operative conditions using electrical stimulation and are indicated for the application of therapeutic deep heat for the treatment of selected medical conditions such as relief of pain, muscle spasms and joint contractures, relief of pain, muscle spasms and joint contractures that may be associated with adhesive capsulitis, bursitis with slight calcification, myositis, soft tissue injuries, shortened tendons due to past injuries and scar tissues, relief of chronic and subchronic pain and joint contractures resulting from capsular tightness and capsular scarring using ultrasound. The combination ultrasound and electrotherapy modes deliver deep heat for pain management (ultrasound therapy) at the same time that electrical stimulation for either pain management or muscle stimulation is being delivered. Only one electrotherapy mode (TENS, High Voltage, Interferential or Premodulated) can be utilized at a time in conjunction with the ultrasound mode.
#### VI. Rationale for Substantial Equivalence
The Reliant BTL 5000 Puls Electrical Stimulation System, BTL 5000 Sono Ultrasound System, and BTL 5000 Combi Combination Electrical Stimulation & Ultrasound systems share the same general indications for use in physical medicine, general and plastic surgery and neurology applications and share the same or similar basic characteristics and features and, therefore, are substantially equivalent to the Chattanooga Group, Inc. Vectra 2C and Vectra 4C (K982317), the Chattanooga Group, Inc. Vectra Pro 2 and Vectra Pro 4 (K982324), the Dynatronics Corporation Dynatron 150 Plus Ultrasound (K935728) , the Excel Tech XL Tek Ultra VM Ultrasound (K001166), and the Empi 300 PV Complete Electrotherapy (K021100) systems.
#### VII. Safety and Effectiveness Information
Validation documentation and a comparison of the technical characteristics and features were provided to demonstrate that the Reliant BTL 5000 Puls Electrical Stimulation System, BTL 5000 Sono Ultrasound System, and BTL 5000 Combi Combination Electrical Stimulation & Ultrasound systems are safe and effective, when indicated in specific applications in the medical specialties of physical medicine, general and plastic surgery and neurology.
{2}------------------------------------------------
### VIII. Conclusion
The Reliant BTL 5000 Puls Electrical Stimulation System, BTL 5000 Sono Ultrasound System, and BTL 5000 Combination Electrical Stimulation & Ultrasound systems were found to be substantially equivalent to the predicate Chattanooga Group, Inc. Vectra 2C and Vectra 4C (K982317), the Chattanooga Group, Inc. Vectra Pro 2 and Vectra Pro 4 (K982324), the Dynatronics Corporation Dynatron 150 Plus Ultrasound (K935728), the Excel Tech XL Tek Ultra VM Ultrasound (K001166), and the Empi 300 PV Complete Electrotherapy (K021100) systems. The Reliant BTL 5000 Puls Electrical Stimulation System, BTL 5000 Sono Ultrasound System, and BTL 5000 Combi Combination Electrical Stimulation & Ultrasound systems share similar indications for use and characteristics and functional features, and thus are substantially equivalent to, the currently marketed predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 4 2003
Reliant Technologies, Inc. C/O: Ms. Anne Worden AL Voss Associates P.O. Box 405 Angels Camp, CA 95221
Re: K023050 Dated: January 3, 2003 Received: January 6, 2003
Trade/Device Name: BTL 5000 Combi Electrical Stimulation & Ultrasound System with rubber electrodes and sponge covers Regulation Numbers: 21 CFR 890.5860, 21 CFR 882.5890, 21 CFR 882.1320 Regulation Names: Ultrasound and muscle simulator, transcutaneous electrical nerve stimulator for pain relief, cutaneous electrodes Regulatory Class: Class II Product Codes: IMG, GZJ, GXY, LIH
Trade/Device Name: BTL 5000 Puls Electrical Stimulation System with rubber electrodes and sponge covers Regulation Numbers: 21 CFR 890.5850, 21 CFR 882.5890, 21 CFR 882.1320 Regulation Names: Powered muscle stimulator, transcutaneous electrical nerve stimulator for pain relief, cutaneous electrodes Regulatory Class: Class II Product Codes: IPF, GZJ, GXY, LIH
