The LR DICOM Controller is a part of the LR 5200 Laser Film Recorder system, which is indicated for use in providing diagnostic quality medical images on film for aid in physician diagnosis, including the printing of images from various digital imaging source modalities, including, but not limited to, Computed Tomography (CT), Magnetic Resonance Imaging (MRI), Ultrasound, Computed Radiography, Digital Mammography, and Nuclear Medicine.
Device Story
LR DICOM Controller is a component of the LR 5200 Laser Film Recorder system. It functions as a network interface, receiving digital image data via Ethernet at speeds up to 100 MB/sec and transmitting it to the film recorder for hard copy output. The device replaces the controller function of the MG3000 component in the predicate system. It is used in clinical settings to facilitate the printing of diagnostic images from various modalities (CT, MRI, Ultrasound, etc.). Physicians use the resulting film for diagnostic purposes. The system supports high-speed network connectivity while maintaining compatibility with non-networked or non-DICOM modalities via the MG3000 interface.
Clinical Evidence
Bench testing only. Performance verified through in-house reliability and imaging performance demonstration tests.
Technological Characteristics
Networked controller for laser film recorder; Ethernet interface (up to 100 MB/sec); modified communications board; compatible with DICOM standard; class II medical image hardcopy device.
Indications for Use
Indicated for use in providing diagnostic quality medical images on film to aid physician diagnosis. Applicable to patients undergoing CT, MRI, Ultrasound, Computed Radiography, Digital Mammography, or Nuclear Medicine imaging. Prescription use only.
Regulatory Classification
Identification
A medical image hardcopy device is a device that produces a visible printed record of a medical image and associated identification information. Examples include multiformat cameras and laser printers.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
11022658
#### 6 2002 510(k) Summary SEP
# LR DICOM Controller
Common/Classification Name: Medical Image Hard Copy Device 21 CFR 892.2040
> Agfa Corporation 10 South Academy Street Greenville, SC 29602-9048
Contact: Jeff Jedlicka; Prepared: August 9, 2002
### LEGALLY MARKETED PREDICATE DEVICES A.
The predicate device is the LR 5200 Printer and Controller, which was cleared originally by FDA on 5 November 1996 as K964414 and also with clarified indications on 27 June 2001 as K012010.
### B. DEVICE DESCRIPTION
The modification involves replacing the controller function of the currently marketed MG3000 component of the LR5200 system so that a new controller can transmit image information from an Ethernet to the LR 5200 at up to 100 MB/sec. The communications board in the LR5200 will also be modified to accept the new communications approach. The MG3000 will still be used in some applications, e.g., to interface a non-networked imaging modality directly to the LR 5200, or to interface a non-DICOM modality to a network.
# C. INTENDED USE
The LR DICOM Controller is a part of the LR 5200 Laser Film Recorder system, which is indicated for use in providing diagnostic quality medical images on film for aid in physician diagnosis, including the printing of images from various digital imaging source modalities, including, but not limited to, Computed Tomography (CT), Magnetic Resonance Imaging (MRI), Ultrasound, Computed Radiography, Digital Mammography, and Nuclear Medicine.
#### D. SUBSTANTIAL EQUIVALENCE SUMMARY
The modified LR 5200 Printer/Controller has the same indications for use as the legally marketed predicate device. The modified device
{1}------------------------------------------------
has the same technological characteristics as the currently marketed This premarket notification will describe most of the device. characteristics of the modified LR DICOM Controller in sufficient detail to assure substantial equivalence. For the few characteristics that may not be precise enough to ensure equivalence, performance data or certifications are provided.
# E. TECHNOLOGICAL CHARACTERISTICS
The technological characteristics are the same in the proposed and predicate devices.
# F. TESTING
The modified LR DICOM Controller has been tested for proper performance to specifications through various in-house reliability and imaging performance demonstration tests.
# G. CONCLUSIONS
This pre-market submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Jeff Jedlicka Manager of Regulatory Affairs Agfa Corporation 10 South Academy Street Mail Stop 100 GREENVILLE SC 29601
Re: K022658
Trade/Device Name: LR DICOM Controller Regulation Number: 21 CFR 892.2040 Regulation Name: Medical image hardcopy device Regulatory Class: II Product Code: 90 LMC Dated: August 7, 2002 Received: August 9, 2002
Dear Mr. Jedlicka:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050,
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nredicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-459- |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-461- |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-461- |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-465- |
| Other | (301) 594-469- |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Snigdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
STATEMENT OF INDICATIONS FOR USE
510(k) Number (if known): 长۵2 - 2658
Device Name:
Indications For Use:
The LR 5200 Laser Film Recorder is indicated for use in providing diagnostic quality medical images on film for aid in physician diagnosis, including the printing of images from various digital imaging source modalities, including, but not limited to, Computed Tomography (CT), Magnetic Resonance Imaging (MRI), Ultrasound, Computed Radiography, Digital Mammography, and Nuclear Medicine.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Vernil C. Bergman
(Division Sign-Off) Division of Reproductive, Abdomina and Radiological Devices 510(k) Number_
Prescription Use
(Per 21 CFR 801.109)
OR
:
Over-The-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.