MODIFICATION TO DISETRONIC D-TRONPLUS INSULIN INFUSION PUMP
K022831 · Disetronic Medical Systems AG · LZG · Sep 11, 2002 · General Hospital
Device Facts
Record ID
K022831
Device Name
MODIFICATION TO DISETRONIC D-TRONPLUS INSULIN INFUSION PUMP
Applicant
Disetronic Medical Systems AG
Product Code
LZG · General Hospital
Decision Date
Sep 11, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Disetronic D-TRONplus Insulin Infusion Pump is intended for the controlled delivery of insulin as prescribed by a physician. It is indicated for patients with insulin dependent Diabetes Mellitus who do not have optimum blood glucose control on conventional insulin injection therapy. Patients for insulin pump therapy must be highly motivated to perform self glucose monitoring on a frequent and regular basis. They must be able to adjust their insulin supply to varying needs depending on actual blood glucose levels, planned meals, physical activities, etc. Patients must be capable of operating the pump. They must also have access to the educational training, support, and follow-up of health care professional experienced in insulin pump therapy.
Device Story
Ambulatory, battery-operated insulin infusion pump; delivers insulin via reservoir piston mechanism driven by stepper motor. User inputs basal rates, boluses, and temporary rate adjustments via four-button interface. Device features infrared interface for data transmission to PC. Used by patients in home/daily settings; programmed by patient or healthcare provider. Output includes insulin delivery and therapy data; data memory readable by third-party diabetes management software. Benefits include improved glycemic control through precise, user-adjustable insulin delivery.
Clinical Evidence
Bench testing only. Device tested for safety and performance in accordance with IEC 60601-2-24 (infusion pump safety), IEC 60601-1 (general safety), and IEC 60601-1-2 (electromagnetic compatibility).
Technological Characteristics
Ambulatory infusion pump; impact-resistant plastic housing; IP X7 rated. Stepper motor-driven piston mechanism. Infrared interface for PC connectivity. Powered by battery. Compatible with standard female luer subcutaneous administration sets. Software-controlled basal/bolus delivery.
Indications for Use
Indicated for patients with insulin-dependent diabetes mellitus requiring better glycemic control than conventional injection therapy. Patients must be motivated, capable of frequent self-glucose monitoring, able to adjust insulin based on glucose levels/meals/activity, capable of operating the pump, and have access to professional training and support.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
SEP 1 1 2002
## 510(k) Summary
D-TRONplus Insulin Infusion Pump
Image /page/0/Picture/3 description: The image contains the alphanumeric string "K022831". The characters are written in a simple, sans-serif font, with a slightly rough or hand-drawn appearance. The text is oriented horizontally and appears to be the primary, if not only, element in the image.
li di
| K | |
|-------------------------|--------------------------------------------------------------------------------------------------|
| Decision Date | |
| Submitter: | Disetronic Medical Systems AG<br>Kirchbergstrasse 190, Postfach<br>CH-3401 Burgdorf, Switzerland |
| Contact: | Sandra Soniec, Phone +41 34 424 4111 |
| Trade/Proprietary Name: | Disetronic D-TRONplus Insulin Infusion Pump |
| Common/Usual Name: | Insulin Infusion Pump and Accessories |
| Classification Name: | Infusion Pump |
## Substantial Equivalence
The D-TRONplus Insulin Infusion Pump is substantially equivalent to the currently marketed D-TRON Insulin Infusion Pump K994186 / K021725.
## Device Description
The Disetronic D-TRONplus Insulin Pump is an ambulatory, battery operated pump that administers small quantities of insulin to the patient. It is compact, shock resistant and fulfills the requirements for IP X7 to maximize patient convenience. The pump housing is made of impact resistant plastic. The total system consists of the pump, adapter, cartridge and the optional PC software that allows convenient programming of the pump or monitoring of the therapy. Furthermore the pump's data memory can be read out and displayed by using third party diabetes management software. The D-TRONplus Insulin Infusion Pump is compatible with commercially available subcutaneous administration sets with standard female luer connectors.
