SUILE WOUND DRESSING

K022967 · Hedonist Biochemical Technologies Co., Ltd. · FRO · Jul 8, 2003 · SU

Device Facts

Record IDK022967
Device NameSUILE WOUND DRESSING
ApplicantHedonist Biochemical Technologies Co., Ltd.
Product CodeFRO · SU
Decision DateJul 8, 2003
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

The SUILE Wound Dressing is indicated for the management of 1st and 2nd degree burns, partial thickness wounds, donor sites and abrasions.

Device Story

SUILE Wound Dressing is a topical medical device used for wound management. It is applied to 1st and 2nd degree burns, partial thickness wounds, donor sites, and abrasions. The device is intended for prescription use by healthcare professionals in clinical settings to facilitate wound healing and protection. It functions as a physical barrier to the wound site.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Wound dressing material; physical barrier function; non-electronic; no software or algorithm components.

Indications for Use

Indicated for management of 1st and 2nd degree burns, partial thickness wounds, donor sites, and abrasions in patients requiring wound care.

Submission Summary (Full Text)

{0}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol often associated with healthcare. The caduceus is depicted with a staff entwined by two snakes and topped with wings. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 0 8 2003 Hedonist Biochemical Technologies Co., LTD c/o Eric Luo, Ph.D. 6833 Saint Lawrence Street Plano, Texas 75024 Re: K022967 Trade/Device Name: SUILE Wound Dressing Regulatory Class: Unclasssified Product Code: FRO Dated: May 2, 2003 Received: May 5, 2003 Dear Dr. Luo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Page 2 - Eric Luo, Ph.D. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Miriam C. Provost Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Page 1 of 1 510(k) Number (if known): K022967 Device Name: SUILE Wound Dressing Indications For Use: The SUILE Wound Dressing is indicated for the management of 1st and 2nd degree burns, partial thickness wounds, donor sites and abrasions. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use X (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) Miriam C. Provost (Division Sign-Off) (Division of General, Restorative and Neurological Devices 510(k) Number K022967
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...