K023034 · Bionet Company, Ltd. · LOS · Nov 14, 2002 · CV
Device Facts
Record ID
K023034
Device Name
CARDIOTOUCH-3000
Applicant
Bionet Company, Ltd.
Product Code
LOS · CV
Decision Date
Nov 14, 2002
Decision
SESE
Submission Type
Special
Indications for Use
The CARDIOTOUCH-3000 is intended to be used under the direct supervision of a licensed healthcare practitioner. The CARDIOTOUCH-3000 is intended to be used by trained operators in a hospital or medical professional's facility environment to record ECG signals from surface electrodes. The device is intended to acquire, analyze, display, and record electrocardiographic information from adult population. The device is not intended for home use. The device is not designated for intracardial use.
Device Story
CARDIOTOUCH-3000 is a 12-channel electrocardiograph; acquires ECG signals via surface electrodes. Device features automated analysis of heart rate, PR/QRS/QT/QTc intervals, and P-R-T axis; provides 25 automated diagnostic interpretations. Operated by trained personnel in hospitals or clinics; requires single-button press for measurement, filtering, parameter extraction, and A4-size report printing. Includes built-in battery for portability in emergency or mobile hospital settings. Provides real-time 12-channel rhythm display. Output assists clinicians in assessing patient cardiac condition; improves chart management via integrated patient/operator information printing.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and comparison of technological characteristics, performance specifications, and intended use with the predicate device.
Technological Characteristics
12-channel ECG; ABS plastic housing; 320x200 dot backlit LCD. Input: floating, isolated, defibrillation-protected; impedance >= 10MΩ. Frequency response: 0.05-150Hz. Power: 100-240Vac 50/60Hz or built-in rechargeable battery. Connectivity: standard 12-lead patient cable. Software: automated parameter calculation and 25-diagnosis auto-interpretation.
Indications for Use
Indicated for adult patients requiring ECG signal acquisition, analysis, and recording. Used by trained operators in clinical or hospital settings under licensed practitioner supervision. Not for home or intracardial use.
{0}------------------------------------------------
## NOV 1 4 2002
# 510(k) Summary As Required by 21 section 807.92 ( c )
KO23034 p. L/3
1-Submitter Name: BIONET Co., LTD
- 2-Address: #501. KICOX Venture Center. 188-5. Guro-Dong, Guro-Gu
- Seoul, South Korea
- +82 2 6300 6419 / +82 2 6300 6418 3-Phone:
- 4-Fax: +82 2 6300 6425
5-Contact Person: Dong-Joo Kang, CEO
6-Date summary prepared: September 6th, 2002
7- Official Correspondent: Mansour Consulting LLC
- 1308 Morningside Park Dr 8- Address:
- Alpharetta, GA 30022 USA
- 9- Phone: (678) 908-8180 (425) 795-9341
- 10- Fax:
- 11- Contact person: Jay Mansour, president
- 12-Device Trade or Proprietary Name: CARDIOTOUCH-3000
13-Device Common or usual name: ECG
14-Device Classification Name: System, ECG Analysis
15-Substantial Equivalency is claimed against the following device: Bionet's EKG-2000 Cardio Care, 510k #k011328 (refer to Appendix 2 for FDA website printout) This notification for CARDIOTOUCH-3000 is of the
SPECIAL type
16-Description of the Device: (as per User Manual pages17 and 18) As an electrocardiograph with 12-channel, CARDIOTOUCH-3000 does not only provide parameters and diagnose a patient automatically for patient's electrocardiogram, but also it improves chart management by printing information on an operator as well as on the patient. Furthermore, for the convenience of operators, it is designed so that an operator needs to press a button once to let the machine measure and record patient's electrocardiogram, apply a filter, extract parameters, diagnose a patient automatically then print out a result on an A-4- size report. In addition, the built-in battery helps an operator carry and use CARDIOTOUCH-3000 easily and conveniently for a hospital on wheel and an emergency.
Product features:
- It prints out 12-channel ECG waveforms on an A-4 size report in . various channel forms including 3ch+1rhy, 6ch+1rhy, and 12ch rhy.
- It records 1ch rhy for 60 seconds and prints it on an A4-size report -
- It displays 12ch rhy simultaneously and continuously in real-time. -
- -It automatically calculates heart rate, PR interval, QRS interval, QT interval, QTc, and P-R-T axis for diagnosis and displays all of them with electrocardiogram on the report.
- Twenty-five diagnoses are available through auto diagnosis ・
- It helps a diagnosis with the changing once-recorded electrocardiogram in a filter setup, signal size, printing speed, channel form and rhythm form and then printing them.
- With a built-in battery, it is easy to carry. -
- Managing a chart is easy and improved because it prints information on a patient and operator at the same time.
3/1
{1}------------------------------------------------
Basic configuration and accessories:
- 1- CARDIOTOUCH-3000 body
- 2- Patient cable (1 EA)
- 3- User Manual (1 VOL.)
