K022794 · Ge Imatron · JAK · Oct 11, 2002 · Radiology
Device Facts
Record ID
K022794
Device Name
E-SPEED EBT SCANNER SYSTEM
Applicant
Ge Imatron
Product Code
JAK · Radiology
Decision Date
Oct 11, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The intended use of the proposed GE Imatron e-Speed EBT Scanner System remains unchanged from the intended uses of prior predicate Imatron devices and other scanners. The e-Speed is designed -- as are all predicate devices -- to produce cross sectional images of the entire human anatomy. Specifically, the e-Speed is intended: - to permit radiologic visualization of the heart, blood vessels, or lymphatic 1. system, during or after injection of a contrast medium; - 2. to function as a diagnostic x-ray system to produce two and three dimensional images of the heart, blood vessels, or lymphatic system from a volume of computer reconstructed cross-sectional images from x-ray transmission data from the same axial plane taken at different angles; - to permit the data from certain three dimensional images to also be presented in 3. time-sequenced or cine fashion; - 4. to permit the data from three dimensional images to be viewed from a point of view moving inside the tissue, called "fly-through" mode for colons, blood vessels, or airways; - to be used for clinical situations requiring determination of specific quantitative న్. information, such as the determination of calcium or other materials in bone, tumors, or organs; - 6. to be used by the physician as an aid in the diagnosis of lung pathology; - 7. for real-time motion studies, vascular and tissue blood flow analysis, and full three-dimensional volume imaging. The e-Speed will make use of previously cleared workstation software products and software applications. Finally, such system is intended to be used in a manner consistent with those devices already classified and set forth in 21 CFR Sections 892.1600 and 892.1750.
Device Story
e-Speed EBT Scanner System utilizes electron beam technology for high-speed cross-sectional imaging. System components include gantry, electron source, patient table, operator console, and HV power supply. Operates by capturing x-ray transmission data at different angles to reconstruct 2D/3D images; supports cine/time-sequenced imaging and 'fly-through' visualization for airways/vessels. Used in clinical settings by physicians for diagnosis of lung pathology, vascular analysis, and quantitative tissue/bone assessment. Output provides cross-sectional images and quantitative data to aid clinical decision-making. Benefits include rapid imaging capabilities for real-time motion studies and detailed anatomical visualization.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, system verification, and validation against established safety and performance standards.
Technological Characteristics
Electron beam tomography (EBT) scanner; utilizes electron beam technology, modernized beam control, and redesigned x-ray data acquisition. Materials equivalent to C150XP (K001550). Compliant with UL2601-1, IEC 60601-1, and 21CFR Subchapter J. Supports 0.050 sec scan mode.
Indications for Use
Indicated for patients requiring cross-sectional imaging of the entire human anatomy, including heart, blood vessels, lymphatic system, lungs, bone, and organs. Used for diagnostic visualization, quantitative analysis (e.g., calcium scoring), and motion/blood flow studies. Prescription use only.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circle. The letters and the circle are white, and the background is black.
GE Imatron
General Flectric Compan 89 Ovster Point Boulevarr h San Francisco CA 941 367-6545, 650 583-9964 A GE Medical Systems Company
Ko22794
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h).
### Submitter:
Kerry A. Spear Regulatory and Compliance Manager, GE Imatron TEL: 650-742-8801 FAX: 650-827-7790 Summary prepared: 19 July 2002
# PRODUCT IDENTIFICATION
| Name: | e-Speed Electron Beam Tomography (EBT) Scanner System |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Name: | Angiographic X-ray System<br>Computed Tomography X-ray System |
| Manufacturer: | GE Imatron<br>389 Oyster Point Boulevard<br>South San Francisco, CA 94080 |
| Distributor: | Same as Manufacturer |
| Marketed Devices: | The e-Speed EBT Scanner System is of comparable type and<br>is substantially equivalent to currently marketed EBT and<br>Computed Tomography X-ray Systems that comply with the<br>same or equivalent standards and have the same intended<br>uses. |
### DEVICE DESCRIPTION
The e-Speed EBT Scanner System ("e-Speed") is composed of a Gantry/Front End Assembly, Rear Electrical/Mechanical Assembly with Electron Source, Patient Table, Operator Console, Computer, HV Power Supply and Power Conditioning and Distribution System, water chiller, and associated accessories.
Materials: Materials and construction are equivalent to the C150XP EBT Scanner System (K001550) and are compliant with UL2601-1, IEC 60601-1, and 21CFR Subchapter J.
{1}------------------------------------------------
Design: The system is designed to be an angiographic X-ray system, and a whole body CT scanner utilizing electron beam technology, a modernized beam control system and redesigned x-ray data acquisition system, a higher power HV supply, an operator console, and the capability to scan at 0.050 sec as the primary mode of operation.
## Indications for Use:
The intended use of the proposed GE Imatron e-Speed EBT Scanner System remains unchanged from the intended uses of prior predicate Imatron devices and other scanners. The e-Speed is designed -- as are all predicate devices -- to produce cross sectional images of the entire human anatomy.
