K022963 · Transonic Systems, Inc. · MQS · Dec 4, 2002 · Gastroenterology, Urology
Device Facts
Record ID
K022963
Device Name
TRANSONIC SYRINGE WARMER, MODEL SYR-1000
Applicant
Transonic Systems, Inc.
Product Code
MQS · Gastroenterology, Urology
Decision Date
Dec 4, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Indications for Use
The Transonic Syringe Warmer (Model SYR1000) is indicated for use with the Transonic HD01 (HD01PMS and HD02) systems to make Cardiac output measurements. The Transonic Syringe Warmer (Model SYR1000) is an accessory to our HD01-CO Hemodialysis Monitor for Cardiac Output (K980906) which are currently marketed as the HD01046 and HD02 systems. The HD01-CO Hemodialysis Monitor for Cardiac Output (HD01P05 and HD02 systems) requires the user to make a 30cc injection of approximately body temperature saline into the hemodialysis blood circuit for the CO measurement. The syringe warmer provides the user a convenient source of the 30cc injection of approximately body temperature saline for this injection.
Device Story
Device is a syringe warmer (Model SYR-1000) used as an accessory to Transonic HD01-CO Hemodialysis Monitor. Input: 30cc saline syringe. Function: warms saline to approximately body temperature. Output: warmed saline for injection into hemodialysis blood circuit. Used in clinical hemodialysis settings by healthcare providers. Benefit: provides convenient source of body-temperature saline required for cardiac output measurement via the HD01-CO monitor.
Clinical Evidence
Bench testing only. No clinical data provided. Finished products undergo physical testing and visual examination per Quality Control Test Procedure documents to ensure performance parameters conform to design specifications.
Technological Characteristics
Heating device for 30cc syringes. Materials bench-tested for suitability; no patient contact, thus no biocompatibility testing required. Operates as an accessory to the HD01-CO Hemodialysis Monitor.
Indications for Use
Indicated for use as an accessory to Transonic HD01/HD02 hemodialysis systems to warm 30cc saline syringes to body temperature for cardiac output measurements.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Predicate Devices
HD01-CO Hemodialysis Monitor for Cardiac Output (K980906)
Vista Dental Syringe Warmer (Class I exempt)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness in Accordance with SMDA'90
September 5, 2002
| Transonic Systems Inc. | |
|------------------------|---------------------------|
| 34 Dutch Mill Rd | Telephone: (607) 257-5300 |
| Ithaca, NY | Fax: (607) 257-7256 |
DEC 0 4 2002
| Contact: | Mark S. Alsberge |
|----------------------|----------------------------------------------------------------------------------------------------------------------------|
| Product Name: | Transonic Syringe Warmer |
| Classification name: | Hemodialysis Access Recirculation Monitoring System<br>Gastroenterology and Urology<br>Class II, 78MQS<br>21 CFR §876.5820 |
# SUBSTANTIAL EQUIVALENCE1 TO:
| 510 (k) Number | Name | Applicant |
|----------------------------|----------------------------------------------------|----------------------------------------|
| K980906 | HD01-CO Hemodialysis Monitor<br>for Cardiac Output | Transonic Systems Inc. |
| Class I exempt | Vista Dental Syringe Warmer | Inter-Med Inc. / Vista Dental Products |
| 872.6100 Anesthetic warmer | | |
# Device Description:
In accordance with section 510(k) for the Federal Food, Drug, and Cosmetic Act, Transonic Systems Inc. intends to introduce into interstate commerce the Transonic Syringe Warmer as an accessory to our HD01-CO Hemodialysis Monitor for Cardiac Output (K980906). The HD01-CO Hemodialysis Monitor for Cardiac Output requires the user to make a 30cc injection of ~ body temperature saline into the hemodialysis blood circuit for the CO measurement. The syringe warmer will provide the user a convenient source of the 30cc injection of ~ body temperature saline for this injection.
<sup>1</sup> The term "substantially equivalent" as use herein is intended to be a determination of substantial equivalence from an FDA regulatory point of view under the Federal Food, Drug, and Cosmetic Act and relates to the fact that the product can be marketed without premarket approval or reclassification. These products may be considered distinct from a patent point of view. The term "substantially equivalent" is not applicable to and diminish any patent claims related to this product or the technology used to manufacture the product.
{1}------------------------------------------------
# 022963 AGE 2 OF 2
#### Material:
The Transonic Syringe Warmer is made of materials which have been bench tested and demonstrated themselves to be suitable for the intended use of this product. The materials have no patient contact; therefore, no biocompatibility testing was completed.
### Substantial equivalence:
The Transonic Syringe Warmer is similar in materials and form to the Vista Dental Anesthetic Syringe Warmer currently marketed by Inter-Med Inc. / Vista Dental Products (Class I Exempt, 872.6100 Anestheic warmer). The Transonic Syringe Warmer introduces a slight modification to the intended use. The Transonic Syringe Warmer will be used as an accessory to our HD01-CO Hemodialysis Monitor for Cardiac Output (K980906). They will be used to warm a 30cc syringe of saline to ~ body temperature for injection into the hemodialysis blood circuit rather than to warm dental anesthetic for injection into the patient. This use difference does not raise any new issues of safety or effectiveness regarding the Transonic Syringe Warmer.
#### Safety and Effectiveness:
All finished products are tested and must meet all required release specifications before distribution. The array of testing required for release includes, but are not limited to; physical testing, visual examination (in process and finished product
The physical testing is defined by Quality Control Test Procedure documents. These tests are established testing procedures that ensure the products performance parameters conform to the product design specifications.
The testing instruction records for each of the individually required procedures are approved, released, distributed and revised in accordance with document control cGMP"s.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the words "Public Health Service" in a simple, sans-serif font. The text is black and appears to be on a white background. The words are arranged on a single line, with "Public" and "Health" being the most prominent words.
Food and Drug Administrat 200 Corporate Boulevard Rockville MD 20850
Mr. Mark S. Alsberge VP Medical and Regulatory Affairs Transonic Systems, Inc. 34 Dutch Mill Road ITHACA NY 14850-9787
- Re: K022963 Trade/Device Name: Transonic Syringe Warmer Model SYR-1000 Regulation Number: 21 CFR 8876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: 78 MOS Dated: September 5, 2002 Received: September 6, 2002
Dear Mr. Alsberge:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
Image /page/2/Picture/10 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing movement or progress.
{3}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): - Transonic Syringe Warmer_(Model SYR1000) Device Name: Indications for Use:
The Transonic Syringe Warmer (Model SYR1000) is indicated for use with the Transonic HD01 (HD01PMS and HD02) systems to make Cardiac output measurements. The Transonic Syringe Warmer (Model SYR1000) is an accessory to our HD01-CO Hemodialysis Monitor for Cardiac Output (K980906) which are currently marketed as the HD01046 and HD02 systems. The HD01-CO Hemodialysis Monitor for Cardiac Output (HD01P05 and HD02 systems) requires the user to make a 30cc injection of approximately body temperature saline into the hemodialysis blood circuit for the CO measurement. The syringe warmer provides the user a convenient source of the 30cc injection of approximately body temperature saline for this injection.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurence of CDRH, Office of Device Evaulation (ODE)
*Prescription Use* √
Emid A. Syeam
o of Reproductive, Abdominal, adiological Devices, 51()(k) Number _
(Optional Format 3-10-98)
(Posted July 1, 1998)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.