K022813 · Lake Region Mfg., Inc. · DQX · Feb 14, 2003 · Cardiovascular
Device Facts
Record ID
K022813
Device Name
STEERABLE GUIDEWIRE
Applicant
Lake Region Mfg., Inc.
Product Code
DQX · Cardiovascular
Decision Date
Feb 14, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
Lake Region's steerable guidewires are intended for use in angiographic procedures to introduce and position catheters and interventional devices within the coronary and peripheral vasculature, including the renal vasculature. The steerability feature allows the guidewire to be torqued to facilitate navigation through the vasculature.
Device Story
Steerable guidewire used in angiographic procedures to navigate catheters and interventional devices through coronary, peripheral, and renal vasculature. Device consists of PTFE-coated stainless steel steerable core; platinum alloy coil at distal tip for radiopacity; solder-secured coil. Features include torqueability for navigation; shapeable straight or preshaped tips; optional MDX (silicone) coating. Available in lengths 130cm-300cm; diameters 0.018". Used by clinicians in clinical settings. Provides mechanical support and guidance for interventional tools; facilitates precise positioning within vascular anatomy.
Clinical Evidence
Bench testing only. No clinical data. Verification testing performed per FDA Coronary and Cerebrovascular Guidewire Guidance and ISO 11070. Internal test protocols verified performance in peripheral and renal vasculature. Visual, tactile, dimensional, and mechanical attributes inspected against established criteria.
Technological Characteristics
PTFE-coated stainless steel core; platinum alloy radiopaque coil; solder joints. Optional MDX (silicone) coating. Lengths 130-300cm; 0.018" diameter. Mechanical steerability via torque transmission. Sterile, single-use. No software or electronic components.
Indications for Use
Indicated for use in angiographic procedures to introduce and position catheters and interventional devices within the coronary, peripheral, and renal vasculature.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
### SECTION 2.0 - SUMMARY OF SAFETY AND EFFECTIVENESS
### August 21, 2002
- 2.1 General Information
FEB 1 4 2003
- Company Name, Address, and Telephone Number 2.1.1
Lake Region Manufacturing, Inc. (LRM) 340 Lake Hazeltine Drive Chaska, MN 55318 Telephone: (952) 448-5111 Fax: (952) 448-3441
Contact Name: Karen Mortensen Regulatory Compliance Specialist
- 2.1.2 Device Trade Name/Proprietary Name
LRM produces guidewires on an OEM basis for other manufacturers, kit assemblers, and distributors. Consequently there are a large number of trade and proprietary names not including or associated with LRM. LRM has no proprietary names of its own to be included with this submission.
- 2.1.3 Device Common Names/Usual Names and Classification Names
These devices are commonly known as coronary and peripheral catheter guidewires. The current classification names and product codes are Wire, Guide, Catheter (74DQX).
- 2.1.4 Establishment Registration Number: 2126666
- 2.1.5 Classification of Devices
Catheter guidewires have been classified as Class II devices by the Circulatory Systems Devices Panel (reference 21 CFR 870.1330).
- 2.1.6 Applicability of Performance Standards
LRM has determined that no mandatory performance standards have been established for these devices under Section 514 of the Medical Amendments to Federal Food. Drug, and Cosmetic Act or by any subsequent regulatory action. LRM has also determined that there are no applicable voluntary standards.
- 2.2 Labels, Labeling, and Advertising
LRM produces cardiovascular and vascular guidewires on an OEM basis for other manufacturers, kit assemblers, and distributors. There is no direct distribution by
{1}------------------------------------------------
LRM. Changes to the customer controlled labels, labeling, or promotional material are at their discretion, including the resolution of any resulting regulatory obligations. A fraction of the total production bears LRM controlled labels and labeling.
- 2.3 Statement of Availability
This summary is being included in the Premarket Notification submission in lieu of a statement of availability.
- 2.4 Device Description
- Utilizing a proprietary process, LRM produces a PTFE coated stainless 2.4.1 steel steerable core. The proximal portion of the core wire is coated with PTFE to provide lubricity and improve wire handling. A platinum alloy coil provides radiopacity in the distal tip. The coil is secured in its location by solder, which is attached to the core. The proximal end of the coil is attached to the core with solder. Additionally, two sections of PTFE are removed from the proximal end of the core to aid in estimating guidewire positioning. The product is offered with a shapeable straight tip or in a preshaped configuration. The guidewires are optionally coated with MDX (silicone). The guidewires are bound by the following parameters:
| Outside Diameter: | .018" |
|-------------------|---------------------------------------------------|
| Lengths: | 130cm - 300cm |
| Tips: | Straight or shaped with various tip flexibilities |
| Flexibility: | Floppy to Support |
#### 2.4.2 Engineering Specifications
The design specifications are the same for the proposed device as they are for the LRM predicate device [reference 510(k) K970376 and K011968]. The finished devices must meet the same basic design criteria.
- 2.5 Substantial Equivalence Data
- 2.5.1 Background Information
The table below lists the differences between the predicate device and the proposed device. Testing was done to ensure the changes to the device met the predetermined performance criteria.