Insulin delivery is accomplished through the reservoir piston mechanism. The piston is advanced by means of a stepper motor driving a rubber stopper forward into the cartridge. The frequency of the motor revolution is controlled by the microprocessor according to the information programmed by the user or their care provider. The rate of the hourly basal infusion for twenty-four hours, subsequent boluses and any temporary increase or reduction in the basal rate can be simply entered using the four buttons on the pump. Additionally, the user controls infusion start, infusion stop and has access to information important to proper monitoring of the pump and the therapy directly on the pump.
The D-TRONblus pump is equipped with an infrared interface that allows data transmission between the pump and a personal computer. Disetronic provides the DiaLog PC software as an accessory to the D-TRONplus pump that facilitates the programming and monitoring of the device.
## Indications for Use
The Disetronic D-TRONplus Insulin Infusion Pump is intended for the controlled delivery of insulin as prescribed by a physician.
It is indicated for patients with insulin dependent diabetes Mellitus who do not have optimum blood glucose control on conventional insulin injection therapy. Patients for insulin pump
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a handwritten string of alphanumeric characters. The string appears to be "K022831". The characters are written in a simple, slightly irregular style, giving the impression of casual handwriting. The image is in black and white, with the characters standing out against the white background.
510(k) Summary D-TRONplus Insulin Infusion Pump
Image /page/1/Picture/2 description: The image shows the logo for DISETROINC. The logo consists of a stylized letter "D" with a plus sign in the upper right corner, followed by the word "DISETRONIC" in bold, sans-serif font. A horizontal line is present below the word.
therapy must be highly motivated to perform self glucose monitoring on a frequent and reqular basis. They must be adjust their insulin supply to varying needs depending on actual blood glucose levels, planned meals, physical activities, etc. Patients must be capable of operating the pump. They must also have access to the educational training. support, and follow-up of health care professional experienced in insulin pump therapy.
### Technological Characteristics
The modifications made to the currently marketed D-TRON Insulin Infusion Pump have not affected the technological characteristics. Therefore, the D-TRONplus Insulin Infusion Pump has the same technological characteristics.
#### Performance Data
The Disetronic D-TRONplus Insulin Infusion Pump has been designed and tested in accordance with IEC 60601-2-24 of the International Electrotechnical Commission: Particular requirements for safety of infusion pumps and controllers. IEC 60601-2-24 incorporates the requirements of IEC 60601-1 for all general safety requirements and IEC 60601-1-2 for Electromagnetic compatibility - requirements and tests. The electronic and mechanical design is not unique and therefore the specifications fully address pump performance.
#### Conclusion
Based on the design equivalency and performance and safety testing. Disetronic Medical Systems has determined that the D-TRONplus Insulin Infusion Pump is substantially equivalent to the devices currently marketed in the United States.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# SEP 1 1 2002
Ms. Sandra Soniec Manager, Regulatory Affairs Disetronic Medical Systems AG Kirchbergstrasse 190, Postfach, CH-3401 Burgdorf, SWITZERLAND
Re: K022831
Trade/Device Name: Disetronic D-TRONplus Insulin Infusion Pump Regulation Number: 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LZG Dated: August 21, 2002 Received: August 26, 2002
Dear Ms. Soniec:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
Page 2 - Ms. Soniec
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618 . Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincerely yours,
Timothy A. Ulatowski
TimothyA. Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### INDICATIONS FOR USE STATEMENT
.
510(k) File Number:
Disetronic D-TRONplus Insulin Infusion Pump Device Name:
Indications For Use:
The Disetronic D-TRONplus Insulin Infusion Pump is intended for the controlled delivery of insulin as prescribed by a physician.
It is indicated for patients with insulin dependent Diabetes Mellitus who do not have optimum blood glucose control on conventional insulin injection therapy. Patients for insulin pump therapy must be highly motivated to perform self glucose monitoring on a frequent and regular basis. They must be able to adjust their insulin supply to varying needs depending on actual blood glucose levels, planned meals, physical activities, etc. Patients must be capable of operating the pump. They must also have access to the educational training, support, and follow-up of health care professional experienced in insulin pump therapy.
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-Th
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
Division of Anesthesiology, General Hos Infection Control Dental D
510(k) Number: K022831
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.