Trade/Device Name: BTL 5000 Sono Ultrasound System Regulation Number: 21 CFR 890.5300 Regulation Name: Ultrasound diathermy Regulatory Class: Class II Product Code: IMI
Dear Ms. Worden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
{4}------------------------------------------------
## Page 2 - Ms. Worden
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark N. Milkuson
Celia M. Witten, Ph.D., M.D Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{5}------------------------------------------------
# Attachment 2 Indications For Use Statement as Requested by FDA
510(k) Number (if Known): K023050
Device Name: Reliant BTL 5000 Puls Electrical Stimulation System, BTL 5000 Sono Ultrasound System, and BTL 5000 Combination Electrical Stimulation & Ultrasound System
Indications For Use:
The Reliant BTL 5000 Puls Electrical Stimulation System and the BTL 5000 Combi Combination Electrical Stimulation & Ultrasound System (and the delivery accessories that are used with them) are indicated for use in the medical specialties of physical medicine, general and plastic surgery and neurology for medical purposes/applications requiring:
Electrical Stimulation for Muscle & Soft Tissue Stimulation/Pain Management:
TENS, Russian, High Voltage Therapy Modes
- Treatment and relaxation of muscle spasms .
- Prevention or retardation of disuse atrophy .
- Increasing local blood circulation .
- Muscle re-education .
- Maintaining or increasing range of motion .
- Immediate postoperative/post-surgical stimulation of calf muscles to prevent venous . thrombosis
TENS, Microcurrent, Interferential and Premodulated Modes
- Symptomatic relief of chronic and/or intractable pain .
- Management of acute pain associated with post-traumatic or post-operative conditions .
## *** Continued on Following Page (page 1 of 2)***
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for
(Division Sign-Off)
Division of General. Restorative
and Neurological Devices
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
510(k) Number K023050
(Optional Format 1-2-96)
Attachment 2 - Page 1 Page Revised 02/05/2003
{6}------------------------------------------------
# Attachment 2 Indications For Use Statement as Requested by FDA
K023050 510(k) Number (if Known):
Device Name: Reliant BTL 5000 Puls Electrical Stimulation System, BTL 5000 Sono Ultrasound System, and BTL 5000 Combi Combination Electrical Stimulation & Ultrasound System
Indications For Use:
## ***Continued from Previous Page (page 2 of 2)***
The Reliant BTL 5000 Sono Ultrasound System and the BTL 5000 Combi Combination Electrical Stimulation & Ultrasound System (and the delivery accessories that are used with them) are indicated for use in the medical specialty of physical medicine for medical purposes/applications requiring:
Ultrasound for Deep Heat/Pain Management:
- · Application of therapeutic deep heat for the treatment of selected chronic and subchronic medical conditions such as:
- > Relief of pain, muscle spasms and joint contractures
- Relief of pain, muscle spasms and joint contractures that may be associated with: >
- Adhesive capsulitis .
- Bursitis with slight calcification .
- . Myositis
- Soft tissue injuries .
- Shortened tendons due to past injuries and scar tissues .
- Relief of chronic and subchronic pain and joint contractures resulting from: >
- Cansular tightness ●
- Capsular scarring .
The combination ultrasound and electrotherapy modes deliver deep heat for pain management (ultrasound therapy) at the same time that electrical stimulation for either pain management or muscle stimulation is being delivered. Only one electrotherapy mode (TENS, High Voltage. Interferential or Premodulated) can be utilized at a time in conjunction with the ultrasound mode.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use (Optional Format 1-2-96)
Premarket Notification, 510(k) Submission for: Reliant BTL 5000 Puls Electrical Stimulation, BTL 5000 Sono Ultrasound & BTL 5000 Combi Electrical Stimulation & Ultrasound Systems
Attachment 2 - Page 2 Page Revised 02/05/2003
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.