MID STREAM PREGNANCY TEST

K023037 · Advanced Diagnostics, Inc. · LCX · Apr 9, 2003 · Clinical Chemistry

Device Facts

Record IDK023037
Device NameMID STREAM PREGNANCY TEST
ApplicantAdvanced Diagnostics, Inc.
Product CodeLCX · Clinical Chemistry
Decision DateApr 9, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1155
Device ClassClass 2

Indications for Use

The Mid Stream Pregnancy test is a home pregnancy urine test designed for qualitative detection of the human chorionic gonadotropin (hCG) in urine. The test is for use as an aid in the diagnosis of early pregnancy.

Device Story

Single-step, visually read, qualitative immunochromatographic lateral flow stick test; detects hCG in urine. Device utilizes sandwich complex format (anti-hCG antibody/hCG/anti-hCG antibody). User (consumer) applies urine to test stick; presence of control line (C) indicates valid test; presence of test line (T) indicates positive result. Used in home setting for early pregnancy detection. Results provide immediate visual feedback to user to aid in pregnancy diagnosis.

Clinical Evidence

Bench testing and consumer studies performed. Correlation study showed 100% agreement with predicate device. Consumer study (n=115) showed 100% agreement with professional results; >99% of consumers found test simple, fast, and accurate. Intra- and inter-assay precision >99%.

Technological Characteristics

Lateral flow immunochromatographic assay; sandwich complex format (anti-hCG antibody/hCG/anti-hCG antibody). Single-use stick form factor. Visual readout. No electronic components or software.

Indications for Use

Indicated for non-professional/Over-The-Counter use for qualitative detection of human chorionic gonadotropin (hCG) in urine as an aid in the diagnosis of early pregnancy.

Regulatory Classification

Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K023037 # Advanced Diagnostics, Inc. APR 0 9 2003 ### 510 (k) Summary | Submitter | Name and address | Advanced Diagnostics, Inc.<br>(Division of Inamco Group of<br>Companies)<br>801 Montrose Avenue<br>S. Plainfield<br>NJ 07080<br>908 754 4880 (Tel)<br>908 754 5181 (Fax) | |--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Contact Person: | Dr. D. M. Chouhan | | | Date Prepared: | 11 September 2002 | | Device name: | Trade Name: Mid Stream Pregnancy Test | | | | Common Name: Human Chorionic Gonadotropin<br>(hCG) Test | | | | Classification Name: 21 CFR section 862.1155,<br>Class II. - A qualitative Human<br>Chorionic Gonadotropin (hCG)<br>test system. | | | Predicate Device | The Mid Stream Pregnancy Test is substantially<br>equivalent to other legally marketed devices for the<br>similar intended use. The device used for comparison<br>study is Acon Midstream Pregnancy Test made by<br>Acon Laboratories, Inc. having a 510K #.K983090. | | | Device Description | A single step, visually read, qualitative immuno-<br>chromatographic single use stick test. | | | Intended Use | The Mid Stream Pregnancy test is a home pregnancy<br>urine test designed for qualitative detection of the<br>human chorionic gonadotropin (hCG) in urine. The test<br>is for use as an aid in the diagnosis of early pregnancy. | | | | 4-1 | Continued.........2 | 801 Montrose Avenue, S. Plainfield, NJ 07080 USA Tel: 908 754 4880 Fax: 908 754 5181 e mail : cs@inamco.com {1}------------------------------------------------ # Advanced Diagnostics, Inc. (2) Summary of the · The intended use and performance characteristics : Both devices are OTC devices intended to use for an Similarities to the early detection of hCG in human urine. Predicate device Technological characteristics: . Both devices are one step, qualitative, visual lateral flow immuno-chromatographic test in a sandwich complex format of anti hCG antibody / hCG/ antihCG antibody. Interpretation of results : . The presence of C line serves as an internal quality control, and the presence of the T line indicates a positive result. Discussion and The comparison studies demonstrated that there was . Conclusion 100% correlation between the results obtained by Mid Stream Pregnancy Test and the predicate device. It indicates that the Mid Stream Pregnancy test is substantially equivalent to the predicate device. . The POL study and the evaluation carried out by different personnel with diverse educational background and work experience agreed 100% with the results expected, indicating that the testing results are highly repeatable. The precision of the test was found to be >99%. . The consumer study carried out with 115 consumers demonstrated that the results obtained from consumers agreed 100% with that obtained from - professionals. No discrepancy was found between the results obtained by Mid Stream Pregnancy Test and the predicate device. Over 99% of the consumers found the Mid Stream Test to be simple. fast, convenient, easy to understand and accurate. Continued............3 4-2 801 Montrose Avenue, S. Plainfield, NJ 07080 USA Tel: 908 754 4880 Fax: 908 754 5181 e mail : cs@inamco.com {2}------------------------------------------------ ## Advanced Diagnostics, Inc. #### (3) - . Based on the results of the consumer study, correlation study and the POL evaluations. we may conclude that Mid Stream Pregnancy Test is safe, as effective and performs as well as the legally marketed OTC device. It's intra and inter-assay precision is >99%. Therefore this device is suitable for use as an OTC Pregnancy Test. 4-3 801 Montrose Avenue, S. Plainfield, NJ 07080 USA Tel: 908 754 4880 Fax: 908 754 5181 e mail : cs@inamco.com {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a seal for the Department of Health & Human Services USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is a stylized image of three wavy lines that resemble a caduceus, a symbol often associated with medicine and healthcare. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 APR 0 9 2003 Dr. D. M. Chouhan President Advanced Diagnostics, Inc. 801 Montrose Avenue South Plainfield, NJ 07080 Re: k023037 Trade/Device Name: Mid Stream Pregnancy Test Regulation Number: 21 CFR 862.1155 Regulation Name: Human chorionic gonadotropin (HCG) test system Regulatory Class: Class II Product Code: LCX Dated: February 5, 2003 Received: February 6, 2003 Dear Dr. Chouhan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {4}------------------------------------------------ Page 2 -- This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Steven Sutman Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### Indications For Use 510 (k) NUMBER (if KNOWN) : _ KO 2 303 1 Mid Stream Pregnancy Test DEVICE NAME : INDICATIONS FOR USE : The Test is a qualitative immunoassay intended for non professional / Over The Counter Use for qualitative detection of the human chorionic gonadotropin (hCG) in urine. The test is for use as an aid in the diagnosis of early pregnancy. Sean Cooger (Divis Division Clinical Laboratory Devices 510(k) number K023037 ( PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDE. ) Concurrence of CDRH, Office of Device Evaluation ( ODE ) Prescription Use ( Per 21 CFR 801.109 ) or Over the Counter Use ( Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...