SAFESTEP SAFETY DENTAL CARTRIDGE INJECTOR, MODEL 33-00000-1
K022959 · MDC Research , Ltd. · EJI · Feb 5, 2003 · Dental
Device Facts
Record ID
K022959
Device Name
SAFESTEP SAFETY DENTAL CARTRIDGE INJECTOR, MODEL 33-00000-1
Applicant
MDC Research , Ltd.
Product Code
EJI · Dental
Decision Date
Feb 5, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6770
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SafeStep™ Safety Dental Cartridge Injector is a sterile, single use disposable injector barrel assembly and a reusable plunger assembly that is indicated for use with pre-filled, 1.8mL, anesthetic cartridges and existing screw-on dental needles for injection of anesthetic solution in oral tissues. The SafeStep™ Safety Dental Cartridge Injector aids in the prevention of needle stick injuries. The SafeStepTM Safety Dental Cartridge Injector incorporates a needle retraction and re-advancing mechanism which may be operating with one hand.
Device Story
SafeStep™ Safety Dental Cartridge Injector is a manual, single-use disposable injector barrel with a reusable plunger assembly; used by dental professionals for delivering anesthetic solution into oral tissues. Device accepts standard 1.8mL pre-filled anesthetic cartridges and screw-on dental needles. Key feature is a one-handed needle retraction and re-advancing mechanism designed to prevent needle stick injuries. Device is purely mechanical; no electronic or software components. Clinical benefit is reduced risk of sharps injuries to the clinician during dental procedures.
Clinical Evidence
Simulated use clinical study compared SafeStep™ Safety Dental Cartridge Injector to the Septodont Safety Plus predicate. Primary endpoint: prevalence of sharps injuries or safety feature failure. Results showed zero sharps injuries and zero failures of the protection mechanism. Evaluator preference and performance were equivalent to the predicate.
Technological Characteristics
Manual, mechanical dental cartridge syringe. Components: single-use disposable injector barrel, reusable plunger assembly. Materials tested for biocompatibility per ISO standards. Sterilization: gamma irradiation (25 kGy) per ANSI/AAMI/ISO 11137-1999. No electronic, software, or energy-based components.
Indications for Use
Indicated for use in patients requiring dental anesthetic injections in oral tissues. Compatible with 1.8mL pre-filled anesthetic cartridges and standard screw-on dental needles. Intended for use by dental professionals to aid in the prevention of needle stick injuries.
Regulatory Classification
Identification
A cartridge syringe is a device intended to inject anesthetic agents subcutaneously or intramuscularly. The device consists of a metal syringe body into which a disposable, previously filled, glass carpule (a cylindrical cartridge) containing anesthetic is placed. After attaching a needle to the syringe body and activating the carpule by partially inserting the plunger on the syringe, the device is used to administer an injection to the patient.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
{0}------------------------------------------------
FEB 0 5 2003
# SUMMARY OF SAFETY AND EFFECTIVENESS SECTION 16:
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92.
## SUBMITTER INFORMATION 16.1
Original Premarket 510(k) Notification
| a. | Company Name: | MDC Research Ltd. |
|----|--------------------------------------|--------------------------------------------|
| b. | Company Address: | 2810 Bunsen Avenue<br>Ventura, CA 93003 |
| c. | Company Phone:<br>Company Facsimile: | (805) 339-0375<br>(805) 339-9751 |
| d. | Contact Person: | David Dowsett<br>Executive V.P. and C.O.O. |
| e. | Date Summary Prepared: | December 11, 2002 |
#### DEVICE IDENTIFICATION 16.2.
| a. Trade/Proprietary Name: | SafeStepTM Safety Dental Cartridge Injector |
|----------------------------|---------------------------------------------|
| b. Classification Name: | Cartridge Syringe<br>21 CFR 872.6770 |
# IDENTIFICATION OF PREDICATE DEVICE 16.3
| Company | Device | 510(k) No. | Date Cleared |
|-----------------------|-------------------------------------|------------|--------------|
| Specialties Septodont | Safety Plus Dental<br>Needle System | K913573 | 11/07/91 |
## 16.4 DEVICE DESCRIPTION
The SafeStep™ Safety Dental Cartridge Syringe consists of a single-use, disposable aspirating syringe-type injector with a separate re-usable plunger assembly indicated for the use in delivering dental anesthetics. The SafeStep™ Safety Dental Cartridge Syringe incorporates a needle safety mechanism that can be activated with one hand.
{1}------------------------------------------------
#### SUBSTANTIAL EQUIVALENCE 16.5
The SafeStep™ Safety Dental Cartridge Injector is substantially equivalent to the Septodont Safety Plus Dental Needle Injection System in terms of function, materials, and intended use. In a simulated use clinical study, the SafeStep™ Safety Dental Cartridge Injector was found to perform equivalently to the predicate device.
