MEDLEY SPO2 MODULE, MODEL 8210 & SPO2 SENSORS & CABLE
K022677 · Alaris Medical Systems, Inc. · DQA · Sep 11, 2002 · Cardiovascular
Device Facts
Record ID
K022677
Device Name
MEDLEY SPO2 MODULE, MODEL 8210 & SPO2 SENSORS & CABLE
Applicant
Alaris Medical Systems, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Sep 11, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The MEDLEY SpO2 Module and accessories are intended for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for adult, pediatric, and neonatal patients in hospitals and hospital-type facilities.
Device Story
The MEDLEY SpO2 Module, Model 8210, is a pulse oximetry system component that integrates with the MEDLEY Programming Module. It utilizes Nellcor pulse oximetry technology to process signals from patient-connected oximetry sensors and cables. The device non-invasively measures functional oxygen saturation of arterial hemoglobin and pulse rate. It is intended for use in hospital and hospital-type facilities by clinical staff. The system provides continuous monitoring data to healthcare providers, enabling assessment of patient oxygenation status and pulse rate, which supports clinical decision-making regarding patient respiratory and cardiovascular health. The module snaps into the Programming Module to form a unified monitoring system.
Clinical Evidence
Bench testing only. Design verification testing, including software verification and validation, was completed to demonstrate that the device performs within required specifications.
Technological Characteristics
The device is a pulse oximetry module utilizing Nellcor technology for signal processing. It consists of an SpO2 module, programming module, patient cables, and sensors. It is designed for continuous monitoring of SpO2 and pulse rate. The system is intended for hospital use and operates via non-invasive optical sensing.
Indications for Use
Indicated for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate in neonatal, pediatric, and adult patients. Suitable for use during motion and no-motion conditions, and for patients with good or poor perfusion in hospital settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
SEP 1 1 2002
Image /page/0/Picture/2 description: The image shows the text "K022677" in a large, handwritten-style font. Below this, in a smaller, more formal font, is the text "ORIGINAL PREMARKET 510(K) NOTIFICATION MEDLEY SPO2 MODULE, MODEL 8210". The image appears to be a document header or title section, possibly from a medical device submission.
## SUMMARY OF SAFETY AND EFFECTIVENESS ALARIS Medical Systems, Inc. MEDLEY SpO2 Module, Model 8210
SUBMITTER'S NAME: ALARIS Medical Systems, Inc. 10221 Wateridge Circle San Diego, CA 92121-2772 (858) 458-7563 (858) 458-6114 FAX
- Renée L. Fluet CONTACT PERSON: Principal Regulatory Affairs Specialist
- August 9, 2002 DATE PREPARED:
Proprietary Name DEVICE NAME: MEDLEY SpO2 Module, Model 8210 SpO2 Sensors and Cables
> Common Name Pulse Oximeter and Sensor Transducer & Electrode Cables
Classification Name Oximeter, DQA (870.2700) Cable, Transducer and Electrode, DSA (870.2900)
MEDLEY SpO2 Module, Model 8220, K010966 PREDICATE DEVICES: Nellcor OxiMax Pulse Oximetry System, K012891
#### DEVICE DESCRIPTION
The MEDLEY SpO2 Module, Model 8210, is essentially the same as the previously cleared device (MEDLEY SpO2 Module, Model 8220). The MEDLEY SpO2 Module (Models 8220 and 8210) and accessories is a device capable of non-invasively monitoring functional oxygen saturation of arterial hemoglobin and pulse rate. The system consists of an SpO2 Module, a Programming Module, connecting cable, and oximetry sensors.
{1}------------------------------------------------
# SUMMARY OF SAFETY AND EFFECTIVENESS Page 2 of 3
The MEDLEY Programming Module (PM) and SpO2 Module snap together. The Nellcor board contains all of the pulse oximetry algorithms used to measure saturation and pulse rate. The patient cable and oximetry sensors, manufactured by Nellcor, work in combination with the MEDLEY SpO2 Module to utilize Nellcor's technology to measure functional oxygen saturation and pulse rate. See Section 5 for a detailed device description and comparison.
## SUBSTANTIAL EQUIVALENCE
The MEDLEY SpO2 Module, Model 8210 is similar to the predicate devices (MEDLEY SpO2 Module, Model 8220, K010966, and the Nellcor OxiMax Pulse Oximetry System, K012891) in that they have the same intended use, operating principles, technological design. incorporate similar materials and manufacturing processes. The changes as described in this Special 510(k) pose no new issues of safety or efficacy. The MEDLEY SpO2 Module, Model 8210 described in this submission. in our opinion, is substantially equivalent to the predicate devices. See Section 6 for substantial equivalence details and a comparison table.
#### INTENDED USE
The MEDLEY SpO2 Module and accessories are intended for continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for adult, pediatric, and neonatal patients in hospitals and hospital-type facilities. This is the same intended use as previously cleared for the MEDLEY SpO2 Module, Model 8220, K010966. A separate Indications for Use statement is provided in Attachment B.
{2}------------------------------------------------
# SUMMARY OF SAFETY AND EFFECTIVENESS Page 3 of 3
## TECHNOLOGY
The MEDLEY SpO2 Module, Model 8210, as described in this Special 510(k) Premarket Notification, and the MEDLEY SpO2 Module, Model 8220 (K010966) are similar devices. The pulse oximeter systems have the same intended use, principle of operation, technology, and functionality. In addition, the MEDLEY SpO2 Module, Model 8210 (with Nellcor technology) and the MEDLEY SpO2 Module, Model 8220 (with Masimo technology) have a similar method of operation and utilize similar accessories. The minor differences as identified in this submission do not alter the fundamental technology of the predicate device.
#### SYSTEM AND SOFTWARE PERFORMANCE
The performance information provided demonstrates that the MEDLEY SpO2 Module, Model 8210 and accessories perform as intended and within the required specifications to demonstrate the safety and effectiveness of the device. All required verification testing for the MEDLEY SpO2 Module, Model 8210 (Nellcor) has been completed. Design verification testing including software V&V is complete, acceptable and design output matches design input. See Section 8 for detailed system and software performance information.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 1 2002
Ms. Renee L. Fluet Principal Regulatory Affairs Specialist Alaris Medical Systems, Incorporated 10221 Wateridge Circle San Diego, California 92121
Re: K022677
Trade/Device Name: MEDLEY SpO2 Module and Accessories Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: August 9, 2002 Received: August 12, 2002
Dear Ms. Fluet:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Ms. Fluet
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646 . Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Timothy A. Ulatowski
Timothy Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Attachment B
Image /page/5/Picture/3 description: The image shows the logo for Alaris Medical Systems. The logo features a stylized star to the left of the word "ALARIS" in a bold, sans-serif font. Below "ALARIS" are the words "MEDICAL SYSTEMS" in a smaller, sans-serif font. The logo is black and white.
MEDLEY SpO2 Module, Model 8210 Indications for Use
510(k) Number:
KO226 17 - - (To Be Assigned By FDA)
Device Trade Name:
MEDLEY SpO2 Module
Indications For Use: The MEDLEY SpO2 Module and accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. It is intended for use with neonatal, pediatric, and adult patients during both no motion and motion conditions, and for patients who are either well or poorly perfused, in hospitals and hospital-type facilities.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ﮐﮯ Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
Signature
(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number: K
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.