Comparative evaluation of 113 serum samples (42 negative, 71 positive) comparing the VisiQuant ANA Test System to the VisiQuant ANA Test Kit.
—
Indications for Use
The VisiQuant™ Anti-Nuclear Antibody Test System is intended for the visual determination of anti-nuclear antibodies (ANA's) immunofluorescence pattern(s) and the semiquantitative measurement of ANA's in human serum with a single dilution as an aid in the in-vitro diagnosis of auto-immune and connective tissue diseases such as systemic lupus erythematosus (SLE) and Sjogren Syndrome.
Device Story
VisiQuant™ ANA Test System is an in-vitro diagnostic device for ANA detection in human serum. It consists of the previously cleared VisiQuant™ ANA Test Kit and a VisiQuant™ Microscope. The system uses a CCD camera to capture images of prepared slides. An operator views these images to identify fluorescence patterns and determine positivity. For positive samples, software calculates fluorescence intensity units (FIU) and interpolates ANA titers using a 4-parameter calibration curve generated from calibrators assayed in the same run. The system generates laboratory and calibration reports. It is intended for clinical laboratory use to aid in diagnosing auto-immune and connective tissue diseases. The device benefits patients by providing standardized, semi-quantitative ANA titer results to support clinical decision-making.
Clinical Evidence
Bench testing only. Comparative evaluation of 113 serum samples between the subject system and the predicate device. Qualitative agreement: 95.2% for negative results (40/42) and 100% for positive results (71/71). Pattern agreement for positive samples was 97.2% (69/71). Quantitative correlation of ANA titers (n=71) showed a correlation coefficient of 0.917 with a regression equation of Log VTM = 0.0578 + 0.966 Log VT.
Technological Characteristics
System includes a fluorescent microscope, CCD camera, and software. Sensing principle: immunofluorescence microscopy. Software performs image capture, fluorescence intensity unit (FIU) calculation, and 4-parameter curve interpolation for titer determination. Connectivity: standalone system. Sterilization: not applicable (in-vitro diagnostic).
Indications for Use
Indicated for patients suspected of having auto-immune or connective tissue diseases, including systemic lupus erythematosus (SLE) and Sjogren's Syndrome. Used for the visual determination of ANA immunofluorescence patterns and semi-quantitative measurement of ANA in human serum.
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
{0}------------------------------------------------
K022800
# 510(k) Summary of Safety and Effectiveness Information VisiQuant™ Anti-Nuclear Antibody Test System Revised September 20, 2002
NOV 2 5 2002 Hyperion, Inc. 14100 S.W. 136th Street Miami, FL. 33186 Contact Person: Radha Goolabsingh at 305-238-3020 x202 or Victor Rana at 305-238-3020 x208, or by facsimile at 305-232-7375.
| Trade or Proprietary Name: | VisiQuant™ Anti-Nuclear Antibody Test System |
|----------------------------|------------------------------------------------------------------------------|
| Common or Usual Name: | VisiQuant™ ANA Test System |
| Classification Name: | Anti nuclear antibody immunological test system |
| Registration Number: | Manufacturer |
| | Hyperion, Inc.<br>14100 S.W. 136th Street<br>Miami, Florida 33186<br>1028110 |
The proposed VisiQuant™ ANA System is substantially equivalent to the VisiQuant™ ANA Test Kit using a specially equipped fluorescent microscope, previously cleared under Document Control No., K013213, on 6/04/2002. The proposed and the predicate device are both in-vitro diagnostic products, intended for the visual determination of anti-nuclear antibodies (ANA's) immunofluorescence pattern(s) and the semi-quantitative measurement of ANA's in human serum sample with a single dilution as an aid in the diagnosis of auto-immune and connective tissue diseases such as systemic lupus erythematosus (SLE) and Sjogren's Syndrome. There are no changes to the test kit and the VisiQuant Microscope supplied meets the minimum requirements specified for the test kit in K013213.
The proposed VisiQuant™ ANA Test System uses the same VisiQuant™ ANA Test Kit as was previously cleared. Hyperion is now proposing to supply the VisiQuant™ Microscope as an additional reader for use with the VisiQuant™ ANA Test Kit.
The reading of the samples remains the same except that the software will now generate the calibration curve and interpolate the titer. The prepared slide is placed onto the stage of the microscope. The microscope focuses on the images of each well and the CCD camera collects the images. The software captures and stores the images. The operator views the images to determine if the sample is positive and identifies the pattern. For the samples determined to be positive by the operator, the software calculates a fluorescence intensity unit (FIU) value. The proposed software will now generate a 4-parameter curve using the FIUs and ANA titers of Calibrators assayed in the same run, from which the VisiQuant titer of the test sample is interpolated. The proposed software now has the capability of generating calibration curve reports and laboratory reports.
