K022667 · Health Care Mfg., Inc. · KPR · Oct 22, 2002 · Radiology
Device Facts
Record ID
K022667
Device Name
HCMI X-RAY
Applicant
Health Care Mfg., Inc.
Product Code
KPR · Radiology
Decision Date
Oct 22, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
HCMI X-Ray High Frequency Radiographic System is intended for use by a qualified/trained doctor or technician or both to perform general purpose radiographic examinations of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing or lying in the prone or supine position.
Device Story
HCMI X-Ray is a stationary high-frequency radiographic system operating at 120/220 V AC. System includes high-frequency generators (100 kHz, 25-600 mA, 40-125 kVp), radiographic table (elevating or non-elevating), FST-10 tubestand, and optional wall stand with bucky. Operated by qualified doctors or technicians in clinical settings to perform general purpose radiographic examinations. Device generates X-ray exposures for diagnostic imaging of skull, spine, chest, abdomen, and extremities. Healthcare providers use resulting images to support clinical decision-making and diagnosis. Benefits include standardized radiographic imaging capabilities for various patient positions.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness established through bench, electrical safety, EMC, and user testing comparisons against the predicate device.
Technological Characteristics
Stationary high-frequency X-ray system; 100 kHz generators; 120/220 V AC power; includes radiographic table and optional wall stand with bucky. Complies with 21 CFR 1020.30, UL-2601 (IEC-60601), and UL187. Features anatomical programming.
Indications for Use
Indicated for general purpose radiographic examinations of skull, spinal column, chest, abdomen, and extremities in adult and pediatric patients. Performed with patient sitting, standing, or in prone/supine position. Prescription use only.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
OCT 2 2 2002
Image /page/0/Picture/1 description: The image shows a document with the title "510(k) Summary of Safety and Effectiveness". The document includes the logo and contact information for Health Care MFG., INC. The address is listed as 2146 East Pythian St., Springfield, MO 65802. The phone number is 417-864-6511, and the fax number is 417-864-7394. The document also includes the number K022667.
EXHIBIT 2
August 9, 2002 Contact: Katrina Moulder, Official Correspondent
- 1. Identification of the Device:
Proprietary-Trade Name: "HCMI X-Ray" High Frequency Radiographic System
Classification Name: Stationary X-Ray System, Product Code 90 KPR Common/Usual Name: Stationary X-Ray System
- 2. Equivalent legally marketed devices This product is similar in function to the Bennett X-Ray Radiographic System K952672
- 3. Indications for Use (intended use) : "HCMI X-Ray" High Frequency Radiographic System is intended for use by a qualified/trained doctor or technician or both to perform general purpose radiographic examinations of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing or lying in the prone or supine position
- 4. Description of the Device: The "HCMI X-Ray" is a radiographic high frequency stationary x-ray unit which operates from 120/220 V 50-60~ AC. HCMI X-Ray Systems available: They are always sold with a radiographic table, either elevating or non-elevating. A wall stand with bucky is optional. The FST-10 tubestand is supplied.
| HCMI 100 kHz Generators with Anatomical Programming | | | |
|-----------------------------------------------------|----------|-----------|--------|
| Model | mA Range | kVp Range | Max KW |
| HF-300 | 25-300 | 40-125 | 30 |
| HF-600 | 25-600 | 40-125 | 37.5 |
| HF-3SE | 25-300 | 40-125 | 30 |
| HF-3SEE | 25-300 | 40-125 | 30 |
{1}------------------------------------------------
- 5. Safety and Effectiveness, comparison to predicate device. The results of bench and user testing indicates that the new device is as safe and effective as the predicate devices.
| Characteristic | Bennett X-Ray<br>Radiographic System<br>K952672 | "HCMI X-Ray"<br>High Frequency<br>Radiographic<br>System |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| Intended Use: | Intended for use by a<br>qualified/trained doctor or<br>technician on both adult and<br>pediatric subjects for taking<br>diagnostic radiographic<br>exposures of the skull, spinal<br>column, chest, abdomen,<br>extremities, and other body<br>parts. Applications can be<br>performed with the patient<br>sitting, standing, or lying in<br>the prone or supine position. | SAME |
| Performance<br>Standard | 21 CFR 1020.30 | SAME |
| Electrical safety | Electrical Safety per<br>Underwriters Laboratories<br>Standard UL-2601(IEC-<br>60601) and IEC 60601,<br>Underwriters Laboratories<br>Standard UL187: UL<br>Standard for Safety for X-<br>Ray Equipment. | SAME, |
- 6. Substantial Equivalence Chart, "HCMI X-Ray" High Frequency Radiographic System
## 7. Conclusion
After analyzing bench, electrical safety, EMC, and user testing data, it is the conclusion of Health Care Manufacturing, Inc. that the "HCMI X-Ray" High Frequency Radiographic System is as safe and effective as the predicate device, has few technological differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a black and white seal of the Department of Health & Human Services. The seal features a stylized eagle with three lines emanating from its head, representing the department's mission to promote health and well-being. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## OCT 2 2 2002
Health Care MFG., Inc. % Daniel Kamm, P.E. Regulatory Engineer Kamm & Associates PO Box 7007 DEERFIELD IL 60015 USA Re: K022667
Trade/Device Name: "HCMI X-Ray" High Frequency Radiographic System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: 90 KPR Dated: August 9, 2002 Received: August 12, 2002
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Snigdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## j) Indications for Use
KU22667 510(k) Number
Device Name: "HCMI X-Ray" High Frequency Radiographic System
Indications for Use: "HCMI X-Ray" High Frequency Radiographic System is intended for use by a qualified/trained doctor or technician or both to perform general purpose radiographic examinations of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing or lying in the prone or supine position.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use OR Over the Counter Use (Per 21 CFR 801.109)
Nancy C. Brogdon
(Division Sign-Off) Division of Reproductive, Abdomina and Radiological Devices 510(k) Number -
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.