Specifically, the e-Speed is intended:
- to permit radiologic visualization of the heart, blood vessels, or lymphatic 1. system, during or after injection of a contrast medium;
- 2. to function as a diagnostic x-ray system to produce two and three dimensional images of the heart, blood vessels, or lymphatic system from a volume of computer reconstructed cross-sectional images from x-ray transmission data from the same axial plane taken at different angles;
- to permit the data from certain three dimensional images to also be presented in 3. time-sequenced or cine fashion;
- 4. to permit the data from three dimensional images to be viewed from a point of view moving inside the tissue, called "fly-through" mode for colons, blood vessels, or airways;
- to be used for clinical situations requiring determination of specific quantitative న్. information, such as the determination of calcium or other materials in bone, tumors, or organs;
- 6. to be used by the physician as an aid in the diagnosis of lung pathology;
- 7. for real-time motion studies, vascular and tissue blood flow analysis, and full three-dimensional volume imaging.
The e-Speed will make use of previously cleared workstation software products and software applications.
Finally, such system is intended to be used in a manner consistent with those devices already classified and set forth in 21 CFR Sections 892.1600 and 892.1750.
{2}------------------------------------------------
# Kozzągy
## Comparison with Predicate:
It is the opinion of GE Imatron that the e-Speed EBT Scanner System is of a type and substantially equivalent to currently marketed angiographic and whole body X-ray computed tomography systems with respect to design, material composition, energy source, and radiation characteristics. It will comply with the X-ray safety requirements of 21CFR 1020.30, 1020.31, and 1020.33, as well as the safety requirements of UL2601-1, IEC 60601-1 and applicable collateral and particular standards.
# Adverse Effects on Health:
Potential electrical, mechanical, and radiation hazards are identified in a risk management summary (hazard analysis) and controlled by:
- . System verification and validation to ensure performance to specifications. Federal Regulations, and user requirements;
- Adherence to industry and international standards (UL/CSA and IEC).
# CONCLUSIONS
The e-Speed EBT Scanner System does not result in any new potential safety risks and performs as well as or better than devices currently on the market. GE Imatron considers the e-Speed EBT Scanner System to be equivalent to other marketed devices with the same indications for use and meeting similar standards.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services (HHS). The seal features a stylized eagle with three lines representing its body and wings. The eagle faces left and is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES" in a circular arrangement.
Food and Drug Administration 0 Corporate Boulevard Rockville MD 20850
OCT 1 1 2002
General Electric Imatron % Mr. Nishith Desai Primary Third Party Reviewer TÜV Rheinland of North America, Inc. 12 Commerce Road NEWTOWN CT 06470
Re: K022794
Trade/Device Name: e-speed Electron Beam Tomography (EBT) Scanner System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: 90 JAK
Dated: September 27, 2002 Received: September 27, 2002
Dear Mr. Desai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Snogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### STATEMENT OF INTENDED USE
510(K) Number (if known): ____K022794_________________________________________________________________________________________________________________________________________
Device Name: e-Speed EBT Scanner System
Indications for Use:
The intended use of the proposed GE Imatron e-Speed EBT Scanner System remains unchanged from the intended uses of prior predicate Imatron devices and other scanners. The e-Speed is designed - as are all predicate devices - to produce cross sectional images of the entire human anatomy.
Specifically, the e-Speed is intended:
- to permit radiologic visualization of the heart, blood vessels, or lymphatic 1. system, during or after injection of a contrast medium;
- 2. to function as a diagnostic x-ray system to produce two and three dimensional images of the heart, blood vessels, or lymphatic system from a volume of computer reconstructed cross-sectional images from x-ray transmission data from the same axial plane taken at different angles;
- 3. to permit the data from certain three dimensional images to also be presented in time-sequenced or cine fashion;
(Continued on following page)
Concurrence of CDRH, Office of Device Evaluation (ODE)
- OR -
Daniel A. Seymore
(Division Sign-Off) Division of Reproductive. Abdo and Radiological Devices 510(k) Number
Prescription Use (Per 21 CFR 801.109)
Over-The-Counter Use
{6}------------------------------------------------
- to permit the data from three dimensional images to be viewed from a point of 4. view moving inside the tissue, called "fly-through" mode for colons, blood vessels, or airways;
- to be used for clinical situations requiring determination of specific quantitative information, such as the determination of calcium or other materials in bone, tumors, or organs;
న్.
- to be used by the physician as an aid in the diagnosis of lung pathology; 6.
- for real-time motion studies, vascular and tissue blood flow analysis, and full 7. three-dimensional volume imaging.
The e-Speed will make use of previously cleared workstation software products and software applications.
Finally, such system is intended to be used in a manner consistent with those devices already classified and set forth in 21 CFR Sections 892.1600 and 892.1750.
Daniel A. Heyman
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.