#### 16.6 INTENDED USE
The SafeStep™ Safety Dental Cartridge Injector is a sterile, single use disposable injector barrel assembly and a reusable plunger assembly that is indicated for use with pre-filled, 1.8mL, anesthetic cartridges and existing screw-on dental needles for injection of anesthetic solution in oral tissues. The SafeStep™ Safety Dental Cartridge Injector aids in the prevention of needle stick injuries. The SafeStepTM Safety Dental Cartridge Injector incorporates a needle retraction and re-advancing mechanism which may be operating with one hand.
# TECHNOLOGICAL CHARACTERISTICS 16.7
The SafeStep™ Safety Dental Cartridge Injector is equivalent to the Septodont Safety Plus in terms of materials, function and intended use. Both devices are constructed of a single use, disposable injector barrel component and a reusable plunger assembly. Both devices utilize pre-filled 1.8 mL dental anesthetic cartridges and standard screw-on dental needles.
#### 16.8 NON-CLINICAL STUDIES
Performance testing of the SafeStep™ Safety Dental Cartridge Injector included thermal stress, functional attribute, injection force, compression/tension loading before and after retraction, release force, leakage/splatter and aspiration testing. The results of the performance evaluations of the SafeStep™ Safety Dental Cartridge Injector were found to be acceptable in all test cases.
{2}------------------------------------------------
The materials of the SafeSter™ Safety Dental Cartridge Injector were tested for biocompatibility in accordance with ISO standards. The results of the biocompatibility tests, including cytotoxicity, sensitization and acute intracutaneous reactivity, were found to be acceptable.
The SafeStep™ Safety Dental Cartridge Injector is sterilized by gamma irradiation using a dosage of 25 kGy. Sterilization validation is to be performed on the final form-fill-seal production package in accordance with ANSI/AAMI/ISO 1137-1999 using Method 1 validation. MDC Research certifies that the SafeStep™ Safety Dental Cartridge Injector will pass ANSV/AAMI/ISO 11137-1994, Method 1 validation prior to marketing of the device.
#### 16.9 CLINICAL STUDIES
MDC Research Ltd. conducted a simulated use clinical study to evaluate the safety and effectiveness of the MDC SafeStep™ Safety Dental Cartridge Injector (SIDCI), for use in the injection of dental anesthetic medication into a patient. The primary research hypothesis being tested is that the use of the MDC SDCI for injection of dental anesthetic medication causes zero prevalence of sharp injuries, or failures of the safety protection feature of the device that could reasonably lead to a sharps injury. The simulated use clinical study evaluated the MDC SafeStep™ Safety Dental Cartridge Injector in comparison to the Septodont Safety Plus predicate device.
The results of the simulated use study demonstrated that the MDC SDCI is equivalent to the Septodont Safety Plus predicate device in terms of performance and evaluator preference. Further, there were no adverse events reported (i.e., sharps injuries) and no failures of the protection feature of the device that could reasonably lead to a sharps injury.
{3}------------------------------------------------
# 16.10 CONCLUSIONS
This notification contains all information required by 21 CFR 807.87. A completed copy of the Premarket Notification 510(k) Reviewer's Checklist is provided in this submission. Performance testing, non-clinical and clinical evaluations of the SafeStep™ Safety Dental Cartridge Injector show that the device performs as intended. Comparison the SafeStep™ Safety Dental Cartridge Injector to the predicate devices show that the device is substantially equivalent.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized representation of a human figure in profile, with three curved lines above it, resembling a bird in flight. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 0 5 2003
MDC Research Limited C/O Ms. Carol Patterson Patterson Consulting Group, Incorporated 21911 Erie Lane Lake Forest, California 92630
Re: K022959
Trade/Device Name: SafeStep™ Safety Dental Cartridge Injector Regulation Number: 872.6770 Regulation Name: Cartridge Syringe Regulatory Class: II Product Code: EJI Dated: December 11, 2002 Received: December 13, 2002
Dear Ms. Patterson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Patterson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Susan Runne
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
K022959
# INDICATION FOR USE
| 510(k) Number: | K022959 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | SafeStep™ Safety Dental Cartridge Injector |
| Indications for Use: | The SafeStep™ Safety Dental Cartridge Injector is a sterile, single use disposable injector barrel assembly and a reusable plunger assembly that is indicated for use with pre-filled, 1.8mL, anesthetic cartridges and existing screw-on dental needles for injection of anesthetic solution in oral tissues. The SafeStep™ Safety Dental Cartridge Injector aids in the prevention of needle stick injuries. The SafeStep™ Safety Dental Cartridge Injector incorporates a needle retraction and re-advancing mechanism which may be operated with one hand. |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | X |
|------------------|---|
|------------------|---|
(Per 21 CFR 801.109)
OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
Ken Mnhuy for M.Sir
(Division Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
510(k) Number: K022959
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.