{2}------------------------------------------------
| Item | Proposed Device Differences from LRM Predicate cleared<br>under 510(k) K970376 and K011968 |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Raw Materials | Core:<br>No change<br>Coil:<br>No change<br>Extension system:<br>No change |
| Assembly Process | No significant change to assembly processes |
| Physical<br>Characteristics | No change except to expand the scope of the overall length<br>parameters from:<br>175cm – 300cm<br>to:<br>130cm – 300cm<br><br>and to allow radiopaque coil length ranges from 2 cm to<br>30 cm (was either 2 cm or 30 cm). |
| Labeling/IFU | The only change to the IFU will be to reference the<br>specific use regarding renal vasculature in addition to the<br>already stated coronary and peripheral vasculature use. |
| Intended Use | No change to intended use except to clarify the guidewire<br>can be used in the renal peripheral vasculature. |
| Anatomical Sites | No change |
| Target Population | No change |
| Performance Testing | No change |
| Safety Characteristics | No change |
| Biocompatibility | No change |
| Risk Analysis | No change |
- 2.5.2 In order to demonstrate equivalence of the proposed device, LRM performed testing to established requirements listed in FDA guidance document entitled Coronary and Cerebrovascular Guidewire Guidance, issued January 1995, and ISO 11070 Sterile Single-use Intravascular Catheter Introducers.
- 2.5.3 Additionally, internal test protocol requirements were utilized to verify the guidewire can be used in the peripheral vasculature, including the renal vasculature.
- 2.5.4 Configurations, including straight and shaped distal tips were inspected to established criteria. These parts were produced following current manufacturing processes and procedures. Test pieces were tested and inspected according to established specific inspection criteria requirements for visual/tactile, dimensional and mechanical attributes.
{3}------------------------------------------------
## 2.6 Qualification and Biocompatibility Test Data
### 2.6.1 Design Control
LRM is in conformance with the design control procedure requirements as specified in 21 CFR 820.30. Risk analysis was completed by means of a Failure Mode and Effect Analysis (FMEA) and all verification and validation activities resulted in the ability to demonstrate that the predetermined acceptance criteria could be met.
# 2.6.2 Material/Product/Process Qualification
LRM has formal quality systems in place to assure that the proposed steerable product will remain equivalent to the predicate product, and that the changes will not have an adverse affect on the safe and effective use of the product. The quality systems include Engineering Change Order Review, Material Qualification, Product Qualification, and Process Qualification. These controls are applied to each product size/group.
### 2.6.3 Biocompatibility Testing
The materials for the proposed device are identical to the predicate device cleared under K011968 and are identically processed and sterilized. Therefore, biocompatibility testing for the proposed device is not required.
### 2.7 Packaging and Sterilization Information
LRM produces guidewires on an OEM basis for other manufacturers, kit assemblers, and distributors. There is no direct distribution by LRM. A portion of the production is private label, sterile packaged to customer specifications, a fraction of that product is provided sterile to the customer.
The single packaged steerable guidewire is placed in a dispenser and then into a Tyvek®/poly pouch. The packaged product may be packaged as five or ten pouchs in a shelf carton, which are typical packaging configurations.
There will be no changes to the sterilization process for the portion of packaged product shipped sterile to the customer. For the product that is shipped bulk, the packaging design and sterilization process parameters are the customer's responsibility. LRM will not recommend that its customers modify their packaging or sterilization procedures as a result of this submission.
### 2.8 Intended Use Statement
Lake Region's steerable guidewires are intended for use in angiographic procedures to introduce and position catheters and interventional devices within the coronary and peripheral vasculature, including the renal vasculature. The steerability feature allows the guidewire to be torqued to facilitate navigation through the vasculature.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of a human figure in profile, with three faces overlapping to represent the department's focus on health and human well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the figure.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Lake Region Manufacturing, Inc. c/o Ms. Karen Mortensen Regulatory Compliance 340 Lake Hazeltine Drive Chaska, MN 55318-1029
Re: K022813
Trade Name: Coronary Peripheral and Renal Steerable Guidewires Regulation Number: 21 CFR 870.1330 Regulation Name: Wire, Guide, Catheter Regulatory Class: Class II (two) Product Code: DQX Dated: December 16, 2002 Received: December 17, 2002
Dear Ms. Mortensen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
FEB 1 4 2003
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 – Ms. Karen Mortensen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
VDeila Tuller
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known): Unknown
Device Name: Steerable Guidewires
Indications for Use:
Lake Region's steerable guidewires are intended for use in angiographic procedures to introduce and position catheters and interventional devices within the coronary and peripheral vasculature, including the renal vasculature. The steerability feature allows the guidewire to be torqued to facilitate navigation through the vasculature.
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use *X* Or Over-The-Counter Use __
(PER 21 CFR 801.109)
PREMARKET NOTIFICATION
(Division Sign-Off)
9-Division of Cardiovascular Devices
510(k) Number K022813
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.