- 4- ECG Paper (2 EA)
- 5- Power cable (1 EA)
## Options:
- 1- Battery (built-in type)
- 2- Protective Ground wire (1 EA)
- 3- Electrode (10 EA)
- 4- ECG cream (1 EA)
- 5- Hanger (1 EA)
- 6- Cart (1 EA)
17-Intended use of the device: (Indications for use typed on a separate FDA form) The CARDIOTOUCH-3000 is intended to be used under the direct supervision of a licensed healthcare practitioner. The CARDIOTOUCH-3000 is intended to be used by trained operators in a hospital or medical professional's facility environment to record ECG signals from surface electrodes. The device is intended to acquire, analyze, display, and record electrocardiographic information from adult population. The device is not intended for home use. The device is not designated for intracardial use.
18-Safety and effectiveness of the device:
This device is safe and effective as the predicate device, being a deviation from the previously cleared Bionet's product EKG-2000 Cardio Care.
This is better expressed in the tabulated comparison (Paragraph 19 below) 19-Summary comparing technological characteristics with other predicate device:
Please find below a tabulated comparison supporting that CARDIOTOUCH-3000 is substantially equivalent to other medical devices in commercial distribution. Also, Equivalency detailed chart path is attached.
| FDA file reference number | 510k #k011328 |
|--------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Attachments inside<br>notification submission file | 510k summary print out |
| TECHNOLOGICAL<br>CHARACTERISTICS | Comparison result |
| Indications for use | Identical:<br>Both CardioCare-2000 and CardioTouch-3000<br>are detecting and measuring as below parameters<br>to provide and help doctors for making figure out<br>the patient's vital condition.<br>• Accurate Heart Rate<br>• Electrocardiograph wave<br>• Printing the current ECG wave and related<br>numeric values |
| Target population | Identical:<br>Predicate device and this device are used by<br>cardiologist and sports expert etc. for watching<br>heart patient condition. |
| | K023034<br>p.313 |
| Design | Similar:<br>CardioCare-2000 have 2line 16character LCD<br>and CardioTouch-3000 have 320 X 200 dot with<br>back lit LCD. |
| Materials | Identical:<br>CardioCare-2000 and CardioTouch-3000 are<br>made of ABS materials. |
| Performance | Identical:<br>ECG leads : Standard 12 leads<br>Sensitivity : 5,10,20, Auto<br>Input circuit : floting input, isolated and<br>defibrillation protected.<br>Input impedence : $\geq$ 10MΩ<br>Input circuit current : $\leq \pm$ 5mV<br>Calibration Voltage : 1mV±2%<br>Time Constant : 100dB or better<br>Frequency response : 0.05~150Hz within -3dB |
| Sterility | Not applicable |
| Biocompatibility | Identical:<br>Both of them are satisfied Biocompatibility<br>through using disposable electrode. |
| Mechanical safety | Identical:<br>A coarse surface and a sharp corners of a device<br>which has cause of patient damage of device are<br>removed and covered. |
| Chemical safety | Not applicable |
| Anatomical sites | Not applicable |
| Human factors | Not applicable |
| Energy used and/or delivered | Identical:<br>Both of them use the rechargeable battery<br>(100V~240Vac 50/60 Hz and Built-in). |
| Compatibility with<br>environment and other<br>devices | CardioCare-2000 and CardioTouch-3000 are<br>suitable for EMI and EMC test. |
| Where used | Identical: Private Check, ER, Hospitals. |
| Standards met | Identical |
| Electrical safety | Identical |
| Thermal safety | Identical |
| Radiation safety | Identical |
{2}------------------------------------------------
3/3
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three abstract human profiles. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Bionet Company Ltd. c/o Mr. Jay Mansour President Mansour Consulting LLC 1308 Morningside Park Drive Alpharetta, GA 30022
Re: K023034
Trade Name: CARDIOTOUCH-3000 Regulation Number: Unclassified Regulation Name: ECG Analysis System Product Code: LOS Dated: October 8, 2002 Received: October 21, 2002
Dear Mr. Mansour:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
### Page 2 - Mr. Jay Mansour
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Elias Mallis
Bram D. Zuckerman, M.D Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 510(k) Number (if known): KO23034
## Device Name: CARDIOTOUCH-3000
## Indications for Use:
The CARDIOTOUCH-3000 is intended to be used under the direct supervision of a licensed healthcare practitioner. The CARDIOTOUCH-3000 is intended to be used by trained operators in a hospital or medical professional's facility environment to record ECG signals from surface electrodes. The device is intended to acquire, analyze, display, and record electrocardio information from adult population. The device is not intended for home use. The device is not designated for intracardial use.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) ----------------------------------------------------
Concurence of CDRH, Office of Device Evaulation (ODE)
X
(Optional Format 3-10-98)
8)
Dias Malli
Division of Cardiovascular & Respiratory Devices
510(k) Number K023034
C3
Prescription Use
(Per 21 C.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.