{1}------------------------------------------------
#### Method Comparison:
A comparative evaluation of the proposed VisiQuant™ ANA Test System demonstrated substantial equivalence to the VisiQuant™ ANA Test Kit using the specially equipped fluorescent microscope. The results are shown in the following tables and diagrams:
| VisiQuant™ ANA Test Kit | No. of samples with VisiQuant™ ANA Test System | | |
|-------------------------|------------------------------------------------|----------|-------|
| | Negative | Positive | Total |
| Negative | 40 | 2 | 42 |
| Positive | 0 | 71 | 71 |
| Total | 40 | 73 | 113 |
Agreement for negative results = 40/42 = 95.2% Agreement for positive results = 71/71 = 100%
Table3. Qualitative Results: The following table shows the pattern agreement for the 71 samples found to be positive by both the VisiQuant™ ANA Test Kit and the VisiQuant™ ANA Test System.
| VisiQuant™ ANA Test Kit | No. of samples with VisiQuant™ ANA Test System | | | | | | |
|-------------------------|------------------------------------------------|----|---|---|-----|-----|-------|
| | H | S | N | C | H,S | H,N | Total |
| H | 31 | | | | | | 31 |
| S | | 19 | | | 1 | | 20 |
| N | | | 1 | | | | 1 |
| C | | | | 2 | | | 2 |
| H,S | 1 | | | | 6 | | 7 |
| H,N | | | | | | 10 | 10 |
| Total | 32 | 19 | 1 | 2 | 7 | 10 | 71 |
Negative (Neg); Homogenous (H); Speckled (S); Nucleolar (N); Centromere (C)
Pattern agreement for samples positive by both tests = 69/71 = 97.2 %:
Table 4. Discrepant samples: Negative (Neg.), Not applicable (n.a); Speckled (S); Homogenous (H). Slight differences in reading could account for these discrepancies.
| Sample ID | VisiQuant™ ANA Test Kit | | VisiQuant™ ANA Test System | |
|-----------|-------------------------|---------|----------------------------|---------|
| | Titer | Pattern | Titer | Pattern |
| Y-SLE17 | 406 | S | 314 | H,S |
| X-E4 | 713 | H,S | 526 | H |
| Y-SS2 | Neg. | n.a. | 130 | H |
| W-14 | Neg. | n.a. | 97 | H,S |
{2}------------------------------------------------
Seventy-one (71) serum samples, positive by both the proposed VisiQuant™ ANA Test System and the VisiQuant™ ANA Test Kit using the specially equipped fluorescent microscope were compared for ANA titers. The following figure shows the correlation with all patterns (Homogenous, Speckled, Nucleolar, Centromere and mixed).
Image /page/2/Figure/3 description: This image is a scatter plot comparing ANA titers from two different VisiQuant ANA test systems. The x-axis represents the ANA titer of the VisiQuant ANA Test kit (VT), while the y-axis represents the ANA titer of the VisiQuant ANA Test System (VTM). The plot includes data points for various antibody patterns, such as Homogeneous, Speckled, Nucleolar, Centromere, Homog. & speckled, and Homog. & Nucleolar. A regression line and 95% confidence intervals are also displayed on the plot.
Fig. 1. Scatter diagram showing correlation in ANA titers for 71 positive results between VisiQuant™ ANA Test System and the VisiQuant™ ANA Test Kit using the specially equipped fluorescent microscope, with a correlation coefficient of 0.917 and a regression equation of Log VTM = 0.0578 + 0.966 Log VT. The Linear Regression and its 95% confidence intervals are also shown.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular pattern around the symbol. The caduceus is rendered in black, and the text is also in black.
Food and Drug Administrat 2098 Gaither Road Rockville MD 20850
Ms. Radha Goolabsingh Director of Quality Assurance and Regulatory Affairs Hyperion, Inc. 14100 S.W. 136th Street Miami, FL 33186
Re:
k022800 Trade/Device Name: VisiQuant™ Anti-Nuclear Antibody Test System Regulation Number: 21 CFR 866.5100 Regulation Name: Antinuclear antibody immunological test system Regulatory Class: Class II Product Code: DHN Dated: October 30, 2002 Received: October 31, 2002
Dear Ms. Goolabsingh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
### Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
# 510(K) NUMBER (IF KNOWN):
DEVICE NAME: VisiQuant™ Anti-Nuclear Antibody Test System
INDICATIONS FOR USE: The VisiQuant™ Anti-Nuclear Antibody Test System is intended for the visual determination of anti-nuclear antibodies (ANA's) immunofluorescence pattern(s) and the semiquantitative measurement of ANA's in human serum with a single dilution as an aid in the in-vitro diagnosis of auto-immune and connective tissue diseases such as systemic lupus erythematosus (SLE) and Sjogren Syndrome.
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence Of CDRH, Office Of Device Evaluation (ODE)
Prescription Use A (Per 21 CFR 801.109)
OR
Over-The-Counter-Use
(Optional Format 1-2-96)
J. Reeves for J. Bautista
(Division Sign-Off) Division of Clinical Laboratory Devi-510(k) Number .
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.